An old looped lead removal, complicated by its knotting over the active right ventricular lead - a case report

被引:0
|
作者
Malecka, Barbara [1 ]
Kutarski, Andrzej [1 ,2 ]
Pietura, Radoslaw [3 ]
Zabek, Andrzej [1 ]
Bednarek, Jacek [1 ]
Lelakowski, Jacek [1 ]
Szczerbo-Trojanowska, Malgorzata [3 ]
机构
[1] Krakowski Szpital Specialisty Jana Pawla 2, Klin Oddzial Elektrokardiol, PL-31302 Krakow, Poland
[2] Uniwersytet Med, Katedra & Klin Kardiol, Lublin, Poland
[3] Uniwersytet Med, Zaklad Radiol Zabiegowej & Neuroradiol, Lublin, Poland
关键词
cardiac pacing; cardiac pacing complications; lead extraction;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We present a 59 year old patient with advanced heart failure who underwent lead extraction. The lead was implanted 11 years ago, and 2 years later looped spontaneously following its fracture. The transvenous extraction process through inferior vena cava was complicated by knotting of the inactive lead over the active ventricular lead. The patient was pacemaker-dependent. Finally, the old lead was successfuly explanted. The active intact lead was used to create resynchronisation system.
引用
收藏
页码:865 / 868
页数:4
相关论文
共 50 条
  • [41] Left ventricular pacemaker lead insertion through the foramen ovale - A case report
    Altun, A
    Akdemir, O
    Erdogan, O
    Aslan, O
    Ozbay, G
    ANGIOLOGY, 2002, 53 (05) : 609 - 611
  • [42] TRANS-FEMORAL REMOVAL OF A CHRONICALLY IMPLANTED PACEMAKER LEAD - REPORT OF A CASE
    COPE, C
    LARRIEU, AJ
    ISAACSON, CS
    WOLK, LA
    GHOSH, SC
    ROTHKOPF, B
    ANNALS OF THORACIC SURGERY, 1986, 42 (03): : 329 - 330
  • [43] Identification and management of a subacute right ventricular perforation by an active-fixation permanent pacemaker lead in a dog
    Garcia-Guasch, L.
    Sa Borges, C.
    Sousa, P.
    Manubens Grau, J.
    JOURNAL OF VETERINARY CARDIOLOGY, 2019, 22 : 113 - 120
  • [44] Reccurrent pericardial chest pain:: A case of late right ventricular perforation after implantation of a transvenous active-fixation ICD lead
    Kautzner, J
    Bytesník, J
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (01): : 116 - 118
  • [45] SHOT THROUGH THE HEART: A NON-SURGICAL APPROACH TO MANAGING RIGHT VENTRICULAR LEAD PERFORATION COMPLICATED BY CARDIAC TAMPONADE
    Rao, Anila
    Tait, Don
    Rao, Lakshmi G.
    Mohan, Jay
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2023, 81 (08) : 3257 - 3257
  • [46] An unusual case of a subacute right ventricular perforation from a pacemaker lead with subsequent left hemothorax
    Dean, Ryan K.
    Subedi, Rogin
    Gill, Dalvir
    Nat, Amitpal
    COR ET VASA, 2018, 60 (05) : E479 - E481
  • [47] A rare case of late right ventricular perforation by a passive-fixation permanent pacemaker lead
    Tziakas, Dimitrios
    Alexoudis, Apostolos
    Konstantinou, Fotios
    Chalikias, Georgios
    Stakos, Dimitrios
    Bougioukas, Georgios
    EUROPACE, 2009, 11 (07): : 968 - 969
  • [48] Delayed right-ventricular perforation by pacemaker lead; a rare complication in a 12-year-old girl
    Aykan, Hayrettin Hakan
    Akin, Alper
    Ertugrul, Ilker
    Karagoz, Tevfik
    TURK KARDIYOLOJI DERNEGI ARSIVI-ARCHIVES OF THE TURKISH SOCIETY OF CARDIOLOGY, 2015, 43 (02): : 185 - 187
  • [49] Giant right atrial thrombus associated with ICD lead externalized conductors: a case report
    de Waroux, Jean-Benoit le Polain
    Scavee, Christophe
    Marchandise, Sebastien
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2018, 2 (02)
  • [50] LEAD-INTOXICATION DUE TO RETAINED BULLET IN RIGHT HIP - A CASE-REPORT
    OVARTLARNPORN, B
    PRAKAITIP, D
    JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND, 1985, 68 (11): : 612 - 616