The relationship between measurement method and corneal structure on apparent intraocular pressure in glaucoma and ocular hypertension

被引:9
|
作者
Nessim, Maged [1 ]
Mollan, Susan P. [2 ]
Wolffsohn, James S. [3 ]
Laiquzzaman, Mohammad [4 ]
Sivakumar, Subramaniam [4 ]
Hartley, Stephanie [1 ]
Shah, Sunil [1 ,3 ,4 ]
机构
[1] City Hosp, Birmingham & Midland Eye Ctr, Birmingham B18 7QH, W Midlands, England
[2] Univ Hosp Birmingham NHS Trust, Queen Elizabeth Hosp, Dept Ophthalmol, Birmingham B15 2WB, W Midlands, England
[3] Aston Univ, Ophthalm Res Grp, Birmingham B4 7ET, W Midlands, England
[4] Heart England NHS Fdn Trust, Solihull Hosp, Birmingham B91 2JL, W Midlands, England
来源
CONTACT LENS & ANTERIOR EYE | 2013年 / 36卷 / 02期
关键词
Cornea; Intraocular pressure; Glaucoma; open angle; Ocular hypertension; Ocular tonometry; GOLDMANN APPLANATION TONOMETRY; DYNAMIC CONTOUR TONOMETRY; OPEN-ANGLE GLAUCOMA; NORMAL-TENSION GLAUCOMA; RESPONSE ANALYZER; BIOMECHANICAL PROPERTIES; NONCONTACT TONOMETRY; THICKNESS; HYSTERESIS; REDUCTION;
D O I
10.1016/j.clae.2012.11.001
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To analyse the relationship between measured intraocular pressure (IOP) and central corneal thickness (CCT), corneal hysteresis (CH) and corneal resistance factor (CRF) in ocular hypertension (OHT), primary open-angle (POAG) and normal tension glaucoma (NTG) eyes using multiple tonometry devices. Methods: Right eyes of patients diagnosed with OHT (n=47), normal tension glaucoma (n=17) and POAG (n=50) were assessed, IOP was measured in random order with four devices: Goldmann applanation tonometry (GAT); Pascal(R) dynamic contour tonometer (DCT); Reichert(R) ocular response analyser (ORA); and Tono-Pen(R) XL. CCT was then measured using a hand-held ultrasonic pachymeter. CH and CRF were derived from the air pressure to corneal reflectance relationship of the ORA data. Results: Compared to the GAT, the Tonopen and ORA Goldmann equivalent (IOPg) and corneal compensated (IOPcc) measured higher IOP readings (F=19.351, p<0.001), particularly in NTG (F=12.604, p<0.001). DCT was closest to Goldmann IOP and had the lowest variance. CCT was significantly different (F=8.305, p<0.001) between the 3 conditions as was CH (F=6.854, p=0.002) and CRF (F=19.653, p<0.001). IOPcc measures were not affected by CCT. The DCT was generally not affected by corneal biomechanical factors. Conclusion: This study suggests that as the true pressure of the eye cannot be determined non-invasively, measurements from any tonometer should be interpreted with care, particularly when alterations in the corneal tissue are suspected. (C) 2012 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:57 / 61
页数:5
相关论文
共 50 条
  • [1] Relationship between intraocular pressure and parameters of obesity in ocular hypertension
    Ahn, Min Won
    Lee, Ji Woong
    Shin, Jong Hoon
    Lee, Jong Soo
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2020, 13 (05) : 794 - 800
  • [2] Relationship between intraocular pressure and parameters of obesity in ocular hypertension
    Min Won Ahn
    Ji Woong Lee
    Jong Hoon Shin
    Jong Soo Lee
    International Journal of Ophthalmology, 2020, (05) : 794 - 800
  • [3] Central corneal thickness and intraocular pressure in ocular hypertension
    Consuelo Gutierrez
    Jose R. V. Casaponsa
    Francisco Munoz
    Gema Rebolleda
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2001, 239 : 463 - 463
  • [4] Central corneal thickness and intraocular pressure in ocular hypertension
    Gutierrez, C
    Casaponsa, JRV
    Munoz, F
    Rebolleda, G
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2001, 239 (06) : 463 - 463
  • [5] Relationship Between Fuchs Endothelial Corneal Dystrophy Severity and Glaucoma and/or Ocular Hypertension
    Nagarsheth, Mehul
    Singh, Annapurna
    Schmotzer, Brian
    Babineau, Denise C.
    Sugar, Joel
    Lee, W. Barry
    Iyengar, Sudha K.
    Lass, Jonathan H.
    ARCHIVES OF OPHTHALMOLOGY, 2012, 130 (11) : 1384 - 1388
  • [6] Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of Intraocular pressure
    Broman, Aimee Teo
    Congdon, Nathan G.
    Bandeen-Roche, Karen
    Quigley, Harry A.
    JOURNAL OF GLAUCOMA, 2007, 16 (07) : 581 - 588
  • [7] Intraocular pressure and progression in exfoliative eyes with ocular hypertension or glaucoma
    Harju, M
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 2000, 78 (06): : 699 - 702
  • [8] Intraocular pressure control and persistence on treatment in glaucoma and ocular hypertension
    Tingey, D
    Bernard, LM
    Grima, DT
    Miller, B
    Lam, A
    CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2005, 40 (02): : 161 - 169
  • [9] The Relationship Between Corneal Biomechanics and Optic Nerve Head Biomechanics in Glaucoma and Ocular Hypertension
    Bayoumi, A.
    Descovich, D.
    Hafez, A. S.
    Lesk, M. R.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [10] Relationship Between Corneal Biomechanical Properties, Central Corneal Thickness, and Intraocular Pressure Across the Spectrum of Glaucoma
    Kaushik, Sushmita
    Pandav, Surinder Singh
    Banger, Anupam
    Aggarwal, Kanika
    Gupta, Amod
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2012, 153 (05) : 840 - 849