Pretreatment Dose Verification in Volumetric Modulated Arc Therapy Using Liquid Ionization Chamber

被引:6
|
作者
Kakade, Nitin R. [1 ,2 ]
Kumar, Rajesh [1 ]
Sharma, Sunil Dutt [1 ,2 ]
Mittal, Vikram [3 ]
Datta, D. [1 ,2 ]
机构
[1] Bhabha Atom Res Ctr, Radiol Phys & Advisory Div, CT & CRS Bldg, Mumbai 400094, Maharashtra, India
[2] Homi Bhabha Natl Inst, Mumbai, Maharashtra, India
[3] PD Hinduja Natl Hosp, Dept Radiotherapy, Mumbai, Maharashtra, India
关键词
Collection efficiency; dose verification; liquid ion chamber; volumetric modulated arc therapy; GENERAL RECOMBINATION CORRECTION; COLLECTION EFFICIENCY; 2-DOSE-RATE METHOD; IMRT VERIFICATION; TETRAMETHYLSILANE; ISOOCTANE; FILM;
D O I
10.4103/jmp.JMP_108_18
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of the present study was to evaluate the practicability of liquid ionization chamber (LIC) for pretreatment dose verification of the advanced radiotherapy techniques such as volumetric modulated arc therapy (VMAT). Materials and Methods: The dosimetric characteristics of LIC such as repeatability, sensitivity, monitor unit linearity, dose rate dependence, angular dependence, voltage-current response, and output factors were investigated in 6 MV therapeutic X-ray beams. The LIC was cross-calibrated against 0.125-cc air-filled thimble ionization chamber. A dedicated dosimetry insert made up of Perspex to incorporate the LIC at proper location in the intensity-modulated radiation therapy thorax phantom was locally fabricated. The collection efficiency and ion recombination correction factor was determined using the two-dose rate method. Pretreatment dose verification measurement of VMAT treatment plans were carried out using the liquid ionization chamber as well as small volume (0.125 cc) air-filled thimble ionization chamber. The measured dose values by the two dosimeters and TPS calculated dose at a given point were compared. Results: The relative percentage differences between the TPS calculated and measured doses were within +/- 1.57% for LIC and +/- 2.21% for 0.125 cc ionization chamber, respectively. Conclusions: The measured dose values by the two dosimeters and TPS calculated dose at a given point were found comparable suggesting that the LIC could be a good choice of dosimeter for pretreatment dose verification in VMAT.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 50 条
  • [1] A Prediction Model for Pretreatment Dose Verification of Volumetric-Modulated Arc Therapy (VMAT) Using Machine Learning
    Li, G.
    Yuan, Q.
    Bai, S.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E612 - E612
  • [2] Dose verification for respiratory-gated volumetric modulated arc therapy
    Qian, Jianguo
    Xing, Lei
    Liu, Wu
    Luxton, Gary
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (15): : 4827 - 4838
  • [4] DOSIMETRIC VERIFICATION OF RAPIDARC™ VOLUMETRIC MODULATED ARC THERAPY
    Medin, J.
    Kjaer-Kristoffersen, F.
    Ceberg, S.
    Back, S.
    Gustavsson, H.
    Archambault, Y.
    Bocanek, J.
    Cozzi, L.
    Eberhard, V.
    Gomola, I.
    Wiberg, I.
    Korreman, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S47 - S48
  • [5] Three-dimensional dose reconstruction-based pretreatment dosimetric verification in volumetric modulated arc therapy for prostate cancer
    Jeong, Yuri
    Oh, Jeong Geun
    Kang, Jeong Ku
    Moon, Sun Rock
    Lee, Kang Kyoo
    [J]. RADIATION ONCOLOGY JOURNAL, 2020, 38 (01): : 60 - 67
  • [6] Monte Carlo Dose Verification for Volumetric Modulated Arc Therapy in Multiple Brain Metastases
    Tominaga, H.
    Araki, F.
    Kanetake, N.
    Sakata, J.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3546 - 3547
  • [7] External Beam Pulsed Low Dose Radiotherapy Using Volumetric Modulated Arc Therapy: Planning, Delivery and Verification
    Tyagi, N.
    Yang, K.
    Sandhu, R.
    Yan, D.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3997 - 3997
  • [8] Verification of Volumetric Modulated Arc Therapy Plans with Independent Three Dimensional Dose Computation Algorithm
    Kathirvel, M.
    Subramanian, V. Shanmuga
    Arun, G.
    Swamy, S. Thirumalai
    Ramalingam, K.
    Kala, S.
    Kumar, S. Ashok
    Anantharaman, A.
    Karthikeyan, K.
    Kumar, N. Jagdish
    Arivarasan, I.
    Babaiah, M.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [9] Dose verification of volumetric modulated arc therapy (VMAT) by use of in-treatment linac parameters
    Haga A.
    Sakumi A.
    Okano Y.
    Itoh S.
    Saotome N.
    Kida S.
    Igaki H.
    Shiraishi K.
    Yamashita H.
    Ohtomo K.
    Nakagawa K.
    [J]. Radiological Physics and Technology, 2013, 6 (2) : 335 - 342
  • [10] Verification of dose distribution for volumetric modulated arc therapy total marrow irradiation in a humanlike phantom
    Surucu, Murat
    Yeginer, Mete
    Kavak, Gulbin O.
    Fan, John
    Radosevich, James A.
    Aydogan, Bulent
    [J]. MEDICAL PHYSICS, 2012, 39 (01) : 281 - 288