Effect of bearing size on the long-term wear, wear debris, and ion levels of large diameter metal-on-metal hip replacements -: An in vitro study

被引:64
|
作者
Leslie, Ian [1 ]
Williams, Sophie [1 ]
Brown, Chris [1 ]
Isaac, Graham [2 ]
Jin, Zhongmin [1 ]
Ingham, Eileen [1 ]
Fisher, John [1 ]
机构
[1] Univ Leeds, Inst Med & Biol Engn, Leeds, W Yorkshire, England
[2] DePuy Int Ltd, Leeds, W Yorkshire, England
基金
英国医学研究理事会; 英国工程与自然科学研究理事会;
关键词
hip surface replacement; wear; metal-on-metal; wear debris; ion levels;
D O I
10.1002/jbm.b.31087
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Hip resurfacing arthroplasty has become a popular alternative to conventional hip surgery. Surface replacements with bearing sizes of 55 mm (n = 5) and 39 rum (n = 5) were tested in a hip simulator for 15 million cycles (Me). Wear debris was isolated from the serum lubricant and characterized by field emmitting gun scanning electorn microscopy, and ion levels were measured via inductively coupled plasma mass spectroscopy at intervals throughout the test. The 39 mm bearings showed significantly greater bedding in volumetric wear (2.58 mm(3)) compared with the 55 mm bearings (1.15 mm(3)). There was no significant difference between the steady state wear rates (1-15 Me) between the two sizes (0.10 and 0.09 mm(3)/Mc, respectively); however, this parity only became clear after 7 Mc. The wear debris isolated was oval in morphology with a mean particle size of 28 nm and a range of 9-108 urn. The Co levels measured at 0.13 Mc were significantly greater than at 3.6 Mc for both bearing sizes (10926 ppb and 176 ppb, respectively). After 0.5 Mc, the Co levels from the 39 mm bearings were significantly higher than the 55 rum (11,007 vs. 1475 ppb). The wear results support previous findings showing that increasing the femoral head size decreased volumetric bedding in wear. The ion levels measured suggest both bearing sizes have similar initial wear rates; however, the 55 mm bearings reach steady state wear more rapidly. (c) 2008 Wiley Periodicals, Inc.
引用
收藏
页码:163 / 172
页数:10
相关论文
共 50 条
  • [21] Effect of wear, acetabular cup inclination angle, load and serum degradation on the friction of a large diameter metal-on-metal hip prosthesis
    Saikko, Vesa
    [J]. CLINICAL BIOMECHANICS, 2019, 63 : 1 - 9
  • [22] Long-term survival analysis of cementless large-diameter head metal-on-metal total hip arthroplasty
    Holappa, Elli
    Kettunen, Jukka
    Miettinen, Hannu
    Kroger, Heikki
    Miettinen, Simo
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (07) : 4437 - 4446
  • [23] Long-term survival analysis of cementless large-diameter head metal-on-metal total hip arthroplasty
    Elli Holappa
    Jukka Kettunen
    Hannu Miettinen
    Heikki Kröger
    Simo Miettinen
    [J]. Archives of Orthopaedic and Trauma Surgery, 2023, 143 : 4437 - 4446
  • [24] Effect of head size on wear properties of metal-on-metal bearings of hip prostheses, and comparison with wear properties of metal-on-polyethylene bearings using hip simulator
    Okazaki, Yoshimitsu
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 31 : 152 - 163
  • [25] Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement A CONSEQUENCE OF EXCESS WEAR
    Langton, D. J.
    Jameson, S. S.
    Joyce, T. J.
    Hallab, N. J.
    Natu, S.
    Nargol, A. V. F.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (01): : 38 - 46
  • [26] Does bearing size influence metal ion levels in large-head metal-on-metal total hip arthroplasty? A comparison of three total hip systems
    James Smith
    David Lee
    Kamal Bali
    Pam Railton
    David Kinniburgh
    Peter Faris
    Deborah Marshall
    Brian Burkart
    James Powell
    [J]. Journal of Orthopaedic Surgery and Research, 9
  • [27] Does bearing size influence metal ion levels in large-head metal-on-metal total hip arthroplasty? A comparison of three total hip systems
    Smith, James
    Lee, David
    Bali, Kamal
    Railton, Pam
    Kinniburgh, David
    Faris, Peter
    Marshall, Deborah
    Burkart, Brian
    Powell, James
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2014, 9
  • [28] Serum Metal Ion Levels in Cementless Metal-On-Metal Total Hip Arthroplasty: Long-Term Follow-Up Trends
    Kim, Chul-Ho
    Ryu, Jae Jung
    Jeong, Mi Yeon
    Kim, Ji Wan
    Chang, Jae Suk
    Yoon, Pil Whan
    [J]. JOURNAL OF ARTHROPLASTY, 2019, 34 (03): : 534 - 537
  • [29] 'Severe' wear challenge to 'as-cast' and 'double heat-treated' large-diameter metal-on-metal hip bearings
    Bowsher, JG
    Nevelos, J
    Williams, PA
    Shelton, JC
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 135 - 143
  • [30] Cushion bearings versus large diameter head metal-on-metal bearings in total hip arthroplasty: a short-term metal ion study
    Moroni, Antonio
    Nocco, Emanuele
    Hoque, Martha
    Diremigio, Elena
    Buffoli, Dario
    Cantu, Francesco
    Catalani, Simona
    Apostoli, Pietro
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2012, 132 (01) : 123 - 129