Accuracy of oscillometric blood pressure monitors for the detection of atrial fibrillation: a systematic review

被引:0
|
作者
Verberk, Willem J. [1 ]
de Leeuw, Peter W.
机构
[1] Maastricht Univ, Cardiovasc Res Inst Maastricht CARIM, Maastricht, Netherlands
关键词
arrhythmia; atrial fibrillation; blood pressure measurement; oscillometric blood pressure monitor; primary care; screening; stroke; EUROPEAN-SOCIETY; GENERAL-PRACTICE; PREVALENCE; GUIDELINES; MANAGEMENT; TRENDS; HOME; ESC;
D O I
10.1586/ERD.12.46
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Atrial fibrillation (AF) is a significant risk factor for stroke and early detection of AF may help to identify patients in need of treatment. Automated blood pressure (BP) monitors with implemented AF or arrhythmia detection systems may be a useful tool for early diagnosis of AF. A systematic review (Medline/PubMed, Embase, Cochrane) of studies was performed to assess the accuracy of modified BP monitors (for diagnosing AF). A total of five studies (four tests in the physician's office and one at home) were selected. For the most accurate AF detection, three sequential BP measurements should be performed. Direct comparison against a 12-lead ECG showed that the highest sensitivity, 97% (95% CI: 94-100%), for detecting AF was obtained when three readings were assessed with two or three AF-positive readings. The highest specificity (97%) was obtained when performing three measurements, of which all three must be AF positive. The modified BP monitor (Microlife Corporation, Taipei, Taiwan) has high potential in improving AF screening.
引用
收藏
页码:635 / 640
页数:6
相关论文
共 50 条
  • [1] Opportunistic detection of atrial fibrillation using blood pressure monitors: a systematic review
    Kane, Sarah A.
    Blake, James R.
    McArdle, Frank J.
    Langley, Philip
    Sims, Andrew J.
    OPEN HEART, 2016, 3 (01):
  • [2] Accuracy of oscillometric blood pressure measurement in atrial fibrillation
    Feenstra, Rudolf K.
    Allaart, Cor P.
    Berkelmans, Gijs F. N.
    Westerhof, Berend E.
    Smulders, Yvo M.
    BLOOD PRESSURE MONITORING, 2018, 23 (02) : 59 - 63
  • [3] Should Oscillometric Blood Pressure Monitors Be Used in Patients With Atrial Fibrillation?
    Myers, Martin G.
    Stergiou, George S.
    JOURNAL OF CLINICAL HYPERTENSION, 2015, 17 (07): : 565 - 566
  • [4] Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation
    Lamb, Tyler S.
    Thakrar, Amar
    Ghosh, Mahua
    Wilson, Merne P.
    Wilson, Thomas W.
    CLINICAL AND INVESTIGATIVE MEDICINE, 2010, 33 (01): : E54 - E62
  • [5] Predictive validity of automated oscillometric blood pressure monitors for screening atrial fibrillation: a systematic review and meta-analysis
    Park, Seong-Hi
    Jun, Kyung Ja
    Choi, Yun-Kyoung
    EXPERT REVIEW OF MEDICAL DEVICES, 2019, 16 (06) : 503 - 514
  • [6] Impact of Atrial Fibrillation on the Accuracy of Oscillometric Blood Pressure Monitoring
    Pagonas, Nikolaos
    Schmidt, Sven
    Eysel, Joerg
    Compton, Friederike
    Hoffmann, Clemens
    Seibert, Felix
    Hilpert, Justus
    Tschoepe, Carsten
    Zidek, Walter
    Westhoff, Timm H.
    HYPERTENSION, 2013, 62 (03) : 579 - 584
  • [7] Accuracy of automated oscillometric blood pressure monitors in the detection of cardiac arrhythmias
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Zhang, Yi
    Wang, Jie
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2009, 14 (02) : 91 - 92
  • [8] Diagnostic accuracy of an oscillometric blood pressure monitor for atrial fibrillation screening
    Li, Linyi
    Bing, Sen
    Liu, Kui
    Jiang, Jun
    Zhang, Jieqiong
    Chen, Changsheng
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2023, 28 (03) : 144 - 148
  • [9] AUTOMATED DETECTION OF ATRIAL FIBRILLATION DURING ROUTINE OSCILLOMETRIC MEASUREMENT OF BLOOD PRESSURE: SYSTEMATIC REVIEW AND META-ANALYSIS
    Kyriakoulis, Konstantinos G.
    Kollias, Anastasios
    Chardouvelis, Panagiotis
    Ntousopoulos, Vasileios
    Menti, Ariadni
    Destounis, Antonios
    Stergiou, George S.
    JOURNAL OF HYPERTENSION, 2023, 41 : E20 - E20
  • [10] The impact of atrial fibrillation on accuracy of oscillometric blood pressure measurement: effect of ventricular rate
    Feng Xie
    Jingsong Xu
    Ling-lin Xia
    Xiao Luo
    Zhenzhen Jiang
    Yanqing Wu
    Hai Su
    Hypertension Research, 2020, 43 : 518 - 524