Total hip arthroplasty by using a cementless ultrashort stem: A subject-specific finite element analysis for a young patient clinical case

被引:2
|
作者
Epasto, Gabriella [1 ]
Foti, Albina [2 ]
Guglielmino, Eugenio [1 ]
Rosa, Michele A. [2 ]
机构
[1] Univ Messina, Dept Elect Engn Chem & Ind Engn, I-98166 Messina, Italy
[2] Univ Messina, Dept Special Surg, Div Traumatol & Orthoped, I-98166 Messina, Italy
关键词
Biomechanics; orthopedics; finite element analysis; computed tomography; total hip arthroplasty; short-stemmed implant; PROXIMAL FEMUR; FEMORAL COMPONENT; BONE-RESORPTION; LOAD-TRANSFER; LONG BONES; STRESS; MODELS; VALIDATION; PROSTHESIS; UNCERTAINTIES;
D O I
10.1177/0954411913482267
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this article, a subject-specific finite element analysis has been developed to study a clinical case of a surgically misaligned hip prosthesis with an ultrashort stem. It was set out to study the strain energy density pattern, comparing the results obtained with computed tomography images. The authors developed two other numerical models: the first one analyzes the stress and strain distributions in the healthy femur (without prosthesis) and the second one analyzes the same boneimplant biomechanical system of the clinical case but assuming the prosthesis in the proper position. The misaligned prosthesis produced an overload at the proximal posterior plane of the femur, as confirmed by computed tomography images, which detect the formation of new bone. The numerical model of the correctly positioned prosthesis demonstrated that the bone is not overloaded and that the position of neutral axis does not significantly shift from the physiological condition.
引用
收藏
页码:757 / 766
页数:10
相关论文
共 50 条
  • [21] Radiographic and clinical analysis of cementless acetabular fixation in total hip arthroplasty
    张晖
    裴福兴
    杨静
    沈彬
    石锐
    Chinese Journal of Traumatology, 2005, (05) : 263 - 268
  • [22] The influence of contact ratio and its location on the primary stability of cementless total hip arthroplasty: A finite element analysis
    Reimeringer, M.
    Nuno, N.
    JOURNAL OF BIOMECHANICS, 2016, 49 (07) : 1064 - 1070
  • [23] Predicting the subject-specific primary stability of cementless implants during pre-operative planning: Preliminary validation of subject-specific finite-element models
    Reggiani, B.
    Cristofolini, L.
    Varini, E.
    Viceconti, M.
    JOURNAL OF BIOMECHANICS, 2007, 40 (11) : 2552 - 2558
  • [24] Total hip arthroplasty for dysplastic coxarthrosis using a cementless Wagner Cone stem
    Giuseppe Solarino
    Giovanni Vicenti
    Andrea Piazzolla
    Francesco Maruccia
    Angela Notarnicola
    Biagio Moretti
    Journal of Orthopaedics and Traumatology, 2021, 22
  • [25] Total Hip Arthroplasty Using a Cylindrical Cementless Stem in Patients With a Small Physique
    Nakamura, Yoshihide
    Mitsui, Hiromasa
    Kikuchi, Akira
    Toh, Satoshi
    Katano, Hiroshi
    JOURNAL OF ARTHROPLASTY, 2011, 26 (01): : 77 - 81
  • [26] Total hip arthroplasty for dysplastic coxarthrosis using a cementless Wagner Cone stem
    Solarino, Giuseppe
    Vicenti, Giovanni
    Piazzolla, Andrea
    Maruccia, Francesco
    Notarnicola, Angela
    Moretti, Biagio
    JOURNAL OF ORTHOPAEDICS AND TRAUMATOLOGY, 2021, 22 (01)
  • [27] Computational modeling of polyethylene wear in total hip arthroplasty using patient-specific kinematics-coupled finite element analysis
    Peng, Yun
    Arauz, Paul
    An, Shuai
    Kwon, Young-Min
    TRIBOLOGY INTERNATIONAL, 2019, 129 : 162 - 166
  • [28] Finite element analysis of the tibial bone graft in cementless total knee arthroplasty
    Koji Totoribe
    Etsuo Chosa
    Go Yamako
    Hiroaki Hamada
    Koki Ouchi
    Shutaro Yamashita
    Gang Deng
    Journal of Orthopaedic Surgery and Research, 13
  • [29] Bony impingement affects range of motion after total hip arthroplasty: A subject-specific approach
    Kessler, Oliver
    Pati, Shantanu
    Stefan, Wirth
    Mayr, Eckart
    Colwell, Clifford W., Jr.
    D'Lima, Darryl D.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2008, 26 (04) : 443 - 452
  • [30] Static structural analysis of different stem designs used in total hip arthroplasty using finite element method
    Chethan, K. N.
    Zuber, Mohammad
    Bhat, Shyamasunder N.
    Shenoy, Satish B.
    Kini, Chandrakant R.
    HELIYON, 2019, 5 (06)