Compact Two Degrees-of-Freedom External Fixator System for Correction of Persistent Clubfoot Deformity

被引:3
|
作者
Wu, Ying Ying [1 ]
Plakseychuk, Anton [2 ]
Shimada, Kenji [3 ]
机构
[1] Carnegie Mellon Univ, Dept Biomed Engn, 5000 Forbes Ave, Pittsburgh, PA 15213 USA
[2] Univ Pittsburgh, Med Ctr, Magee Womens Hosp, Ctr Bone & Joint, 300 Halket St, Pittsburgh, PA 15213 USA
[3] Carnegie Mellon Univ, Dept Mech Engn, 5000 Forbes Ave, Pittsburgh, PA 15213 USA
关键词
PONSETI METHOD; FOOT; MANAGEMENT; RELAPSE;
D O I
10.1115/1.4043109
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Bone deformities are often complex three-dimensional (3D) deformities, and correcting them is difficult. To correct persistent clubfoot deformity in adolescents or adults, an external fixator is sometimes used to encourage tissue growth and preserve healthy tissues. However, it is difficult to set up, resulting in long surgeries and steep learning curves for surgeons. It is also bulky and obstructs patient mobility. In this paper, we introduce a new approach of defining clubfoot deformity correction as a six degrees-of-freedom (6DOF) correction, and then reducing it to just two degrees-of-freedom (2DOF) using the axis-angle representation. Therefore, only two physical trajectory joints are needed, which in turn enables a more compact fixator design. A computer planner was developed to minimize the bulk of the external fixator, and to optimize the distraction schedule to avoid overstretching the soft tissues. This reduces the learning curve for surgeons and shortens surgery time. To validate the system, a patient-specific clubfoot simulator was developed, and four experiments were performed on the clubfoot simulator. The accuracy of midfoot correction was 11mm and 3.5 deg without loading, and 41mm and 11.7 deg with loading. While the external fixator has to be more rigid to overcome resistance against correction, the surgical system itself was able to achieve accurate correction in less than 2 h. This is an improvement from the current method, which takes 2.5-4.5 h.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Stability of three degrees-of-freedom auto-parametric system
    He, Ji-Huan
    Amer, T. S.
    Abolila, A. F.
    Galal, A. A.
    ALEXANDRIA ENGINEERING JOURNAL, 2022, 61 (11) : 8393 - 8415
  • [32] Foot Deformity Correction with Hexapod External Fixator, the Ortho-SUV Frame™
    Takata, Munetomo
    Vilensky, Victor A.
    Tsuchiya, Hiroyuki
    Solomin, Leonid N.
    JOURNAL OF FOOT & ANKLE SURGERY, 2013, 52 (03): : 324 - 330
  • [33] Use of an external ring fixator for correction of an acquired angular limb deformity in a donkey
    Eggleston, RB
    Mueller, POE
    Chambers, JN
    Bentley, A
    JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2000, 217 (08): : 1186 - +
  • [34] Efficacy and Complications of External Fixator-Assisted Correction in Cubitus Varus Deformity
    Chauhan, Shrey
    Gupta, Parmanand
    Soni, Ashwani
    INDIAN JOURNAL OF ORTHOPAEDICS, 2025, 59 (02) : 156 - 163
  • [35] Normal mode localization for a two degrees-of-freedom system with quadratic and cubic non-linearities
    Bhattacharyya, R
    Jain, P
    Nair, A
    JOURNAL OF SOUND AND VIBRATION, 2002, 249 (05) : 909 - 919
  • [36] Humeral lengthening and proximal deformity correction with monorail external fixator in young adults
    Hongjiang Ruan
    Yi Zhu
    Shen Liu
    Qinglin Kang
    International Orthopaedics, 2018, 42 : 1107 - 1111
  • [37] Humeral lengthening and proximal deformity correction with monorail external fixator in young adults
    Ruan, Hongjiang
    Zhu, Yi
    Liu, Shen
    Kang, Qinglin
    INTERNATIONAL ORTHOPAEDICS, 2018, 42 (05) : 1107 - 1111
  • [38] Hemicallotasis for correction of varus deformity of the proximal tibia using a unilateral external fixator
    Oh, Chang-Wug
    Kim, Sung-Jung
    Park, Sung-Ki
    Kim, Hee-June
    Kyung, Hee-Soo
    Cho, Hwan-Sung
    Park, Byung-Chul
    Ihn, Joo-Chul
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2011, 16 (01) : 44 - 50
  • [39] A hexapod robot external fixator for computer assisted fracture reduction and deformity correction
    Seide, K
    Faschingbauer, M
    Wenzl, ME
    Weinrich, N
    Juergens, C
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2004, 1 (01): : 64 - 69
  • [40] On Structural Patterns of Mechanical Systems with Impacts with One and Two Degrees-of-Freedom
    Blazejczyk-Okolewska, Barbara
    MATHEMATICAL PROBLEMS IN ENGINEERING, 2013, 2013