Efficacy of percutaneous fibrin sheath stripping in restoring patency of tunneled hemodialysis catheters

被引:33
|
作者
Brady, PS [1 ]
Spence, LD [1 ]
Levitin, A [1 ]
Mickolich, CT [1 ]
Dolmatch, BL [1 ]
机构
[1] Cleveland Clin Fdn, Dept Radiol, Hosp HB6, Cleveland, OH 44195 USA
关键词
D O I
10.2214/ajr.173.4.10511171
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. Our objective was to determine the effectiveness of percutaneous fibrin sheath stripping as a method of restoring function to failing tunneled hemodialysis catheters. SUBJECTS AND METHODS. A total of 131 percutaneous fibrin sheath stripping procedures were performed on 100 failing tunneled hemodialysis catheters in 91 patients. Only the initial stripping procedure of the first catheter inserted in each patient was included fur analysis. Patients were excluded if an additional cause of catheter failure was noted at the time of percutaneous fibrin stripping. Failure of the hemodialysis catheter was defined as inability to sustain an average blood flow rate of 250 ml/min or more in a hemodialysis session. Patients were followed up until the time of catheter failure, catheter removal, or a second stripping. Poststripping primary patency and complication rates were determined. RESULTS. The technical success of the procedure was 95.6%. Median follow-up was 16 weeks (range, 0-128 weeks). The overall median duration of primary patency after the first stripping was 89 days (3 months). No statistically significant difference in primary patency rates was seen between patients who underwent fibrin sheath demonstration by contrast injection and those who did not (p = .71). Female patients were statistically more likely to have catheter failure after catheter stripping than were male patients (p = .02). The route of catheter insertion did not significantly influence poststripping patency rates. No complications were associated with the procedure. CONCLUSION. Percutaneous fibrin sheath stripping is a safe, effective method of restoring patency to ailing hemodialysis catheters when the failure is due to fibrin sheath formation and other causes are excluded.
引用
收藏
页码:1023 / 1027
页数:5
相关论文
共 46 条
  • [1] Restoration of patency in failing tunneled hemodialysis catheters:: A comparison of catheter exchange, exchange and balloon disruption of the fibrin sheath, and femoral stripping
    d'Othée, Bertrand Janne
    Tham, Jacques C.
    Sheiman, Robert G.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (06) : 1011 - 1015
  • [2] Management of fibrin sheath associated with tunneled hemodialysis catheters
    Beathard, GA
    Arnold, P
    Litchfield, T
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2003, 14 : 241A - 242A
  • [3] Fibrin sheath stripping versus catheter exchange for the treatment of failed tunneled hemodialysis catheters: Randomized clinical trial
    Merport, M
    Murphy, TP
    Egglin, TK
    Dubel, GJ
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (09) : 1115 - 1120
  • [4] Maintaining patency of tunneled hemodialysis catheters - Efficacy of aspirin compared to warfarin
    Obialo, CI
    Conner, AC
    Lebon, LF
    SCANDINAVIAN JOURNAL OF UROLOGY AND NEPHROLOGY, 2003, 37 (02): : 172 - 176
  • [5] Patency of Femoral Tunneled Hemodialysis Catheters and Factors Predictive of Patency Failure
    Kirsteen R. Burton
    Lancia L. Q. Guo
    Kong T. Tan
    Martin E. Simons
    Kenneth W. Sniderman
    John R. Kachura
    John R. Beecroft
    Dheeraj K. Rajan
    CardioVascular and Interventional Radiology, 2012, 35 : 1396 - 1402
  • [6] Patency of Femoral Tunneled Hemodialysis Catheters and Factors Predictive of Patency Failure
    Burton, Kirsteen R.
    Guo, Lancia L. Q.
    Tan, Kong T.
    Simons, Martin E.
    Sniderman, Kenneth W.
    Kachura, John R.
    Beecroft, John R.
    Rajan, Dheeraj K.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 35 (06) : 1396 - 1402
  • [7] Infectious outcomes of fibrin sheath disruption in tunneled dialysis catheters
    Waters, Mara
    Huszti, Ella
    Ramirez, Maria Erika
    Lok, Charmaine E.
    JOURNAL OF VASCULAR ACCESS, 2023, 24 (05): : 1091 - 1098
  • [8] Failing hemodialysis catheters: Fibrin sheath stripping versus exchange through existing subcutaneous tunnel
    Szymski, GX
    Rosenblum, JD
    Leef, JA
    Zernich, BW
    RADIOLOGY, 1996, 201 : 1551 - 1551
  • [9] Transvenous removal of fibrin sheaths from tunneled hemodialysis catheters
    Haskal, ZJ
    Leen, VH
    ThomasHawkins, C
    ShlanskyGoldberg, RD
    Baum, RA
    Soulen, MC
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (04) : 513 - 517
  • [10] SURVIVAL OF TUNNELED HEMODIALYSIS CATHETERS AFTER PERCUTANEOUS PLACEMENT
    Weber, Ewa
    Liberek, Tomasz
    Wolyniec, Wojciech
    Rutkowski, Boleslaw
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2015, 30