Patient Radiation Dose Reduction during Transarterial Chemoembolization Using a Novel X-Ray Imaging Platform

被引:5
|
作者
Kohlbrenner, Ryan [1 ]
Kolli, Pallav [1 ]
Taylor, Andrew G. [1 ]
Kohi, Maureen P. [1 ]
Fidelman, Nicholas [1 ]
LaBerge, Jeanne M. [1 ]
Kerlan, Robert K. [1 ]
Agarwal, Vishal K. [1 ]
Lehrman, Evan D. [1 ]
Nanavati, Sujal [1 ]
Avrin, David E. [1 ]
Gould, Robert [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94143 USA
关键词
INTERVENTIONAL RADIOLOGY PROCEDURES; DIGITAL-SUBTRACTION-ANGIOGRAPHY; AREA PRODUCT; FLUOROSCOPY; TECHNOLOGY; QUALITY; TIME; IR;
D O I
10.1016/j.jvir.2015.06.016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate radiation dose reduction in patients undergoing transarterial chemoembolization with the use of a new image acquisition and processing platform. Materials and 'Methods: Radiation-dose data were obtained from 176 consecutive chemoembolization procedures in 135 patients performed in a single angiography suite. From January 2013 through October 2013, 85 procedures were, performed by using our institution's standard fluoroscopic settings. After upgrading the x-ray fluoroscopy system with an image acquisition and processing platform designed to reduce image noise and reduce skin entrance dose, 91 chemoembolization procedures were performed from November 2013 through December 2014. Cumulative dose area product (CDAP), cumulative air kerma (CAK), and total fluoroscopy time were recorded for each procedure. Image quality was assessed by three interventional radiologists blinded to the x-ray acquisition platform used. Results: Patient radiation dose indicators were significantly lower for chemoembolization procedures perfauned with the novel imaging platform. Mean CDAP decreased from 3,033.2 dGy.cm(2) (range, 600.3-9,404.1 dGy.cm(2)) to 1,640.1 dGy.cm(2) (range, 278.6-6,779.9 dGy.cm(2); 45.9% reduction; P<.00001). Mean CAK 'decreased from 1,445.4 mGy (range, 303.6-5,233.7 mGy) to 971.7 mGy (range, 144.2-3,512.0 mGy; 32.8% reduction; P<.0001). A 20.3% increase in mean total fluoroscopy time was noted after upgrading the imaging platform, but blinded analysis of the image quality revealed no significant degradation. Conclusions: Although a small increase in fluoroscopy time was observed, a significant reduction in patient radiation dose was achieved by using the optimized imaging platform, without image quality degradation.
引用
收藏
页码:1331 / 1338
页数:8
相关论文
共 50 条
  • [31] A Preliminary Study on Low-dose X-ray Imaging Using Parametric X-ray
    Sakae, Toshiro
    Hayakawa, Yasushi
    Takahashi, Yumiko
    Kuwada, Takao
    Nakao, Keisuke
    Nogami, Kyoko
    Inagaki, Manabu
    Tanaka, Toshinari
    Sato, Isamu
    Hirayama, Katsunori
    Sekiya, Kotarou
    Kaneda, Takashi
    Suemitsu, Masaaki
    Okada, Hiroyuki
    Yamamoto, Hirotsugu
    Hata, Saya
    Akimoto, Yoshiaki
    Fukumoto, Masahiko
    Makimura, Masaharu
    JOURNAL OF HARD TISSUE BIOLOGY, 2011, 20 (01) : 31 - 35
  • [32] Novel X-ray image noise reduction technology reduces patient radiation dose while maintaining image quality in coronary angiography
    ten Cate, T.
    van Wely, M.
    Gehlmann, H.
    Mauti, M.
    Camaro, C.
    Reifart, N.
    Suryapranata, H.
    de Boer, M. J.
    NETHERLANDS HEART JOURNAL, 2015, 23 (11) : 525 - 530
  • [33] Novel X-ray image noise reduction technology reduces patient radiation dose while maintaining image quality in coronary angiography
    T. ten Cate
    M. van Wely
    H. Gehlmann
    M. Mauti
    C. Camaro
    N. Reifart
    H. Suryapranata
    M.J. de Boer
    Netherlands Heart Journal, 2015, 23 : 525 - 530
  • [34] Impact of digital imaging on radiation doses to the patient during X-ray examination of the urinary tract
    Sjöholm, B
    Geijer, H
    Persliden, J
    ACTA RADIOLOGICA, 2005, 46 (06) : 657 - 661
  • [35] Absorbed Dose to the Conceptus and Patient Size in X-Ray Projection Imaging
    Magill, D.
    Huda, W.
    He, W.
    Tipnis, S.
    Mah, E.
    Frey, G.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [36] Novel X-ray imaging technology enables significant patient dose reduction in interventional cardiology while maintaining diagnostic image quality
    Eloot, Liesbeth
    Thierens, Hubert
    Taeymans, Yves
    Drieghe, Benny
    De Pooter, Jan
    Van Peteghem, Sylvie
    Buytaert, Dimitri
    Gijs, Thomas
    Lapere, Regine
    Bacher, Klaus
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2015, 86 (05) : E205 - E212
  • [37] EVALUATION OF DOSE REDUCTION POTENTIALS OF A NOVEL SCATTER CORRECTION SOFTWARE FOR BEDSIDE CHEST X-RAY IMAGING
    Renger, Bernhard
    Brieskorn, Carina
    Toth, Vivien
    Mentrup, Detlef
    Jockel, Sascha
    Lohoefer, Fabian
    Schwarz, Martin
    Rummeny, Ernst J.
    Noel, Peter B.
    RADIATION PROTECTION DOSIMETRY, 2016, 169 (1-4) : 60 - 67
  • [38] SPOT REGION OF INTEREST IMAGING: A NOVEL FUNCTIONALITY AIMED AT X-RAY DOSE REDUCTION IN NEUROINTERVENTIONAL PROCEDURES
    Borota, Ljubisa
    Patz, Andreas
    RADIATION PROTECTION DOSIMETRY, 2020, 188 (03) : 322 - 331
  • [39] A study on the reduction of radiation dose using carbon fiber-reinforced plastic during X-ray examination of children and infants
    Kwak, Y. -G.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (01)
  • [40] Significant reduction of radiation exposure during electrophysiological procedures with the use of novel x-ray system using a crystalline detector
    Knecht, S.
    Pavlovic, N.
    Reichlin, T.
    Kuhne, M.
    Schar, B.
    Osswald, S.
    Sticherling, C.