Flow injection-assisted optical sensor for determination of iron(II) and iron(III) in natural water

被引:15
|
作者
daConceicao, ACL [1 ]
Tena, MT [1 ]
dosSantos, MMC [1 ]
Goncalves, MLS [1 ]
deCastro, MDL [1 ]
机构
[1] UNIV CORDOBA,FAC SCI,DEPT ANALYT CHEM,E-14004 CORDOBA,SPAIN
关键词
flow injection; sensors; iron; waters;
D O I
10.1016/S0003-2670(96)00595-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An approach for speciation of iron (as Fe(II) and Fe(III)) based on integration of retention of the Fe(III)-SCN complex with detection using a conventional spectrophotometer is proposed here. The device (namely, a flow-cell packed with an exchange resin) has been coupled to a flow-injection manifold with inner-coupled injection valves which enables discrimination between Fe(III) and Fe(II) taking advantage of a redox minicolumn housed in the loop of one of the valves. The method thus proposed has good selectivity with a determination limit of 80 mu gl(-1) and affords a determination range of 80-500 mu gl(-1) for Fe(II) plus Fe(III). Application to synthetic and real samples containing both the oxidation states of iron lead to average recoveries of 103% and 101% for Fe(II) and total iron (Fe(II)+ Fe(III)), respectively, thus showing the usefulness of the overall approach.
引用
收藏
页码:191 / 197
页数:7
相关论文
共 50 条