A compact neural recording interface based on silicon microelectrode

被引:3
|
作者
Han JianQiang [1 ]
Zhang Xu [1 ]
Pei WeiHua [1 ]
Gui Qiang [1 ]
Liu Ming [1 ]
Chen HongDa [1 ]
机构
[1] Chinese Acad Sci, Inst Semicond, State Key Lab Integrated Optoelect, Beijing 100083, Peoples R China
基金
中国国家自然科学基金;
关键词
neural recording interface; silicon microelectrode; CMOS preamplifier; flip-chip bonding; hybrid device; ARRAYS; AMPLIFIER;
D O I
10.1007/s11431-013-5359-2
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A prototype of hybrid neural recording interface has been developed for extracellular neural recording. It consists of a silicon-based plane microelectrode array and a CMOS low noise neural amplifier chip. The neural amplifier chip is designed and implemented in 0.18 mu m N-well CMOS 1P6M technology. The area of the neural preamplifier is only 0.042 mm(2) with a gain of 48.3 dB. The input equivalent noise is 4.73 mu Vrms within pass bands of 4 kHz. To avoid cable tethering for high dense multichannel neural recording interface and make it compact, flip-chip bonding is used to integrate the preamplifier chip and the microelectrode together. The hybrid device measures 3 mmx5.5 mmx330 mu m, which is convenient for implant or in-vivo neural recording. The hybrid device was testified in in-vivo experiment. Neural signals were recorded from hippocampus region of anesthetized Sprague Dawley rats successfully.
引用
收藏
页码:2808 / 2813
页数:6
相关论文
共 50 条
  • [21] Distributed implantation of a flexible microelectrode array for neural recording
    Wei, Chunrong
    Wang, Yang
    Pei, Weihua
    Han, Xinyong
    Lin, Longnian
    Liu, Zhiduo
    Ming, Gege
    Chen, Ruru
    Wu, Pingping
    Yang, Xiaowei
    Zheng, Li
    Wang, Yijun
    MICROSYSTEMS & NANOENGINEERING, 2022, 8 (01)
  • [22] Silicon-based microelectrode arrays for stimulation and signal recording of in vitro cultured neurons
    PAN HaiXian 1
    2 Institute of RF-& OE-ICs
    3 Institute of Microelectronics
    Science China(Information Sciences), 2011, 54 (10) : 2199 - 2208
  • [23] Silicon-based microelectrode arrays for stimulation and signal recording of in vitro cultured neurons
    Pan HaiXian
    Lu XiaoYing
    Wang ZhiGong
    Ren TianLing
    Fang Tao
    Zhang Jie
    Zhou ChangJian
    Wang LiGang
    SCIENCE CHINA-INFORMATION SCIENCES, 2011, 54 (10) : 2199 - 2208
  • [24] Silicon-based microelectrode arrays for stimulation and signal recording of in vitro cultured neurons
    HaiXian Pan
    XiaoYing Lü
    ZhiGong Wang
    TianLing Ren
    Tao Fang
    Jie Zhang
    ChangJian Zhou
    LiGang Wang
    Science China Information Sciences, 2011, 54 : 2199 - 2208
  • [25] AN IMPLANTABLE CMOS CIRCUIT INTERFACE FOR MULTIPLEXED MICROELECTRODE RECORDING ARRAYS
    JI, J
    WISE, KD
    IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1992, 27 (03) : 433 - 443
  • [26] Brain-Silicon Interface for High-Resolution in vitro Neural Recording
    Aziz, Joseph N. Y.
    Genov, Roman
    Bardakjian, Berj L.
    Derchansky, Miron
    Carlen, Peter L.
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2007, 1 (01) : 56 - 62
  • [27] A Compact Recording Array for Neural Interfaces
    Leene, Lieuwe B.
    Liu, Yan
    Constandinou, Timothy G.
    2013 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS), 2013, : 97 - 100
  • [28] Magnetic resonance compatibility of multichannel silicon microelectrode systems for neural recording and stimulation: Design criteria, tests, and recommendations
    Santiesteban, FMM
    Swanson, SD
    Noll, DC
    Anderson, DJ
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (03) : 547 - 558
  • [29] PRACTICAL 24 CHANNEL MICROELECTRODE FOR NEURAL RECORDING IN VIVO.
    Kuperstein, Michael
    Whittington, Douglas A.
    IEEE Transactions on Biomedical Engineering, 1981, BME-28 (03) : 288 - 293
  • [30] Hemispherical Microelectrode Array for Ex Vivo Retinal Neural Recording
    Ha, Yoonhee
    Yoo, Hyun-Ji
    Shin, Soowon
    Jun, Sang Beom
    MICROMACHINES, 2020, 11 (05)