A steady-state high-temperature apparatus for measuring thermal conductivity of ceramics

被引:34
|
作者
Filla, BJ
机构
[1] Materials Reliability Division, Natl. Inst. of Std. and Technology, Boulder
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1997年 / 68卷 / 07期
关键词
D O I
10.1063/1.1148202
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A one-sided very-high-temperature guarded hot plate has been built to measure thermal conductivity of monolithic ceramics, ceramic composites, thermal barrier coatings, functional graded materials, and high-temperature metal alloys. It is an absolute, steady-state measurement. device with an operational temperature range of 400-1400 K. Measurements are made in an atmosphere of low-pressure helium. Specimens examined in this apparatus are 70 mm in diameter, with thicknesses ranging between 1 and 8 mm. Optimal specimen thermal conductivities fall in the range of 0.5-30 W/(mK). Internal heated components are composed entirely of high-purity aluminum oxide, boron nitride, beryllium oxide, and fibrous alumina insulation board. Pure nickel and thermocouple-grade platinum-based alloys are the only metals used in the system. Apparatus design, modeling, and operation are described, along with the methods of data analysis that are unique to this system. An analysis of measurement uncertainty yields a combined measurement uncertainty of +/-5%. Experimental measurements on several materials are presented to illustrate the precision and bias of the apparatus.
引用
收藏
页码:2822 / 2829
页数:8