Fast Dynamic Reconstruction Algorithm with Joint Bilateral Filtering for Perfusion C-arm CT

被引:0
|
作者
Manhart, Michael [1 ]
Kowarschik, Markus [2 ]
Fieselmann, Andreas [1 ,3 ]
Yu Deuerling-Zheng [2 ]
Hornegger, Joachim [1 ,3 ]
机构
[1] Univ Erlangen Nurnberg, Pattern Recognit Lab, Dept Comp Sci, Martensstr 3, D-91058 Erlangen, Germany
[2] Siemens AG, Angiog & Intervent X Ray Syst, Forchheim, Germany
[3] Erlangen Grad Sch Adv Opt Technol SAOT, Erlangen, Germany
关键词
Perfusion imaging; dynamic reconstruction; C-arm CT; stroke treatment; MODEL;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Tissue perfusion measurement using C-arm angiography systems is a novel technique with potential high benefit for catheter-guided treatment of stroke in the interventional suite. However, perfusion C-arm CT (PCCT) is challenging: the slow C-arm rotation speed only allows measuring samples of contrast time attenuation curves (TACs) every 5 - 6 s if reconstruction algorithms for static data are used. Furthermore, the peaks of the tissue TACs typically lie in a range of 5 - 30 HU, thus perfusion imaging is very sensitive to noise. We present a dynamic, iterative reconstruction (DIR) approach to reconstruct TACs described by a weighted sum of linear spline functions. The optimization problem is solved using an appropriate initialization and a Landweber-based optimization strategy with a modified backprojection step. To reduce noise a novel regularization technique based on joint bilateral filtering (JBF) is introduced. The algorithm is evaluated using simulation data created with a dynamic cylindrical phantom, a realistic digital brain phantom and real measured data from an animal study with a canine stroke model. Results indicate that the DIR algorithm qualitatively and quantitatively improves reconstructed TACs and perfusion maps compared to classical Feldkamp (FDK) reconstruction. For the brain phantom study the Pearson correlation (PC) of the reconstructed cerebral blood flow (CBF) maps to the ground truth increased from 0.82 (FDK) to 0.87 (DIR). For the canine study the PC of the CBF maps to co-registered perfusion CT maps increased from 0.61 (FDK) to 0.73 (DIR).
引用
收藏
页码:2304 / 2311
页数:8
相关论文
共 50 条
  • [31] TV-regularized Iterative Image Reconstruction on a Mobile C-ARM CT
    Pan, Yongsheng
    Whitaker, Ross
    Cheryauka, Arvi
    Ferguson, Dave
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [32] Artifact Reduction in Spare-view Image Reconstruction in C-arm CT
    Xia, Dan
    Chang, Yu-Bing
    Zhang, Zheng
    Chen, Buxin
    Manak, Joe
    Sidky, Emil Y.
    Pan, Xiaochuan
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [33] In vitro evaluation of the imaging accuracy of C-arm conebeam CT in cerebral perfusion imaging
    Ganguly, A.
    Fieselmann, A.
    Boese, J.
    Rohkohl, C.
    Hornegger, J.
    Fahrig, R.
    MEDICAL PHYSICS, 2012, 39 (11) : 6652 - 6659
  • [34] C-arm CT soft tissue 3-D imaging with a mobile C-arm
    Ritter, D
    Mitschke, M
    Graumann, R
    CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 3 - 6
  • [35] SPATIO-TEMPORALLY REGULARIZED 4-D CARDIOVASCULAR C-ARM CT RECONSTRUCTION USING A PROXIMAL ALGORITHM
    Taubmann, Oliver
    Unberath, Mathias
    Lauritsch, Guenter
    Achenbach, Stephan
    Maier, Andreas
    2017 IEEE 14TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI 2017), 2017, : 52 - 55
  • [36] Feasibility of a new C-arm Angiography (DYNA-CT) based three-dimensional Coronary Reconstruction Algorithm
    Rittger, Harald
    Rieber, Johannes
    Sinha, Anil
    Schmidt, Martin
    Brachmann, Johannes
    AMERICAN JOURNAL OF CARDIOLOGY, 2009, 104 (6A): : 164D - 165D
  • [37] CT Imaging with a mobile C-arm prototype
    Cheryauka, Arvi
    Tubbs, Dave
    Langille, Vinton
    Kalya, Prabhanjana
    Smith, Brady
    Cherone, Rocco
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [38] Improved C-arm cardiac cone beam CT based on alternate reconstruction and segmentation
    Bai, X. (jackybxz163@163.com), 1600, Elsevier Ltd (13):
  • [39] Adrenal vein sampling with C-arm CT
    Kinnison, Malonnie
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 19 (01) : 153 - 153
  • [40] Improved C-arm cardiac cone beam CT based on alternate reconstruction and segmentation
    Liu, Bo
    Zhou, Fugen
    Bai, Xiangzhi