Direct thermography-a new in vitro method to characterize temperature kinetics of ablation catheters

被引:1
|
作者
Fiek, M. [4 ]
Gindele, F. [1 ]
von Bary, C. [1 ]
Muessig, D. [2 ]
Lucic, A. [3 ]
Hoffmann, E. [1 ]
Reithmann, C. [1 ]
Steinbeck, G. [1 ]
机构
[1] Univ Munich, Med Hosp 1, Munich, Germany
[2] Univ Erlangen Nurnberg, Dept Biomed Engn, D-91054 Erlangen, Germany
[3] Tech Univ Munich, Dept Thermodynam A, D-80290 Munich, Germany
[4] Klinikum Pasing, Med Hosp 1, D-81241 Munich, Germany
关键词
Thermography; Radiofrequency ablation; Ablation catheter; In vitro model; RADIOFREQUENCY LESION FORMATION; NODAL REENTRANT TACHYCARDIA; HEAT-TRANSFER COEFFICIENT; SLOW PATHWAY ABLATION; TISSUE TEMPERATURES; THERMAL-PROPERTIES; FINITE-ELEMENT; PULMONARY VEIN; VIVO; MYOCARDIUM;
D O I
10.1007/s10840-013-9815-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
For the treatment of increasingly complex cardiac arrhythmias, new catheter designs as well as alternative energy sources are constantly being developed. However, there is presently no in vitro method available for assessment of the temperature changes induced at various myocardial levels during energy delivery. Therefore, our study was aimed at developing an in vitro model to record and display the temperature kinetics during ablation in the entire muscle cross section. A sapphire glass pane was inserted into one wall of the in vitro experimental set-up. Due to its thermodynamic properties, the temperature distribution in an adjacent cross section of the cardiac muscle can be measured exactly (+/- 1 A degrees C) through this pane by means of a thermography camera. Computer-supported image processing enables the colour-coded and two-dimensional display of the temperature kinetics during the energy application at any location of the myocardial cross section (+/- 0.5 mm). This new measuring methodology was validated by direct temperature measurements utilizing several intramyocardial thermo elements. This new method allows a temporal and spatial analysis of the temperature phenomena during ablation without the interference and spatial limitation of intramyocardial temperature probes. New ablation technologies can thus be evaluated, independent of the catheter configuration or source of energy used.
引用
收藏
页码:53 / 59
页数:7
相关论文
共 43 条
  • [1] Direct thermography—a new in vitro method to characterize temperature kinetics of ablation catheters
    M. Fiek
    F. Gindele
    C. von Bary
    D. Muessig
    A. Lucic
    E. Hoffmann
    C. Reithmann
    G. Steinbeck
    Journal of Interventional Cardiac Electrophysiology, 2013, 38 : 53 - 59
  • [2] A method for the in vitro testing of cardiac ablation catheters
    Hsu, SS
    Hoh, L
    Rosenbaum, RM
    Rosen, A
    Walinsky, P
    Greenspon, AJ
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1996, 44 (10) : 1841 - 1847
  • [3] Method for the in vitro testing of cardiac ablation catheters
    Jefferson Medical Coll, Philadelphia, United States
    IEEE Trans Microwave Theory Tech, 10 pt 2 (1841-1847):
  • [4] Method to characterize phase transformation kinetics at high temperature
    Seymour, Kevin C.
    McCormack, Scott J.
    Ribero, Daniel
    Kriven, Waltraud M.
    JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 2024,
  • [5] A new method to characterize the kinetics of cholinesterases inhibited by carbamates
    Xiao, Qiaoling
    Zhou, Huimin
    Wei, Hong
    Du, Huaqiao
    Tan, Wen
    Zhan, Yiyi
    Pistolozzi, Marco
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2017, 144 : 175 - 182
  • [6] New Measurement Method to Characterize Piezoelectric SAW Substrates at Very High Temperature
    Nicolay, P.
    Elmazria, O.
    Sarry, F.
    Aubert, T.
    Bouvot, L.
    Hehn, M.
    2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, : 1877 - +
  • [7] Thermopixelgraphy: a new method for analyzing the skin temperature of the lower limbs through infrared thermography
    Fernandes, Alex de Andrade
    Marins, Joao Carlos Bouzas
    de Andrade, Andre Gustavo Pereira
    Albuquerque, Maicon Rodrigues
    Brito, Ciro Jose
    da Silva, Cristiano Diniz
    do Valle, Myrian Augusta Araujo Neves
    de Assis, Miller Gomes
    Garcia, Emerson Silami
    Pimenta, Eduardo Mendonca
    JOURNAL OF THERMAL ANALYSIS AND CALORIMETRY, 2024, 149 (22) : 12737 - 12745
  • [8] New Polish prototypes of 4-and 8-mm-tip nonirrigated radiofrequency ablation catheters: an in vitro study
    Kaczmarek, Krzysztof
    Cygankiewicz, Iwona
    Strzelecki, Aleksander
    Wranicz, Jerzy K.
    Ptaszynski, Pawel
    KARDIOLOGIA POLSKA, 2020, 78 (11) : 1142 - 1147
  • [9] A new method for the solution of the point kinetics equations in the presence of Newtonian temperature feedback
    Khoshahval, Farrokh
    Akbari, Morteza
    PROGRESS IN NUCLEAR ENERGY, 2020, 119
  • [10] New calibration method to measure rake face temperature of the tool during dry orthogonal cutting using thermography
    Soler, D.
    Aristimuno, P. X.
    Saez-de-Buruaga, M.
    Garay, A.
    Arrazola, P. J.
    APPLIED THERMAL ENGINEERING, 2018, 137 : 74 - 82