The effect of a disposable probe cover on pulse oximetry
被引:3
|作者:
Cheung, P
Hardman, JG
Whiteside, R
机构:
[1] Ipswich Gen Hosp, Dept Anaesthesia, Ipswich, Qld, Australia
[2] Ipswich Gen Hosp, Dept Anaesthesia & Intens Care, Ipswich, Qld, Australia
关键词:
monitoring : pulse oximeter;
cover;
D O I:
10.1177/0310057X0203000215
中图分类号:
R614 [麻醉学];
学科分类号:
100217 ;
摘要:
The re-use of pulse oximeter probes presents the possibility, of between-patient contamination. Use of a disposable polyethylene cover may, reduce this risk. In a controlled, prospective study we examined the effect of such a cover oil the accuracy of pulse oximetry. Each of ten volunteer subjects was monitored simultaneously by two identical Nellcor pulse oximeters, one with a plastic cover and the other, without a cover, used as a control. The pulse oximetry (SpO(2)) reading for each probe was recorded while subjects breathed 21% O-2 and again while they breathed 10% O-2. The probe cover was then swapped onto the other probe and the recordings were repeated. Ninety-five per cent limits of agreement in SpO(2) (mean difference in SpO(2) (1.95 x standard deviation of difference) between covered and non-covered probes were -0.6% to 0.6% while breathing 21% oxygen and -2.0% to 2.9% while breathing 10% oxygen. We conclude that a protective plastic sheath may, induce a small error in pulse oximetry reading that is most marked during hypoxaemia. This error is unlikely to be of clinical significance.
引用
收藏
页码:211 / 214