Validation of the Tensoval Duo Control II blood pressure monitor for clinic use and self-measurement according to the British Hypertension Society protocol and the European Society of Hypertension International Protocol revision 2010
被引:6
|作者:
de Greeff, Annemarie
[1
]
Shennan, Andrew H.
[1
]
机构:
[1] Kings Coll London, Kings Hlth Partners, Womens Hlth Acad Ctr, London SE1 7EH, England
关键词:
accuracy;
adult;
auscultatory;
blood pressure;
British Hypertension Society;
European Society of Hypertension;
oscillometric;
validation;
ACCURACY ASSESSMENT;
MEASURING DEVICES;
ADULTS;
D O I:
10.1097/MBP.0b013e328360fb52
中图分类号:
R6 [外科学];
学科分类号:
1002 ;
100210 ;
摘要:
Objective The Tensoval Duo Control II is an automated upper arm device that uses a combination of oscillometric and auscultatory technology to determine blood pressure noninvasively. The accuracy of this device was assessed according to the British Hypertension Society (BHS) protocol and the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) in an adult population. Methods Ethical approval was obtained. Eighty-five and 33 adult individuals, respectively, were recruited to fulfil the requirements of each protocol. Trained observers took nine sequential same-arm measurements alternating between a mercury sphygmomanometer and the device. The device had to achieve at least a B grade for both systolic and diastolic pressures to pass the BHS protocol and had to fulfil the criteria of all three phases of the ESH-IP2 protocol to receive recommendation. Results The device achieved an A/A grading for the BHS protocol and passed all three phases of the ESH-IP2 protocol. The mean difference +/- SD for the BHS/ESH protocols, respectively, was -1.8 +/- 6.5/ -0.7 +/- 5.7 mmHg for systolic pressure and 1.9 +/- 5.1/2.4 +/- 4.5mmHg for diastolic pressure. The device maintained its A/A grading throughout the low-pressure, medium-pressure and high-pressure ranges. Conclusion The Tensoval Duo Control II device is recommended for clinical and home use according to both the BHS and the ESH-IP2 standard. (c) 2013 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:161 / 166
页数:6
相关论文