Design and Benchtop Validation of a Handheld Integrated Dynamic Breast Imaging System for Noninvasive Characterization of Suspicious Breast Lesions

被引:9
|
作者
Xu, Ronald X. [1 ,3 ]
Ewing, Joseph [2 ]
El-Dahdah, Hamid [1 ]
Wang, Bei [3 ]
Povoski, Stephen P. [4 ,5 ,6 ]
机构
[1] Ohio State Univ, Dept Biomed Engn, Columbus, OH 43210 USA
[2] Ohio State Univ, Dept Mech Engn, Columbus, OH 43210 USA
[3] Ohio State Univ, Dept Elect & Comp Engn, Columbus, OH 43210 USA
[4] Ohio State Univ, Div Surg Oncol, Arthur G James Canc Hosp, Columbus, OH 43210 USA
[5] Ohio State Univ, Richard J Solove Res Inst, Columbus, OH 43210 USA
[6] Ohio State Univ, Ctr Comprehens Canc, Columbus, OH 43210 USA
关键词
Near infrared imaging and spectroscopy; Ultrasound; Breast cancer; Dynamic stimulation;
D O I
10.1177/153303460800700609
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
We have developed a portable, handheld, integrated, dynamic breast imaging system that integrates a near infrared tissue oximeter, clinical ultrasound, and two pressure sensors for noninvasive detection of pressure-induced structural and functional dynamics of Suspicious breast lesions. A series of benchtop tests were conducted to validate multiple performance characteristics of the integrated dynamic near infrared/ultrasound breast imaging system (idNIRUS), including the reconstruction of the absorptive heterogeneities and the generation of the dynamic compression stimuli. In absorptive heterogeneity testing, we reconstructed the absorption coefficients of transparent polypropylene tubing circulated with a skim milk-India ink mixture and embedded in a gel wax tissue simulating phantom. High linear correlations (R-2 greater than 0.989) were observed between the reconstructed and the measured absorption coefficients of the embedded tubing. In dynamic compression testing, five volunteer operators generated ten successive compression sessions by compressing the idNIRUS imager on a breast self examination wearable model following the computer simulated pressure profile. The manually generated pressure profiles demonstrated an accuracy of 95.7% and operator-dependent variation of less than 5%. The results of the current benchtop tests will help to optimize the most appropriate testing conditions for our future planned clinical trial.
引用
收藏
页码:471 / 481
页数:11
相关论文
共 50 条
  • [1] Application of Breast Scintigraphy for Patients with Suspicious (Breast Imaging-Reporting and Data System IV) Breast Lesions
    Azarpeikan, Ali Reza
    Omranipour, Ramesh
    Mahmoodzadeh, Habibollah
    Miri, Seyed Rouhollah
    Mohammadzadeh, Narjes
    Derakhshan, Farhang
    Farzanefar, Saeed
    Abbasi, Mehrshad
    ADVANCED BIOMEDICAL RESEARCH, 2023, 12 (01):
  • [2] Effectiveness of breast electrical impedance imaging for clinically suspicious breast lesions
    Daglar, G.
    Senol, K.
    Yakut, Z., I
    Yuksek, Y. N.
    Tutuncu, T.
    Tez, M.
    Yesiltepe, C. H.
    BRATISLAVA MEDICAL JOURNAL-BRATISLAVSKE LEKARSKE LISTY, 2016, 117 (09): : 505 - 510
  • [3] Dynamic magnetic resonance imaging in evaluating suspicious breast lesions: Correlation with pathologic findings
    Snow, RD
    Dyess, DL
    Harpen, MD
    Kreisberg, CN
    Tucker, JA
    SOUTHERN MEDICAL JOURNAL, 1998, 91 (06) : 527 - 532
  • [4] MR-tomographic characterization of suspicious breast lesions with dynamic susceptibility contrast sequences.
    Dannert, S
    Krück, W
    Schick, F
    Claussen, CD
    Müller-Schimpfle, M
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2001, 173 (01): : 38 - 43
  • [5] Magnetic resonance-guided biopsy of suspicious breast lesions with a handheld vacuum biopsy device
    Gebauer, B.
    Bostanjoglo, M.
    Moesta, K. T.
    Schneider, W.
    Schlag, P. M.
    Felix, R.
    ACTA RADIOLOGICA, 2006, 47 (09) : 907 - 913
  • [6] Optical imaging of the breast: evaluation of deoxyhemoglobin concentration alteration in 166 patients with suspicious breast lesions
    Petrillo A.
    Catalano O.
    Fusco R.
    Filice S.
    Vallone P.
    Setola S.
    Granata V.
    Raiano C.
    Avino F.
    Di Bonito M.
    Botti G.
    European Radiology Experimental, 2 (1)
  • [7] HIGH-RESOLUTION MR IMAGING OF SUSPICIOUS BREAST-LESIONS
    OREL, SG
    SCHNALL, MD
    HOLLAND, GA
    YEH, IT
    TROUPIN, RH
    RADIOLOGY, 1992, 185 : 246 - 246
  • [8] Critical imaging analysis of suspicious non-palpable breast lesions
    Pereira, Renato de Oliveira
    da Silva, Benedito Borges
    REVISTA DA ASSOCIACAO MEDICA BRASILEIRA, 2020, 66 (12): : 1610 - 1612
  • [9] Fast and Noninvasive Characterization of Suspicious Lesions Detected at Breast Cancer X-Ray Screening: Capability of Diffusion-weighted MR Imaging with MIPs
    Bickelhaupt, Sebastian
    Laun, Frederik B.
    Tesdorff, Jana
    Lederer, Wolfgang
    Daniel, Heidi
    Stieber, Anne
    Delorme, Stefan
    Schlemmer, Heinz-Peter
    RADIOLOGY, 2016, 278 (03) : 689 - 697
  • [10] Microwave Breast Imaging System Prototype with Integrated Numerical Characterization
    Haynes, Mark
    Stang, John
    Moghaddam, Mahta
    INTERNATIONAL JOURNAL OF BIOMEDICAL IMAGING, 2012, 2012