Validation of the Tensoval Duo Control II blood pressure monitor for clinic use and self-measurement according to the British Hypertension Society protocol and the European Society of Hypertension International Protocol revision 2010

被引:6
|
作者
de Greeff, Annemarie [1 ]
Shennan, Andrew H. [1 ]
机构
[1] Kings Coll London, Kings Hlth Partners, Womens Hlth Acad Ctr, London SE1 7EH, England
关键词
accuracy; adult; auscultatory; blood pressure; British Hypertension Society; European Society of Hypertension; oscillometric; validation; ACCURACY ASSESSMENT; MEASURING DEVICES; ADULTS;
D O I
10.1097/MBP.0b013e328360fb52
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The Tensoval Duo Control II is an automated upper arm device that uses a combination of oscillometric and auscultatory technology to determine blood pressure noninvasively. The accuracy of this device was assessed according to the British Hypertension Society (BHS) protocol and the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) in an adult population. Methods Ethical approval was obtained. Eighty-five and 33 adult individuals, respectively, were recruited to fulfil the requirements of each protocol. Trained observers took nine sequential same-arm measurements alternating between a mercury sphygmomanometer and the device. The device had to achieve at least a B grade for both systolic and diastolic pressures to pass the BHS protocol and had to fulfil the criteria of all three phases of the ESH-IP2 protocol to receive recommendation. Results The device achieved an A/A grading for the BHS protocol and passed all three phases of the ESH-IP2 protocol. The mean difference +/- SD for the BHS/ESH protocols, respectively, was -1.8 +/- 6.5/ -0.7 +/- 5.7 mmHg for systolic pressure and 1.9 +/- 5.1/2.4 +/- 4.5mmHg for diastolic pressure. The device maintained its A/A grading throughout the low-pressure, medium-pressure and high-pressure ranges. Conclusion The Tensoval Duo Control II device is recommended for clinical and home use according to both the BHS and the ESH-IP2 standard. (c) 2013 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:161 / 166
页数:6
相关论文
共 50 条
  • [1] Validation of Transtek blood pressure monitor TMB-1491 for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Tian, Huiyong
    Zeng, Sijian
    Zhong, Xiaoyan
    Gong, Wei
    Liu, Wenjun
    BLOOD PRESSURE MONITORING, 2015, 20 (05) : 280 - 282
  • [2] Validation of the iHealth BP7 wrist blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Wang, Qing
    Zhao, Huadong
    Chen, Wan
    Li, Ni
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2014, 19 (01) : 54 - 57
  • [3] Validation of the Medisana MTP Plus upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Erdem, Emre
    Aydogdu, Turkan
    Akpolat, Tekin
    BLOOD PRESSURE MONITORING, 2011, 16 (01) : 43 - 46
  • [4] Validation of the Globalcare GCE603 automated blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Song, Cheng
    Yu, Yang
    Lu, Bao-Chuan
    Yan, Xi-Ling
    BLOOD PRESSURE MONITORING, 2020, 25 (05) : 291 - 294
  • [5] Validation of the iHealth BP3 upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Chen, Can
    Shang, Fujun
    Wang, Jiepin
    Chen, Jianghong
    Ji, Na
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2012, 17 (06) : 253 - 256
  • [6] Validation of the YuWell YE690A upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Chen, Qi
    Lei, Lei
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2017, 22 (05) : 295 - 297
  • [7] Validation of the Andon KD-5917 automatic upper arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Guo, Wan-Gang
    Li, Bing-Ling
    He, Yong
    Xue, Yu-Sheng
    Wang, Hai-Yan
    Zheng, Qiang-Sun
    Xiang, Ding-Cheng
    BLOOD PRESSURE MONITORING, 2014, 19 (04) : 242 - 245
  • [8] Validation in the General Population of the iHealth Track Blood Pressure Monitor for Self-Measurement According to the European Society of Hypertension International Protocol Revision 2010: Descriptive Investigation
    Mazoteras-Pardo, Victoria
    Becerro-De-Bengoa-Vallejo, Ricardo
    Elena Losa-Iglesias, Marta
    Lopez-Lopez, Daniel
    Palomo-Lopez, Patricia
    Rodriguez-Sanzl, David
    Calvo-Lobo, Cesar
    JMIR MHEALTH AND UHEALTH, 2019, 7 (03):
  • [9] Validation of the Andon KD-5851 upper arm blood pressure monitor, for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Wu, Lili
    Jiao, Yinghui
    Wang, Chengdong
    Chen, Lei
    Di, Dalin
    Zhang, Haiyan
    BLOOD PRESSURE MONITORING, 2015, 20 (04) : 228 - 231
  • [10] Validation of the Medipro MediCare 100f upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Yi, Jun
    Wan, Yi
    Pan, Feng
    Yu, Xiaorong
    Zhao, Huadong
    Shang, Fujun
    Xu, Yongyong
    BLOOD PRESSURE MONITORING, 2011, 16 (04) : 197 - 199