Direct thermography-a new in vitro method to characterize temperature kinetics of ablation catheters

被引:1
|
作者
Fiek, M. [4 ]
Gindele, F. [1 ]
von Bary, C. [1 ]
Muessig, D. [2 ]
Lucic, A. [3 ]
Hoffmann, E. [1 ]
Reithmann, C. [1 ]
Steinbeck, G. [1 ]
机构
[1] Univ Munich, Med Hosp 1, Munich, Germany
[2] Univ Erlangen Nurnberg, Dept Biomed Engn, D-91054 Erlangen, Germany
[3] Tech Univ Munich, Dept Thermodynam A, D-80290 Munich, Germany
[4] Klinikum Pasing, Med Hosp 1, D-81241 Munich, Germany
关键词
Thermography; Radiofrequency ablation; Ablation catheter; In vitro model; RADIOFREQUENCY LESION FORMATION; NODAL REENTRANT TACHYCARDIA; HEAT-TRANSFER COEFFICIENT; SLOW PATHWAY ABLATION; TISSUE TEMPERATURES; THERMAL-PROPERTIES; FINITE-ELEMENT; PULMONARY VEIN; VIVO; MYOCARDIUM;
D O I
10.1007/s10840-013-9815-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
For the treatment of increasingly complex cardiac arrhythmias, new catheter designs as well as alternative energy sources are constantly being developed. However, there is presently no in vitro method available for assessment of the temperature changes induced at various myocardial levels during energy delivery. Therefore, our study was aimed at developing an in vitro model to record and display the temperature kinetics during ablation in the entire muscle cross section. A sapphire glass pane was inserted into one wall of the in vitro experimental set-up. Due to its thermodynamic properties, the temperature distribution in an adjacent cross section of the cardiac muscle can be measured exactly (+/- 1 A degrees C) through this pane by means of a thermography camera. Computer-supported image processing enables the colour-coded and two-dimensional display of the temperature kinetics during the energy application at any location of the myocardial cross section (+/- 0.5 mm). This new measuring methodology was validated by direct temperature measurements utilizing several intramyocardial thermo elements. This new method allows a temporal and spatial analysis of the temperature phenomena during ablation without the interference and spatial limitation of intramyocardial temperature probes. New ablation technologies can thus be evaluated, independent of the catheter configuration or source of energy used.
引用
收藏
页码:53 / 59
页数:7
相关论文
共 43 条