Reducing analytical variation between point-of-care and laboratory HbA1c testing

被引:4
|
作者
Malkani, Samir [1 ]
Korpi-Steiner, Nichole [2 ]
Rao, Lokinendi V. [2 ]
机构
[1] Univ Massachusetts, Dept Med, Sch Med, Amherst, MA 01003 USA
[2] Univ Massachusetts, Dept Pathol, Med Sch Worcester, Amherst, MA 01003 USA
关键词
calibration; HbA1c; point-of-care systems; GLYCEMIC CONTROL; HEMOGLOBIN A1C; AVAILABILITY;
D O I
10.1111/1753-0407.12009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Point-of-care (POC) HbA1c testing allows for timely treatment changes, improved glycemic control, and patient and provider satisfaction. Substantial variation between POC and laboratory HbA1c results has been reported. At our university hospital diabetes clinic, we observed significant negative bias in HbA1c with the DCA Vantage (TM) (Siemens Healthcare Diagnostics, Tarrytown, NY, USA) compared with the Tosoh G8 HPLC laboratory analyzer (Tosoh Bioscience, San Francisco, CA, USA). This led us to systematically analyze the bias with the goal of recalibrating the DCA to minimize bias. Methods: We analyzed 45 patient samples, with HbA1c ranging between 5% and 10.8%, concurrently on two DCA analyzers and on the Tosoh G8 machine. The bias for each sample was the difference between the value on the DCA and the Tosoh 08 analyzer. Based on regression equations derived from the data, a correction factor for each DCA analyzer was calculated. The analyzers were recalibrated and retested for bias. Results: At baseline, the mean bias (range) was -0.5229 (+0.1 to -1.3) for Analyzer 1 and -0.5348 (0.0 to -1.6) for Analyzer 2. After recalibration, the mean bias (range) was 0.000 (+0.6 to -0.6) and 0.0003 (+0.5 to -0.5) for Analyzers 1 and 2, respectively, and the systematic negative bias seen prior to the calibration was almost eliminated. Conclusions: We recommend periodic recalibration of POC analyzers to eliminate systematic unidirectional bias and to harmonize results between the POC and central laboratory analyzers within a healthcare system. Calibration may need to be repeated with any change in the reagent lot.
引用
收藏
页码:192 / 196
页数:5
相关论文
共 50 条
  • [1] Variation between Point-of-care and Laboratory HbA1c testing in Clinical Practice
    Fellows, Shawn E.
    Cipriano, Gabriela C.
    [J]. PUERTO RICO HEALTH SCIENCES JOURNAL, 2019, 38 (03) : 189 - 191
  • [2] VARIATION IN POINT-OF-CARE TESTING OF HBA1C IN DIABETES CARE IN GENERAL PRACTICE
    Kristensen, T.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2016, 18 : A64 - A64
  • [3] Variation in Point-of-Care Testing of HbA1c in Diabetes Care in General Practice
    Kristensen, Troels
    Waldorff, Frans Boch
    Nexoe, Jorgen
    Skovsgaard, Christian Volmar
    Olsen, Kim Rose
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2017, 14 (11):
  • [4] Comparison of point-of-care HbA1c test versus standardized laboratory testing
    Sicard, DA
    Taylor, JR
    [J]. ANNALS OF PHARMACOTHERAPY, 2005, 39 (06) : 1024 - 1028
  • [5] Accuracy of point-of-care HbA1c testing in pregnant women
    Culliney, Katherine
    McCowan, Lesley M. E.
    Okesene-Gafa, Karaponi
    Murphy, Rinki
    Rowan, Janet
    Taylor, Rennae S.
    Mckinlay, Christopher J. D.
    [J]. AUSTRALIAN & NEW ZEALAND JOURNAL OF OBSTETRICS & GYNAECOLOGY, 2018, 58 (06): : 643 - 647
  • [6] Comparison between a new point-of-care HbA1c test and a routine laboratory method
    Relker, L.
    Luppa, P. B.
    [J]. CLINICA CHIMICA ACTA, 2024, 558
  • [7] Accuracy and Clinical Utility of a Point-of-Care HbA1c Testing Device
    Knaebel, Jennifer
    Irvin, Benjamin R.
    Xie, Charles Z.
    [J]. POSTGRADUATE MEDICINE, 2013, 125 (03) : 91 - 98
  • [8] Is point-of-care HBA1C testing suitable for diagnosis of diabetes mellitus?
    Tollaanes, M. C.
    Flesche, K.
    Stavelin, A. V.
    Christensen, N. G.
    Sandberg, S.
    [J]. CLINICA CHIMICA ACTA, 2019, 493 : S644 - S644
  • [9] Evaluation of a HbA1c point-of-care analyzer
    Menendez-Valladares, Paloma
    Fernandez-Riejos, Patricia
    Sanchez-Mora, Catalina
    Perez-Perez, Antonio
    Sanchez-Margalet, Victor
    Gonzalez-Rodriguez, Concepcion
    [J]. CLINICAL BIOCHEMISTRY, 2015, 48 (10-11) : 686 - 689
  • [10] Analytical performance of point-of-care device QuikRead go in HbA1c analysis
    Linko-Parvinen, Anna
    Dahlstrom, Johanna
    Pallari, Hanna-Mari
    [J]. JOURNAL OF LABORATORY AND PRECISION MEDICINE, 2023, 8