Two-Port Static Coated Micro Gas Chromatography Column With an Embedded Thermal Conductivity Detector

被引:39
|
作者
Shree Narayanan [1 ]
Alfeeli, Bassam [1 ,2 ]
Agah, Masoud [1 ]
机构
[1] Virginia Polytech Inst & State Univ, Bradley Dept Elect & Comp Engn, VTMEMS Lab, Blacksburg, VA 24061 USA
[2] Kuwait Inst Sci Res, Kuwait 13109, Kuwait
基金
美国国家科学基金会;
关键词
Micro gas chromatography; monolithic microelectromechanical systems integration; separation column; thermal conductivity detector; FLAME IONIZATION DETECTOR; HIGH-PERFORMANCE; FABRICATION; ARRAY; GC;
D O I
10.1109/JSEN.2011.2180017
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper reports a microfabricated 2 x 4 cm gas chromatography chip to separate and detect gases in a two-port structure by embedding a microthermal conductivity detector (mu TCD) within a separation column. A circular on-chip heater is placed on the backside of the monolithic device enabling temperature programming and consequently faster analysis of the heavier components. A unique process enhanced by reactive ion etching lag (RIE lag) is used to achieve multiple etch depths in silicon and restrict the process flow to just three masks. The silicon substrate contains the separation column, the heater, and the tunnels for the TCD electric feed throughs. A Pyrex wafer containing the TCD elements is anodically bonded to the silicon substrate to seal the structure. Performance of a standalone mu TCD fabricated in the same process and integrated in a hybrid fashion is also described. The single-chip design demonstrates successful separation and identification of multicomponent gas mixtures with a performance comparable to that obtained through a flame ionization detector connected in series. Further, on-chip temperature programming capability was utilized to elevate the column temperature to 75 degrees C to exhibit analysis in less than a minute.
引用
收藏
页码:1893 / 1900
页数:8
相关论文
共 50 条
  • [31] Micro gas chromatography column coated with ionic liquid/metal-organic framework composites as the stationary phase
    Zhang, Kangning
    Wu, Xinyu
    Wang, Dazuo
    Hua, Xinyuan
    Zhang, Ze
    Li, Peng
    Tian, Bing
    Wang, Hairong
    Sun, Jianhai
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2023, 33 (11)
  • [32] DETERMINATION OF THE NITROGEN CONCENTRATION IN ARGON ON PPM LEVEL BY GAS-CHROMATOGRAPHY WITH A THERMAL-CONDUCTIVITY DETECTOR
    LASA, J
    DROZDOWICZ, B
    KORUS, A
    CHEMIA ANALITYCZNA, 1994, 39 (06): : 735 - 739
  • [33] New Analytical Method for Measuring the Atomic Weight of Neon Using Gas Chromatography with a Thermal Conductivity Detector
    Kim, Jeong Eun
    Yang, Inseok
    Moon, Dong Min
    Kim, Jin Seog
    Hong, Kiryong
    ACS OMEGA, 2021, 6 (22): : 14476 - 14480
  • [34] Thermal Conductivity Detector for Gas Chromatography: Very Wide Gain Range Acquisition System and Experimental Measurements
    Rastrello, Fabio
    Placidi, Pisana
    Scorzoni, Andrea
    Cozzani, Enrico
    Messina, Marco
    Elmi, Ivan
    Zampolli, Stefano
    Cardinali, Gian Carlo
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2013, 62 (05) : 974 - 981
  • [35] USE OF A SMALL VOLUME THERMAL-CONDUCTIVITY DETECTOR FOR CAPILLARY GAS-CHROMATOGRAPHY - PRACTICAL ASPECTS
    SIMON, R
    WELLS, G
    ACS SYMPOSIUM SERIES, 1984, 250 : 59 - 76
  • [36] DETERMINATION OF CARBON-MONOXIDE IN BLOOD BY GAS-CHROMATOGRAPHY USING A THERMAL-CONDUCTIVITY DETECTOR
    GOLDBAUM, LR
    CHACE, DH
    LAPPAS, NT
    JOURNAL OF FORENSIC SCIENCES, 1986, 31 (01) : 133 - 142
  • [37] Flow-insensitive micro-thermal conductivity detector with semi-diffusion gas channel
    Wang, Jiayu
    Zhou, Hong
    Wang, Yi
    Wu, Ming
    Liu, Yanxiang
    Huang, Gang
    Li, Tie
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2022, 32 (09)
  • [38] Research of NMHCs Measurement by Gas Chromatography Using Two-valve, Two-column and Single FID Detector
    Zhao, Z. Z.
    An, Z. H.
    2015 2ND INTERNATIONAL CONFERENCE ON INTELLIGENT MATERIALS AND MECHATRONICS (IMM 2015), 2015, : 129 - 132
  • [39] Differences in Product Distribution Measured with Flame Ionization Detector Gas Chromatography and Thermal Conductivity Detector Gas Chromatography during the Dimethyl Ether-to-Olefins and Methanol-to-Olefins Processes
    Gao, Yuli
    Cao, Yingqian
    Chen, Sheng-Li
    Wang, Ya
    Zhu, Ruyue
    Sun, Wei
    Zhang, Qi
    Fan, Yu
    ENERGY & FUELS, 2017, 31 (12) : 13266 - 13272
  • [40] PRACTICAL ASPECTS REGARDING THE USE OF A SMALL VOLUME THERMAL-CONDUCTIVITY DETECTOR FOR CAPILLARY GAS-CHROMATOGRAPHY
    SIMON, R
    WELLS, G
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1983, 185 (MAR): : 105 - ANYL