Accuracy of oscillometric blood pressure monitors for the detection of atrial fibrillation: a systematic review

被引:0
|
作者
Verberk, Willem J. [1 ]
de Leeuw, Peter W.
机构
[1] Maastricht Univ, Cardiovasc Res Inst Maastricht CARIM, Maastricht, Netherlands
关键词
arrhythmia; atrial fibrillation; blood pressure measurement; oscillometric blood pressure monitor; primary care; screening; stroke; EUROPEAN-SOCIETY; GENERAL-PRACTICE; PREVALENCE; GUIDELINES; MANAGEMENT; TRENDS; HOME; ESC;
D O I
10.1586/ERD.12.46
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Atrial fibrillation (AF) is a significant risk factor for stroke and early detection of AF may help to identify patients in need of treatment. Automated blood pressure (BP) monitors with implemented AF or arrhythmia detection systems may be a useful tool for early diagnosis of AF. A systematic review (Medline/PubMed, Embase, Cochrane) of studies was performed to assess the accuracy of modified BP monitors (for diagnosing AF). A total of five studies (four tests in the physician's office and one at home) were selected. For the most accurate AF detection, three sequential BP measurements should be performed. Direct comparison against a 12-lead ECG showed that the highest sensitivity, 97% (95% CI: 94-100%), for detecting AF was obtained when three readings were assessed with two or three AF-positive readings. The highest specificity (97%) was obtained when performing three measurements, of which all three must be AF positive. The modified BP monitor (Microlife Corporation, Taipei, Taiwan) has high potential in improving AF screening.
引用
收藏
页码:635 / 640
页数:6
相关论文
共 50 条
  • [31] Auscultatory versus oscillometric blood pressure measurement in patients with atrial fibrillation and arterial hypertension
    Aistėja Šelmytė–Besusparė
    Jūratė Barysienė
    Dovilė Petrikonytė
    Audrius Aidietis
    Germanas Marinskis
    Aleksandras Laucevičius
    BMC Cardiovascular Disorders, 17
  • [32] Reliability of Oscillometric Blood Pressure Monitoring in Atrial Fibrillation Patients Admitted for Electric Cardioversion
    Maselli, Monica
    Giantin, Valter
    Corrado, Domenico
    Franchin, Alessandro
    Attanasio, Francesca
    Pengo, Valentina
    Tramontano, Alessandra
    De Toni, Pietro
    Perissinotto, Egle
    Manzato, Enzo
    JOURNAL OF CLINICAL HYPERTENSION, 2015, 17 (07): : 558 - 564
  • [33] Impact of atrial fibrillation on the accuracy of oscillometric blood pressure monitoring in ICU patients from a large real-world database
    Li, Yuxi
    Wang, Beining
    Fan, Fangfang
    Weng, Haoyu
    Zhang, Long
    Jiang, Jie
    Zheng, Bo
    Zhang, Yan
    Li, Jianping
    JOURNAL OF HYPERTENSION, 2023, 41 (05) : 838 - 844
  • [34] Accuracy of Apple Watch for Detection of Atrial Fibrillation: A Systematic Review and Meta-Analysis
    Yasmin, Farah
    Moeed, Abdul
    Adil, Mariam
    Rahman, Saad Ur
    Wilson, F. Perry
    CIRCULATION, 2023, 148
  • [35] ACCURACY OF APPLE WATCH FOR DETECTION OF ATRIAL FIBRILLATION: A SYSTEMATIC REVIEW AND META-ANALYSIS
    Yasmin, Farah
    Umar, Muhammad
    Moeed, Abdul
    Adil, Mariam
    Asghar, Muhammad Sohaib
    Alraies, M. Chadi
    Wilson, Francis Perry
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 83 (13) : 2399 - 2399
  • [36] ARTIFICIAL INTELLIGENCE FOR THE DETECTION OF ATRIAL FIBRILLATION: A SYSTEMATIC REVIEW AND METANALYSIS OF DIAGNOSTIC ACCURACY STUDIES
    Manetas-Stavrakakis, Nikolaos
    Sotiropoulou, Ioanna Myrto
    Paraskevas, Themistoklis
    Stavrakaki, Stefania Maneta
    Bampatsias, Dimitrios
    Xanthopoulos, Andreas
    Papageorgiou, Nikolaos
    Briasoulis, Alexandros
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 83 (13) : 2645 - 2645
  • [37] Novel methods of testing and calibration of oscillometric blood pressure monitors
    Celler, Branko G.
    Argha, Ahmadreza
    Phu Ngoc Le
    Ambikairajah, Eliathamby
    PLOS ONE, 2018, 13 (08):
  • [38] Comparison of oscillometric blood pressure measurement by two clinical monitors
    Prashanth, G. P.
    INDIAN JOURNAL OF ANAESTHESIA, 2011, 55 (05) : 548 - U125
  • [39] An easy way to verify the accuracy of oscillometric blood pressure monitors by auscultatory method: a smartphone based App
    Zhang, Zhi
    Wang Bingjian
    Xi, Weichun
    Chu, Guang
    Fang, Wang
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 70 (16) : C138 - C138
  • [40] Logistic ordinal regression for the calibration of oscillometric blood pressure monitors
    Barbe, Kurt
    Kurylyak, Yuriy
    Lamonaca, Francesco
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2014, 11 : 89 - 96