Validation of the Tensoval Duo Control II blood pressure monitor for clinic use and self-measurement according to the British Hypertension Society protocol and the European Society of Hypertension International Protocol revision 2010

被引:6
|
作者
de Greeff, Annemarie [1 ]
Shennan, Andrew H. [1 ]
机构
[1] Kings Coll London, Kings Hlth Partners, Womens Hlth Acad Ctr, London SE1 7EH, England
关键词
accuracy; adult; auscultatory; blood pressure; British Hypertension Society; European Society of Hypertension; oscillometric; validation; ACCURACY ASSESSMENT; MEASURING DEVICES; ADULTS;
D O I
10.1097/MBP.0b013e328360fb52
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The Tensoval Duo Control II is an automated upper arm device that uses a combination of oscillometric and auscultatory technology to determine blood pressure noninvasively. The accuracy of this device was assessed according to the British Hypertension Society (BHS) protocol and the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) in an adult population. Methods Ethical approval was obtained. Eighty-five and 33 adult individuals, respectively, were recruited to fulfil the requirements of each protocol. Trained observers took nine sequential same-arm measurements alternating between a mercury sphygmomanometer and the device. The device had to achieve at least a B grade for both systolic and diastolic pressures to pass the BHS protocol and had to fulfil the criteria of all three phases of the ESH-IP2 protocol to receive recommendation. Results The device achieved an A/A grading for the BHS protocol and passed all three phases of the ESH-IP2 protocol. The mean difference +/- SD for the BHS/ESH protocols, respectively, was -1.8 +/- 6.5/ -0.7 +/- 5.7 mmHg for systolic pressure and 1.9 +/- 5.1/2.4 +/- 4.5mmHg for diastolic pressure. The device maintained its A/A grading throughout the low-pressure, medium-pressure and high-pressure ranges. Conclusion The Tensoval Duo Control II device is recommended for clinical and home use according to both the BHS and the ESH-IP2 standard. (c) 2013 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:161 / 166
页数:6
相关论文
共 50 条
  • [21] Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010
    M Abou-Dakn
    C Döhmen
    S Wenzel
    Journal of Human Hypertension, 2017, 31 : 89 - 92
  • [22] Erratum to: Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beate Beime
    Cornelia Deutsch
    Ralf Krüger
    Andreas Wolf
    Peter Müller
    Gertrud Hammel
    Peter Bramlage
    European Journal of Pediatrics, 2017, 176 : 835 - 836
  • [23] Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010
    Abou-Dakn, M.
    Doehmen, C.
    Wenzel, S.
    JOURNAL OF HUMAN HYPERTENSION, 2017, 31 (02) : 89 - 92
  • [24] The QardioArm Blood Pressure App for Self-Measurement in an Obese Population: Validation Study Using the European Society of Hypertension International Protocol Revision 2010
    Mazoteras-Pardo, Victoria
    Becerro-De-Bengoa-Vallejo, Ricardo
    Elena Losa-Iglesias, Marta
    Lopez-Lopez, Daniel
    Palomo-Lopez, Patricia
    Rodriguez-Sanz, David
    Calvo-Lobo, Cesar
    JMIR MHEALTH AND UHEALTH, 2018, 6 (10):
  • [25] Validation of the blood pressure measurement device Erkameter 125 PRO according to the European Society of Hypertension International Protocol revision 2010
    Beime, Beate
    Deutsch, Cornelia
    Krueger, Ralf
    Zimmermann, Erik
    Bramlage, Peter
    BLOOD PRESSURE MONITORING, 2016, 21 (06) : 356 - 360
  • [26] Validation of the blood pressure measurement device Beurer BM 28 according to the European Society of Hypertension International Protocol revision 2010
    Deutsch, Cornelia
    Bramlage, Carsten
    Botta, Beate
    Krueger, Ralf
    Forstner, Klaus
    Bramlage, Peter
    Beime, Beate
    BLOOD PRESSURE MONITORING, 2021, 26 (04) : 292 - 298
  • [27] Validation of the Pangao PG-800B5 for clinical use and self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Chen, Wan
    Zeng, Zhaolin
    Li, Lizhi
    Wan, Xiaofen
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2014, 19 (05) : 302 - 305
  • [28] Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Bilo, Grzegorz
    Zorzi, Cristina
    Munera, Juan E. Ochoa
    Torlasco, Camilla
    Giuli, Valentina
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2015, 20 (05) : 291 - 294
  • [29] Validation of the Omron HEM-7201 upper arm blood pressure monitor, for self-measurement in a high-altitude environment, according to the European Society of Hypertension International Protocol revision 2010
    K Cho
    M Tian
    Y Lan
    X Zhao
    L L Yan
    Journal of Human Hypertension, 2013, 27 : 487 - 491
  • [30] Validation of the PHILIPS DL8760 upper arm blood pressure monitor, in oscillometry mode, for self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010
    Chahine, Mirna N.
    Bou Harb, Souad
    Saad, Abdel Rahman
    Sarkis, Patrick
    Azaki, Alaa
    Harb, Aya
    Allouch, Ali
    Asmar, Roland
    BLOOD PRESSURE MONITORING, 2021, 26 (03) : 237 - 241