Point-of-Care Amperometric Testing of Diabetic Marker ( HbA1c) Using Specific Electroactive Antibodies

被引:19
|
作者
Chopra, Adity [1 ]
Rawat, Sanjay [1 ]
Bhalla, Vijayender [1 ]
Suri, C. Raman [1 ]
机构
[1] Inst Microbial Technol CSIR, Sect 39A, Chandigarh 160036, India
关键词
HbA1c; Ferrocene tagged antibody; Self-assembled monolayer; Electrochemical immunoassay; IMMUNOASSAY; PERFORMANCE; HBA(1C);
D O I
10.1002/elan.201300646
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Specific immune detection of glycated hemoglobin is still a great challenge owing to the small epitopic difference between Hemoglobin (Hb) and HbA1c. We report a new electrochemical immunoassay format for point of care testing of HbA1c. A conducting self-assembled monolayer of mercaptophenyl boronic acid (MPBA) was used as a capture layer for binding of glycated proteins and ferrocene tagged anti-HbA1c antibody (FcAb) as a tracer molecule on a gold screen printed electrode. Validation of the new HbA1c assay was carried out using 6 clinical samples with known HPLC values and a correlation coefficient of 98% was observed.
引用
收藏
页码:469 / 472
页数:4
相关论文
共 50 条
  • [1] Accuracy of point-of-care HbA1c testing in pregnant women
    Culliney, Katherine
    McCowan, Lesley M. E.
    Okesene-Gafa, Karaponi
    Murphy, Rinki
    Rowan, Janet
    Taylor, Rennae S.
    Mckinlay, Christopher J. D.
    [J]. AUSTRALIAN & NEW ZEALAND JOURNAL OF OBSTETRICS & GYNAECOLOGY, 2018, 58 (06): : 643 - 647
  • [2] VARIATION IN POINT-OF-CARE TESTING OF HBA1C IN DIABETES CARE IN GENERAL PRACTICE
    Kristensen, T.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2016, 18 : A64 - A64
  • [3] Variation in Point-of-Care Testing of HbA1c in Diabetes Care in General Practice
    Kristensen, Troels
    Waldorff, Frans Boch
    Nexoe, Jorgen
    Skovsgaard, Christian Volmar
    Olsen, Kim Rose
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2017, 14 (11):
  • [4] Accuracy and Clinical Utility of a Point-of-Care HbA1c Testing Device
    Knaebel, Jennifer
    Irvin, Benjamin R.
    Xie, Charles Z.
    [J]. POSTGRADUATE MEDICINE, 2013, 125 (03) : 91 - 98
  • [5] Is point-of-care HBA1C testing suitable for diagnosis of diabetes mellitus?
    Tollaanes, M. C.
    Flesche, K.
    Stavelin, A. V.
    Christensen, N. G.
    Sandberg, S.
    [J]. CLINICA CHIMICA ACTA, 2019, 493 : S644 - S644
  • [6] Evaluation of a HbA1c point-of-care analyzer
    Menendez-Valladares, Paloma
    Fernandez-Riejos, Patricia
    Sanchez-Mora, Catalina
    Perez-Perez, Antonio
    Sanchez-Margalet, Victor
    Gonzalez-Rodriguez, Concepcion
    [J]. CLINICAL BIOCHEMISTRY, 2015, 48 (10-11) : 686 - 689
  • [7] Impact of HbA1c Point of care testing implementation in the diabetic clinic
    Abdullah, H.
    [J]. CLINICA CHIMICA ACTA, 2024, 558
  • [8] Analysis of a Point-of-Care HbA1c Assay: Is It Time for an HbA1c Error Grid?
    Krouwer, Jan S.
    [J]. JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2020, 14 (05): : 896 - 897
  • [9] Variation between Point-of-care and Laboratory HbA1c testing in Clinical Practice
    Fellows, Shawn E.
    Cipriano, Gabriela C.
    [J]. PUERTO RICO HEALTH SCIENCES JOURNAL, 2019, 38 (03) : 189 - 191
  • [10] Reducing analytical variation between point-of-care and laboratory HbA1c testing
    Malkani, Samir
    Korpi-Steiner, Nichole
    Rao, Lokinendi V.
    [J]. JOURNAL OF DIABETES, 2013, 5 (02) : 192 - 196