Improved Opacification of a Suboptimally Enhanced Pulmonary Artery in Chest CT: Experience Using a Dual-Layer Detector Spectral CT

被引:39
|
作者
Bae, Kyungsoo [1 ,2 ]
Jeon, Kyung Nyeo [1 ,2 ]
Cho, Soo Buem [2 ]
Park, Sung Eun [2 ]
Moon, Jin Il [2 ]
Baek, Hye Jin [1 ,2 ]
Choi, Bo Hwa [2 ]
机构
[1] Gyeongsang Natl Univ, Sch Med, Inst Hlth Sci, Dept Radiol, 90 Chilam Dong, Jinju 660702, Gyeongsang Nam, South Korea
[2] Gyeongsang Natl Univ, Changwon Hosp, Dept Radiol, Chang Won, South Korea
关键词
chest CT; pulmonary artery; spectral CT; suboptimal contrast enhancement; virtual monoenergetic image; virtual monoenergetic imaging (VMI); COMPUTED-TOMOGRAPHY ANGIOGRAPHY; REDUCED IODINE LOAD; IMAGE-QUALITY; CONTRAST ENHANCEMENT; DIAGNOSTIC-ACCURACY; ENERGY; RECONSTRUCTION; OPTIMIZATION; REDUCTION; EMBOLISM;
D O I
10.2214/AJR.17.18537
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The objective of our study was to evaluate the quality of virtual monoenergetic imaging (VMI) from dual-layer detector spectral CT and the effect of virtual monoenergetic images obtained at low energies on the detection of pulmonary embolism (PE) in patients with a suboptimally enhanced pulmonary artery on chest CT. MATERIALS AND METHODS. Of 1552 consecutive chest CT examinations performed with dual-layer detector spectral CT using a routine protocol with a tube voltage of 120 kVp, 79 examinations with suboptimal enhancement of the pulmonary artery (i.e., mean attenuation of pulmonary artery <= 180 HU) were included. The mean attenuation of the pulmonary artery, noise, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) of virtual monoenergetic images obtained at 40-200 keV were compared with those of the conventional 120-kVp images. The virtual monoenergetic images with the best CNR were compared with the 120-kVp images with regard to subjective image quality and diagnostic accuracy for detecting PE. RESULTS. Sufficient attenuation of the pulmonary artery (> 180 HU) was obtained using VMI for 78 of the 79 examinations. The noise levels of the virtual monoenergetic images were gradually increased with decreasing energy level (i.e., kiloelectron volt setting). The CNR and SNR of virtual monoenergetic images at 40-65 keV were significantly higher (both, p < 0.001) than the CNR and SNR of the 120-kVp images. The CNR was the highest at 40 keV for all cases. Diagnostic accuracy for detecting PE was significantly higher for 40-keV images (reader 1: AUC = 0.992, p = 0.033; reader 2: AUC = 0.986, p = 0.043) than for 120-kVp images (reader 1, AUC = 0.911; reader 2, AUC = 0.933). The subjective quality was not different between these two images. CONCLUSION. In chest CT examinations in which the pulmonary artery is suboptimally enhanced, obtaining virtual monoenergetic images at a low energy setting using dual-layer detector spectral CT allows sufficient attenuation of the pulmonary artery to be achieved while preserving image quality and increasing diagnostic performance for detecting PE.
引用
收藏
页码:734 / 741
页数:8
相关论文
共 50 条
  • [1] Image quality evaluation of dual-layer spectral detector CT of the chest and comparison with conventional CT imaging
    Doerner, Jonas
    Hauger, Myriam
    Hickethier, Tilman
    Byrtus, Jonathan
    Wybranski, Christian
    Hokamp, Nils Grosse
    Maintz, David
    Haneder, Stefan
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2017, 93 : 52 - 58
  • [2] Comprehensive analyses with radiological and biological markers of breast cancer on contrast-enhanced chest CT: a single center experience using dual-layer spectral detector CT
    Moon, Jin Il
    Choi, Bo Hwa
    Baek, Hye Jin
    Ryu, Kyeong Hwa
    Park, Sung Eun
    Ha, Ji Young
    Jung, Eun Jung
    Lee, Han Shin
    An, Hyo Jung
    [J]. EUROPEAN RADIOLOGY, 2020, 30 (05) : 2782 - 2790
  • [3] Comprehensive analyses with radiological and biological markers of breast cancer on contrast-enhanced chest CT: a single center experience using dual-layer spectral detector CT
    Jin Il Moon
    Bo Hwa Choi
    Hye Jin Baek
    Kyeong Hwa Ryu
    Sung Eun Park
    Ji Young Ha
    Eun Jung Jung
    Han Shin Lee
    Hyo Jung An
    [J]. European Radiology, 2020, 30 : 2782 - 2790
  • [4] Iodine Imaging at Spectral CT with a Dual-Layer Detector
    Ozguner, O.
    Dhanantwari, A.
    Halliburton, S.
    Wen, G.
    Utrup, S.
    Jordan, D.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3657 - 3657
  • [5] Improved Coronary Artery Visualization Using Virtual Monoenergetic Imaging from Dual-Layer Spectral Detector CT Angiography
    D'Angelo, Tommaso
    Lanzafame, Ludovica R. M.
    Micari, Antonino
    Blandino, Alfredo
    Yel, Ibrahim
    Koch, Vitali
    Gruenewald, Leon D.
    Vogl, Thomas J.
    Booz, Christian
    Bucolo, Giuseppe M.
    Cannizzaro, Maria Teresa
    Ascenti, Giorgio
    Mazziotti, Silvio
    [J]. DIAGNOSTICS, 2023, 13 (16)
  • [6] Objective Image Characterization of Spectral CT with a Dual-Layer Detector
    Ozguner, O.
    Halliburton, S.
    Dhanantwari, A.
    Wen, G.
    Utrup, S.
    Jordan, D.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3836 - 3836
  • [7] Objective image characterization of a spectral CT scanner with dual-layer detector
    Ozguner, Orhan
    Dhanantwari, Amar
    Halliburton, Sandra
    Wen, Gezheng
    Utrup, Steven
    Jordan, David
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (02):
  • [8] The Application of Dual-layer Spectral Detector CT in Abdominal Vascular Imaging
    Lai, Lu-Yao
    Jiang, Ying
    Shu, Jian
    [J]. CURRENT MEDICAL IMAGING, 2023, 19 (14) : 1609 - 1615
  • [9] Noninvasive assessment of liver fibrosis by dual-layer spectral detector CT
    Morita, Koichiro
    Nishie, Akihiro
    Ushijima, Yasuhiro
    Takayama, Yukihisa
    Fujita, Nobuhiro
    Kubo, Yuichiro
    Ishimatsu, Keisuke
    Yoshizumi, Tomoharu
    Maehara, Junki
    Ishigami, Kousei
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2021, 136
  • [10] Improved Sensitivity and Reader Confidence in CT Colonography Using Dual-Layer Spectral CT: A Phantom Study
    Obmann, Markus M.
    An, Chansik
    Schaefer, Amanda
    Sun, Yuxin
    Wang, Zhen J.
    Yee, Judy
    Yeh, Benjamin M.
    [J]. RADIOLOGY, 2020, 297 (01) : 99 - 107