Feedback system of a liquid-nitrogen-cooled double-crystal monochromator: design and performances

被引:133
|
作者
Proux, O
Nassif, V
Prat, A
Ulrich, O
Lahera, E
Biquard, X
Menthonnex, JJ
Hazemann, JL
机构
[1] Univ Grenoble 1, Lab Geophys Interne & Tectonophys, CNRS, UMR, F-38400 St Martin Dheres, France
[2] CEA Grenoble, DRFMC, SP2M, NRS, F-38054 Grenoble, France
[3] CNRS, Cristallog Lab, F-38042 Grenoble, France
关键词
double-crystal monochromator; cryogenic cooling; parallelism adjustment; X-ray absorption spectroscopy;
D O I
10.1107/S0909049505037441
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new set-up is reported of an indirect cryogenic cooling system for a double-crystal monochromator which runs on the BM30b/FAME beamline at the ESRF (Grenoble, France). This device has been conceived to limit the vibrations on the first diffracting crystal and to maintain it at a constant temperature. These points are crucial for maximizing the beamline stability. Moreover, the relative angular position of the second crystal can be dynamically adjusted by a piezoelectric transducer coupled with a feedback system in order to always be at the maximum of the rocking curve during an X-ray absorption spectroscopy scan. The temperature is stabilized to an accuracy of 0.01 degrees, with two principal consequences. The energy resolution is close to the theoretical value [Delta E/E=5.6x10(-6) for Si(220)] and the precision of the energy positioning is extremely good even if the power load changes. A feedback mechanism allows a permanent and automatic optimization of the angle between the two crystals of the monochromator. The intensity of the monochromatic beam remains optimized (i) when the intensity of the electron beam decreases in the storage ring and (ii) during an energy scan.
引用
收藏
页码:59 / 68
页数:10
相关论文
共 50 条
  • [1] Fixed exit double-crystal monochromator with one single cooled crystal
    Souvorov, A
    Drakopoulos, M
    Freund, A
    Snigireva, I
    Snigirev, A
    Erko, A
    Gudat, W
    Abrosimov, N
    Rossolenko, S
    Schroder, W
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 413 (2-3): : 427 - 430
  • [2] A CHROMATIC PREMIRROR SYSTEM FOR A DOUBLE-CRYSTAL MONOCHROMATOR
    VANDERLAAN, G
    PADMORE, HA
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 291 (1-2): : 225 - 227
  • [3] Neutron double-crystal monochromator
    Mikerov, VI
    Vedernikov, VI
    Soldatova, EN
    Tukarev, VA
    [J]. INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1996, 39 (01) : 13 - 17
  • [4] TECHNOLOGY FOR LIQUID-NITROGEN-COOLED COMPUTERS
    LONGSWORTH, RC
    STEYERT, WA
    [J]. IEEE TRANSACTIONS ON ELECTRON DEVICES, 1987, 34 (01) : 4 - 7
  • [5] PERFORMANCE OF A LIQUID-NITROGEN-COOLED, THIN SILICON CRYSTAL MONOCHROMATOR ON A HIGH-POWER, FOCUSED WIGGLER SYNCHROTRON BEAM
    ROGERS, CS
    MILLS, DM
    LEE, WK
    KNAPP, GS
    HOLMBERG, J
    FREUND, A
    WULFF, M
    ROSSAT, M
    HANFLAND, M
    YAMAOKA, H
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (06): : 3494 - 3499
  • [6] A DOUBLE-CRYSTAL MONOCHROMATOR FOR EXAFS AT BSRL
    XIE, YN
    YANG, H
    SHAN, Q
    XIA, SJ
    XIAN, D
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01): : 956 - 957
  • [7] THE ETA10 LIQUID-NITROGEN-COOLED SUPERCOMPUTER SYSTEM
    CARLSON, DM
    SULLIVAN, DC
    BACH, RE
    RESNICK, DR
    [J]. IEEE TRANSACTIONS ON ELECTRON DEVICES, 1989, 36 (08) : 1404 - 1413
  • [8] USE OF SPUTTERED MULTILAYERS IN A DOUBLE-CRYSTAL MONOCHROMATOR
    COWAN, PL
    LINDLE, DW
    SOUTHWORTH, SH
    KARLIN, BA
    PERERA, RCC
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 291 (1-2): : 219 - 220
  • [9] DESIGN AND PERFORMANCE OF THE ADVANCED LIGHT-SOURCE DOUBLE-CRYSTAL MONOCHROMATOR
    JONES, G
    RYCE, S
    LINDLE, DW
    KARLIN, BA
    WOICIK, JC
    PERERA, RCC
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02): : 1748 - 1750
  • [10] A double-crystal monochromator for neutron stress diffractometry
    V. T. Em
    A. M. Balagurov
    V. P. Glazkov
    I. D. Karpov
    P. Mikula
    N. F. Miron
    V. A. Somenkov
    V. V. Sumin
    J. Šaroun
    M. N. Shushunov
    [J]. Instruments and Experimental Techniques, 2017, 60 : 526 - 532