Material selection in the design of the femoral component of cemented total hip replacement

被引:29
|
作者
El-Sheikh, HF [1 ]
MacDonald, BJ [1 ]
Hashmi, MSJ [1 ]
机构
[1] Dublin City Univ, Sch Mech & Mfg Engn, Dublin 9, Ireland
关键词
artificial hip; prosthesis; finite clement method;
D O I
10.1016/S0924-0136(01)01128-1
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The finite clement stress analysis technique can be used to optimise both design and materials selection in many load-bearing components. Three-dimensional finite element models for a femur implanted with a cemented prosthesis were constructed with a representative physiological loading condition. The effect of prosthesis Young's modulus and cement Young's modulus on stresses in the prosthesis stem and cement layer was studied. A range of materials was considered for the stem and cement, enabling the derivation of two equations predicting the maximum principal stress in either material when any combination of materials is used. The materials selection criterion is proposed based on fatigue strength. (C) 2002 Published by Elsevier Science B.V.
引用
收藏
页码:309 / 317
页数:9
相关论文
共 50 条
  • [1] THE SURVIVAL OF THE CEMENTED FEMORAL COMPONENT OF A TOTAL HIP-REPLACEMENT
    RITTER, MA
    CAMPBELL, ED
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1989, (243) : 143 - 147
  • [2] BENDING AND FRACTURE OF THE FEMORAL COMPONENT IN CEMENTED TOTAL HIP-REPLACEMENT
    JONES, PR
    HUKINS, DWL
    PORTER, ML
    DAVIES, KE
    HARDINGE, K
    TAYLOR, CJ
    [J]. JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (01): : 9 - 15
  • [3] ASEPTIC LOOSENING OF THE FEMORAL COMPONENT IN CEMENTED TOTAL HIP-REPLACEMENT
    JONES, PR
    HUKINS, DWL
    PORTER, ML
    DAVIES, KE
    TIMPERLEY, AJ
    HARDINGE, K
    TAYLOR, CJ
    [J]. JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (05): : 379 - 384
  • [4] Cemented Calcar Replacement Femoral Component in Revision Hybrid Total Hip Arthroplasty
    Ito, Hiroshi
    Tanino, Hiromasa
    Yamanaka, Yasuhiro
    Nakamura, Toshiki
    Minami, Akio
    Matsuno, Takeo
    [J]. JOURNAL OF ARTHROPLASTY, 2011, 26 (02): : 236 - 243
  • [5] BENDING AND FRACTURE OF THE FEMORAL COMPONENT IN CEMENTED TOTAL HIP-REPLACEMENT - COMMENT
    PAUL, JP
    [J]. JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (05): : 444 - 444
  • [6] Femoral component failure in canine cemented total hip replacement: a report of two cases
    Yates, G. D.
    Wasik, S. M.
    Edwards, G. A.
    [J]. AUSTRALIAN VETERINARY JOURNAL, 2010, 88 (06) : 225 - 230
  • [7] Femoral stem wear in cemented total hip replacement
    Zhang, H-Y
    Blunt, L.
    Jiang, X-Q
    Brown, L.
    Barrans, S.
    Zhao, Y.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H5) : 583 - 592
  • [8] Revision total hip arthroplasty with cemented femoral component
    Nouri, Habib
    Kallel, Sofiene
    Slimane, Mohamed Hadj
    Meherzi, Mohamed Hedi
    Ouertatani, Moez
    Karray, Salahedine
    [J]. EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2008, 18 (05): : 327 - 332
  • [9] Revision total hip arthroplasty with cemented femoral component
    Habib Nouri
    Sofiene Kallel
    Mohamed Hadj Slimane
    Mohamed Hédi Meherzi
    Moez Ouertatani
    Salahedine Karray
    [J]. European Journal of Orthopaedic Surgery & Traumatology, 2008, 18 : 327 - 332
  • [10] Material selection for femoral component of total knee replacement using comprehensive VIKOR
    Bahraminasab, Marjan
    Jahan, Ali
    [J]. MATERIALS & DESIGN, 2011, 32 (8-9) : 4471 - 4477