Conceptual design of a cryogen-free μMRI device

被引:4
|
作者
Authelet, G. [1 ]
Poirier-Quinot, M. [2 ]
Ginefri, J-C [2 ]
Bonelli, A. [1 ]
Baudouy, B. [1 ]
机构
[1] Unversite Paris Saclay, CEA, Irfu, F-91191 Gif Sur Yvette, France
[2] Univ Paris Sclay, Univ Paris Sud, CNRS, IR4M,UMR8081, Orsay, France
来源
关键词
NATURAL CIRCULATION LOOP; HEAT-TRANSFER; SURFACE COIL;
D O I
10.1088/1757-899X/278/1/012122
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
To perform Micro Magnetic Resonance Imaging (mu MRI) analysis on small regions such as skins, articulations or small animals, the required spatial resolution implies to dramatically improve the sensitivity of the detection. One way to go is to use small radio frequency superconducting coil that allow, among others, increasing significantly the signal-to-noise ratio. The RF probe, constituted of an optimized YBaCuO film coil cooled below nitrogen temperature, must be located no further than few millimeters from the biological region to be imaged in a clinical MRI magnet. To fulfill the medical environment and constraints, a cryogen free cooling scheme has been developed to maintain the superconducting coil at the working temperature. The cryogenic design is based on a pulse tube cryocooler and solid thermal links inserted in a non-magnetic cryostat to avoid creating any electromagnetic perturbations to the MRI magnet and the measurements. We report here the conceptual design of the cryogenic system with the required thermal performances, the corresponding layout and architecture of the system as well as the main technical challenges met for the construction.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Risk reduction benefits of cryogen-free refrigeration
    Cournoyer, Michael E.
    Pecos, James M.
    Chunglo, Steve D.
    Bonner, Charles
    Foster, Lynn A.
    Maez, Rudy J.
    Punjak, Wayne A.
    Wilburn, Dianne W.
    JOURNAL OF CHEMICAL HEALTH & SAFETY, 2012, 19 (03): : 29 - 33
  • [32] Conceptual Design of a 33 T Cryogen-Free Magnet REBCO Insert: Mechanical Aspects and Protection Against Thermal Runaway
    Badel, Arnaud
    Takahashi, Kohki
    Zampa, Alexndre
    Okada, Tatsunori
    Tsuchiya, Yuji
    Awaji, Satoshi
    Uto, Tatsuro
    Tosaka, Taizo
    Takewa, Hiroyuki
    Nezuka, Hayato
    Hanai, Satoshi
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2024, 34 (05) : 1 - 5
  • [33] Scanning SQUID microscopy in a cryogen-free cooler
    Shperber, Yishai
    Vardi, Naor
    Persky, Eylon
    Wissberg, Shai
    Huber, Martin E.
    Kalisky, Beena
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (05):
  • [34] Cryogen-free one hundred microkelvin refrigerator
    Yan, Jiaojie
    Yao, Jianing
    Shvarts, Vladimir
    Du, Rui-Rui
    Lin, Xi
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2021, 92 (02):
  • [35] Structural analysis of a cryogen-free refrigerator for space
    Emes, MR
    Hepburn, ID
    Ray, RJ
    Worth, LBC
    CRYOGENICS, 2001, 41 (11-12) : 771 - 779
  • [36] Advances in the first cryogen-free hybrid magnet
    Watanabe, K
    Nishijima, G
    Awaji, S
    Takahashi, K
    Koyama, K
    Motokawa, M
    Ishizuka, A
    Hasebe, T
    Sakuraba, J
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2004, 14 (02) : 388 - 392
  • [37] Design of YBCO Insert Coil for a Cryogen-Free 22 T Superconducting Magnet
    Miyazaki, Hiroshi
    Iwai, Sadanori
    Tosaka, Taizo
    Tasaki, Kenji
    Hanai, Satoshi
    Ioka, Shigeru
    Watanabe, Kazuo
    Awaji, Satoshi
    Oguro, Hidetoshi
    Fujita, Shinji
    Daibo, Masanori
    Iijima, Yasuhiro
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2014, 24 (03)
  • [38] Cryogen-free hybrid magnet for magnetic levitation
    Watanabe, K
    Takahashi, K
    Mogi, I
    Nishijima, G
    Awaji, S
    Motokawa, M
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2003, 386 : 485 - 489
  • [39] Wide Variety of Experiments Using a Cryogen-Free 27.5 T Hybrid Magnet and a Cryogen-Free 18.1 T Superconducting Magnet
    Watanabe, K.
    Awaji, S.
    Oguro, H.
    JOURNAL OF LOW TEMPERATURE PHYSICS, 2013, 170 (5-6) : 503 - 510
  • [40] Test Results for a 1.5 T MRI System Utilizing a Cryogen-free YBCO Magnet
    Slade, Robert A.
    Parkinson, Benjamin J.
    Walsh, Rowan M.