Latest Progress on the Iseult/INUMAC Whole Body 11.7 T MRI Magnet

被引:6
|
作者
Vedrine, P. [1 ]
Maksoud, W. Abdel [1 ]
Aubert, G. [1 ]
Beaudet, F. [2 ]
Belorgey, J. [1 ]
Bermond, S. [1 ]
Berriaud, C.
Bredy, Ph. [1 ]
Bresson, D. [2 ]
Donati, A. [1 ]
Dubois, O. [1 ]
Gilgrass, G. [3 ]
Juster, F. P. [1 ]
Lannou, H. [1 ]
Meuris, C. [1 ]
Molinie, F. [1 ]
Nusbaum, M. [2 ]
Nunio, F. [1 ]
Payn, A. [1 ]
Schild, T. [1 ]
Scola, L. [1 ]
Sinanna, A. [1 ]
机构
[1] CEA Saclay, Irfu, F-91191 Gif Sur Yvette, France
[2] ALSTOM Magnets & Superconductors, F-90000 Belfort, France
[3] Siemens Healthcare MR, Eynsham OX29 4BP, Witney, England
关键词
Magnetic resonance imaging; niobium titanium; superconducting magnet;
D O I
10.1109/TASC.2011.2176093
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The Whole Body 11.7 T MRI Magnet is an actively shielded magnet system, with a stored energy of 338 MJ and an inductance of 308 H. Operating at a homogeneous field level of 11.75 T within a 90 cm warm bore, the cryostat has external dimensions of 4.8 m in diameter and 5.0 m in length. It is part of the Iseult/Inumac project, a French-German initiative focused on very-high-magnetic-field molecular imaging to improve sensitivity, spatial, temporal, and spectral resolution for preclinical and/or clinical MR systems. After the qualification of two first unit lengths of 820 m, the NbTi conductor with a current of 1483 A is now being produced at Luvata Waterbury. Winding of the main coil, made of 170 double pancakes, is starting at Alstom Belfort. Several pieces of equipment have already been delivered to the Neurospin site, CEA Saclay,; including the main refrigerator produced by Air Liquide. Several prototypes have been tested and confirmed the soundness of the magnet design. This paper describes the 11.7 T magnet and the latest progress in its design and fabrication.
引用
收藏
页数:4
相关论文
共 50 条
  • [41] Globally Optimal Algorithm for Design of 0.7 T Actively Shielded Whole-Body Open MRI Superconducting Magnet System
    Ni, Zhipeng
    Hu, Geli
    Li, Lankai
    Yu, Gongqi
    Wang, Qiuliang
    Yan, Luguang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2013, 23 (03)
  • [42] A simulation-based design study of superconducting zonal shim coil for a 9.4 T whole-body MRI magnet
    Kim, Geonyoung
    Choi, Kibum
    Park, Jeonghwan
    Bong, Uijong
    Bang, Jeseok
    Hahn, Seungyong
    PROGRESS IN SUPERCONDUCTIVITY AND CRYOGENICS, 2020, 22 (01): : 12 - 16
  • [43] Systematic analysis of the quench process performance and simulation of the 9.4 T-800 mm whole-body MRI magnet
    Chen, Shunzhong
    Wang, Lei
    Zhang, Zili
    Wang, Yaohui
    Wang, Hui
    Cheng, Junsheng
    Dai, Yinming
    Wang, Qiuliang
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2023, 36 (04):
  • [44] Thermal and Electrical Behaviour of the Persistent Current Switch for a Whole-Body Superconducting MRI Magnet
    Nandawadekar, Ajit
    Soni, V
    Suman, N.
    Ram, Sankar T.
    Kumar, R.
    Saini, S. K.
    Sharma, R. G.
    Singh, Mukhtiar
    Kar, Soumen
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2021, 31 (05)
  • [45] A cryogen-free superconducting magnet with 95 cm warm bore for whole body MRI
    Xu, BX
    Schnurer, S
    Obasih, K
    Mruzek, M
    Ige, O
    Lochner, R
    Helbing, J
    Mantone, D
    IEEE TRANSACTIONS ON MAGNETICS, 1996, 32 (04) : 2637 - 2639
  • [46] Design, construction and performance testing of a 1.5 T cryogen-free low-temperature superconductor whole-body MRI magnet
    Wang, Yulin
    Wang, Sucong
    Liang, Ping
    Brusic, Vladimir
    Zeng, Jie
    Liu, Bingjian
    Ge, Yuwei
    Zheng, Jianjun
    Zhang, Jingfeng
    Wang, Chengbo
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2023, 36 (04):
  • [47] Thermal Analysis of the 4 K Cryostat Within a 1.5 T Conduction-Cooled Superconducting Magnet for Whole-Body MRI Application
    Qu, Hongyi
    Wu, Huafang
    Feng, Zhichao
    Xiao, Jing
    Chen, Qi
    Wang, Hui
    Wang, Qiuliang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2024, 34 (08)
  • [48] Numerical analysis of the coupling loss induced quench protection for a 1.5T whole-body MgB2 MRI magnet
    Poole, Charles
    Baig, Tanvir
    Deissler, Robert J.
    Martens, Michael
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2017, 30 (10):
  • [49] Composite support system for 1.5 T whole-body MRI cryostat
    Suman, Navneet
    Kar, Soumen
    Siddiquee, Arshad Noor
    MATERIALS TODAY-PROCEEDINGS, 2022, 62 : 309 - 317
  • [50] On the passive shimming of a 7 T whole-body MRI superconducting magnet: Implementation with minimized ferromagnetic materials usage and operable magnetic force control
    Wang, Wenchen
    Wang, Yaohui
    Wang, Hui
    Cheng, Junsheng
    Qu, Hongyi
    Wang, Cong
    Niu, Chaoqun
    Liu, Feng
    MEDICAL PHYSICS, 2023, 50 (10) : 6514 - 6524