Imaging of prompt gamma rays emitted during delivery of clinical proton beams with a Compton camera: feasibility studies for range verification

被引:106
|
作者
Polf, Jerimy C. [1 ]
Avery, Stephen [2 ]
Mackin, Dennis S. [3 ]
Beddar, Sam [3 ]
机构
[1] Univ Maryland, Sch Med, Dept Radiat Oncol, Baltimore, MD 21201 USA
[2] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
[3] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2015年 / 60卷 / 18期
关键词
proton therapy; range verification; in vivo imaging; prompt gamma; THERAPY; PET; PATIENT; UNCERTAINTIES; EMISSION; ALGORITHM; POSITION; DESIGN;
D O I
10.1088/0031-9155/60/18/7085
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this paper is to evaluate the ability of a prototype Compton camera (CC) to measure prompt gamma rays (PG) emitted during delivery of clinical proton pencil beams for prompt gamma imaging (PGI) as a means of providing in vivo verification of the delivered proton radiotherapy beams. A water phantom was irradiated with clinical 114 MeV and 150 MeV proton pencil beams. Up to 500 cGy of dose was delivered per irradiation using clinical beam currents. The prototype CC was placed 15 cm from the beam central axis and PGs from 0.2 MeV up to 6.5 MeV were measured during irradiation. From the measured data (2D) images of the PG emission were reconstructed. (1D) profiles were extracted from the PG images and compared to measured depth dose curves of the delivered proton pencil beams. The CC was able to measure PG emission during delivery of both 114 MeV and 150 MeV proton beams at clinical beam currents. 2D images of the PG emission were reconstructed for single 150 MeV proton pencil beams as well as for a 5 x 5 cm mono-energetic layer of 114 MeV pencil beams. Shifts in the Bragg peak (BP) range were detectable on the 2D images. 1D profiles extracted from the PG images show that the distal falloff of the PG emission profile lined up well with the distal BP falloff. Shifts as small as 3 mm in the beam range could be detected from the 1D PG profiles with an accuracy of 1.5 mm or better. However, with the current CC prototype, a dose of 400 cGy was required to acquire adequate PG signal for 2D PG image reconstruction. It was possible to measure PG interactions with our prototype CC during delivery of proton pencil beams at clinical dose rates. Images of the PG emission could be reconstructed and shifts in the BP range were detectable. Therefore PGI with a CC for in vivo range verification during proton treatment delivery is feasible. However, improvements in the prototype CC detection efficiency and reconstruction algorithms are necessary to make it a clinically viable PGI system.
引用
收藏
页码:7085 / 7099
页数:15
相关论文
共 50 条
  • [21] Classification of Compton Camera Based Prompt Gamma Imaging for Proton Radiotherapy by Random Forests
    Barajas, Carlos A.
    Kroiz, Gerson C.
    Gobbert, Matthias K.
    Polf, Jerimy C.
    [J]. 2021 INTERNATIONAL CONFERENCE ON COMPUTATIONAL SCIENCE AND COMPUTATIONAL INTELLIGENCE (CSCI 2021), 2021, : 308 - 311
  • [22] Development of a Compton Camera for Online Range Monitoring of Laser-Accelerated Proton Beams via Prompt-Gamma Detection
    Thirolf, P. G.
    Lang, C.
    Aldawood, S.
    v.d. Kolff, H. G.
    Maier, L.
    Schaart, D. R.
    Parodi, K.
    [J]. INPC 2013 - INTERNATIONAL NUCLEAR PHYSICS CONFERENCE, VOL. 2, 2014, 66
  • [23] Proton therapy treatment verification with prompt gamma rays and fast neutrons - a feasibility study
    Pettersen, H. E. S.
    Alagoz, E.
    Hysing, L. B.
    Koegler, T.
    Lathouwers, D.
    Lionheart, W.
    Obhodas, J.
    Pausch, G.
    Ratliff, H. M.
    Rovituso, M.
    Schellhammer, S.
    Setterdahl, L. M.
    Skjerdal, K.
    Sudac, D.
    Turko, J. A.
    Ytre-Hauge, K. S.
    Meric, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S557 - S559
  • [24] Proton range monitoring using prompt gamma camera in clinical configuration
    Janssens, G.
    Smeets, J.
    Roellinghoff, F.
    Prieels, D.
    Stichelbaut, F.
    Celani, A.
    Perali, I.
    Fiorini, C.
    Clementel, E.
    Sterpin, E.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S158 - S159
  • [25] A Prompt Gamma-Ray Spectroscopy System for Clinical Studies of in Vivo Proton Range Verification
    Verburg, J.
    Bortfeld, T.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3757 - 3757
  • [26] Detection of mixed-range proton pencil beams with a prompt gamma slit camera
    Priegnitz, M.
    Helmbrecht, S.
    Janssens, G.
    Perali, I.
    Smeets, J.
    Stappen, F. Vander
    Sterpin, E.
    Fiedler, F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (02): : 855 - 871
  • [27] Simultaneous imaging of prompt gamma photons and prompt X-rays during irradiation of proton beams to human torso phantom at clinical dose level
    Yamamoto, Seiichi
    Yamashita, Tomohiro
    Kobashi, Yusuke
    Yabe, Takuya
    Akagi, Takashi
    Yamaguchi, Mitsutaka
    Kawachi, Naoki
    Kamada, Kei
    Yoshikawa, Akira
    Yokokawa, Hiromu
    Sagisaka, Mayu
    Kataoka, Jun
    [J]. JOURNAL OF INSTRUMENTATION, 2023, 18 (07)
  • [28] Range verification of passively scattered proton beams based on prompt gamma time patterns
    Testa, Mauro
    Min, Chul Hee
    Verburg, Joost M.
    Schuemann, Jan
    Lu, Hsiao-Ming
    Paganetti, Harald
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (15): : 4181 - 4195
  • [29] 3D prompt gamma imaging for proton beam range verification
    Draeger, E.
    Mackin, D.
    Peterson, S.
    Chen, H.
    Avery, S.
    Beddar, S.
    Polf, J. C.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (03):
  • [30] A 3 Dimensional Prompt Gamma Imaging System for Range Verification in Proton Radiotherapy
    Draeger, E.
    Chen, H.
    Mackin, D.
    Peterson, S.
    Avery, S.
    Beddar, S.
    Polf, J.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3757 - 3757