Stereotactic radiosurgery for intradural spine tumors using cone-beam CT image guidance

被引:10
|
作者
Monserrate, Andres [1 ]
Zussman, Benjamin [1 ]
Ozpinar, Alp [1 ]
Niranjan, Ajay [1 ]
Flickinger, John C. [2 ,3 ]
Gerszten, Peter C. [1 ,2 ]
机构
[1] Univ Pittsburgh, Dept Neurol Surg, Med Ctr, Pittsburgh, PA 15260 USA
[2] Univ Pittsburgh, Med Ctr, Dept Radiat Oncol, Pittsburgh, PA USA
[3] Pittsburgh Canc Inst, Pittsburgh, PA 15213 USA
关键词
spine radiosurgery; intradural tumors; spine tumors; cone-beam CT; CONFORMAL RADIOTHERAPY; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; PARASPINAL TUMORS; CYBERKNIFE; TOLERANCE; SYSTEM; CORD;
D O I
10.3171/2016.9.FOCUS16356
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective Cone-beam CT (CBCT) image guidance technology has been widely adopted for spine radiosurgery delivery. There is relatively little experience with spine radiosurgery for intradural tumors using CBCT image guidance. This study prospectively evaluated a series of intradural spine tumors treated with radiosurgery. Patient setup accuracy for spine radiosurgery delivery using CBCT image guidance for intradural spine tumors was determined. Methods Eighty-two patients with intradural tumors were treated and prospectively evaluated. The positioning deviations of the spine radiosurgery treatments in patients were recorded. Radiosurgery was delivered using a linear accelerator with a beam modulator and CBCT image guidance combined with a robotic couch that allows positioning correction in 3 translational and 3 rotational directions. To measure patient movement, 3 quality assurance CBCTs were performed and recorded in 30 patients: before, halfway, and after the radiosurgery treatment. The positioning data and fused images of planning CT and CBCT from the treatments were analyzed to determine intrafraction patient movements. From each of 3 CBCTs, 3 translational and 3 rotational coordinates were obtained. Results The radiosurgery procedure was successfully completed for all patients. Lesion locations included cervical (22), thoracic (17), lumbar (38), and sacral (5). Tumor histologies included schwannoma (27), neurofibromas (18), meningioma (16), hemangioblastoma (8), and ependymoma (5). The mean prescription dose was 17 Gy (range 12-27 Gy) delivered in 1-3 fractions. At the halfway point of the radiation, the translational variations and standard deviations were 0.4 +/- 0.5, 0.5 +/- 0.8, and 0.4 +/- 0.5 mm in the lateral (x), longitudinal (y), and anteroposterior (z) directions, respectively. Similarly, the variations immediately after treatment were 0.5 +/- 0.4, 0.5 +/- 0.6, and 0.6 +/- 0.5 mm along x, y, and z directions, respectively. The mean rotational angles were 0.3 degrees +/- 0.4 degrees, 0.3 degrees +/- 0.4 degrees, and 0.3 degrees +/- 0.4 degrees along yaw, roll, and pitch, respectively, at the halfway point and 0.5 degrees +/- 0.5 degrees, 0.4 degrees +/- 0.5 degrees, and 0.2 degrees +/- 0.3 degrees immediately after treatment. Conclusions Radiosurgery offers an alternative treatment option for intradural spine tumors in patients who may not be optimal candidates for open surgery. CBCT image guidance for patient setup for spine radiosurgery is accurate and successful in patients with intradural tumors.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Radiosurgery for benign tumors of the spine using the Synergy S with cone-beam computed tomography image guidance
    Gerszten, Peter C.
    Chen, Stephanie
    Quader, Mubina
    Xu, Yuanguang
    Novotny, Josef, Jr.
    Flickinger, John C.
    [J]. JOURNAL OF NEUROSURGERY, 2012, 117 : 197 - 202
  • [2] Cone-beam CT based image-guidance for extracranial stereotactic radiotherapy of intrapulmonary tumors
    Guckenberger, Matthias
    Meyer, Juergen
    Wilbert, Juergen
    Baier, Kurt
    Mueller, Gerd
    Wulf, Joern
    Flentje, Michael
    [J]. ACTA ONCOLOGICA, 2006, 45 (07) : 897 - 906
  • [3] Intradural Spine Tumors Radiosurgery Delivery Accuracy using CBCT Image Guidance
    Shankar, V.
    Shreya, V. Sai
    Bhavya, P.
    Vyas, H.
    Haritha, C.
    Deopujari, C. E.
    Velho, V.
    Turel, K.
    Andar, U.
    Saxena, U.
    Basu, T.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E688 - E688
  • [4] Determination of Stereotactic Shifts for Frame-Based Stereotactic Radiosurgery Using Cone-Beam Computed Tomography Image Guidance
    Hadley, A.
    Seneviratne, D.
    Serago, C. F.
    Herchko, S.
    Welch, E.
    Peterson, J. L.
    Trifiletti, D. M.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E763 - E763
  • [5] Frameless radiosurgery on a conventional Linac utilizing cone-beam CT image guidance
    Remeijier, P.
    Minkema, D.
    Betgen, A.
    deWit, L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S13 - S13
  • [6] Frameless stereotactic treatment of brain Metastases using cone-beam CT based image guidance
    Remeijer, P.
    Minkema, D.
    Betgen, A.
    DeWit, L.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S687 - S688
  • [7] Stereotactic Cone-Beam CT Guided Radiosurgery of Trigeminal Neuralgia
    Ziemer, B.
    Natsuaki, Y.
    Perez-Andujar, A.
    Morin, O.
    Weethee, B.
    Ma, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E658 - E659
  • [8] Frameless stereotactic treatment of brain metastases using cone-beam CT based image guidance.
    Remeijer, P.
    Minkema, D.
    Betgen, A.
    DeWit, L.
    van Herk, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S195 - S195
  • [9] Stereotactic Radiosurgery for Malignant Intradural and Intramedullary Tumors of the Spine
    Elibe, Erinma
    Boyce-Fappiano, David
    Lee, Ian
    Rock, Jack
    Siddiqui, M. Salim
    Siddiqui, Farzan
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 101 (02): : E25 - E25
  • [10] Stereotactic Radiosurgery for Malignant Intradural and Intramedullary Tumors of the Spine
    Elibe, E.
    Boyce-Fappiano, D.
    Siddiqui, S.
    Lee, I.
    Rock, J.
    Siddiqui, F.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E518 - E518