Microvascular Imaging Using Compressed Sensing at 7T MRI: A Preliminary Study

被引:6
|
作者
Park, Chan-A [1 ]
Kang, Daehun [2 ,3 ]
Son, Young-Don [4 ,5 ]
Kim, Hang-Keun [4 ,5 ]
Kang, Chang-Ki [4 ,6 ]
机构
[1] Korea Basic Sci Inst, Bioimaging Res Team, Ochang, Chungbuk, South Korea
[2] Tohoku Univ, Grad Sch Informat Sci, Sendai, Miyagi 980, Japan
[3] Tohoku Univ, Kansei Fukushi Res Inst, Sendai, Miyagi 980, Japan
[4] Gachon Univ, Neurosci Res Inst, Inchon, South Korea
[5] Gachon Univ, Coll Hlth Sci, Dept Biomed Engn, Inchon, South Korea
[6] Gachon Univ, Coll Hlth Sci, Dept Radiol Sci, Inchon, South Korea
关键词
HIGH-RESOLUTION MRI;
D O I
10.1007/s00723-015-0713-0
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
The aim of this study was to obtain high-resolution images with a short acquisition time using a compressed sensing (CS) technique for microvascular imaging, especially of the lenticulostriate arteries (LSAs), with 7T magnetic resonance imaging (MRI). Initially, two different radio frequency (RF) coils, birdcage and multi-channel phased array, were tested to compare images of micro-vessels acquired using the CS technique with those acquired using conventional magnetic resonance angiography (MRA) at 7T MRI. High-resolution CS MRA as well as low- and high-resolution conventional time-of-flight MRA images obtained with a multi-channel RF coil was compared to evaluate the feasibility of the CS technique for microvascular imaging. Signal intensities were then used to quantitatively estimate vessel sizes. Compared to low-resolution MRA, the CS technique improved the depiction of the LSAs and provided a substantially reduced partial volume effect of the vessels. Despite the short acquisition time of the CS technique, it clearly depicted the LSA morphology almost as well as the high-resolution MRA. The measured size difference between micro-vessels that were analyzed using the CS technique and those analyzed using high-resolution MRA was about 20 %; however, the difference between the high-resolution and the low-resolution MRA was as high as 100 %. This study evaluated the feasibility of the CS technique for microvascular imaging with 7T MRI. A CS technique that combines high-resolution imaging with a short acquisition time would be a very effective method for microvascular imaging in both research and clinical studies.
引用
收藏
页码:1189 / 1197
页数:9
相关论文
共 50 条
  • [1] Microvascular Imaging Using Compressed Sensing at 7T MRI: A Preliminary Study
    Chan-A Park
    Daehun Kang
    Young-Don Son
    Hang-Keun Kim
    Chang-Ki Kang
    Applied Magnetic Resonance, 2015, 46 : 1189 - 1197
  • [2] Compressed sensing sodium MRI of cartilage at 7T: Preliminary study
    Madelin, Guillaume
    Chang, Gregory
    Otazo, Ricardo
    Jerschow, Alexej
    Regatte, Ravinder R.
    JOURNAL OF MAGNETIC RESONANCE, 2012, 214 : 360 - 365
  • [3] Microvascular imaging of symptomatic and asymptomatic MCA steno-occlusive patients using ultra-high-field 7T MRI : a preliminary comparative study
    Lee, Y.
    Kang, C.
    Kim, K.
    Park, C.
    Kim, H. Y.
    Min, Y.
    Park, C. W.
    Kim, Y.
    Cho, Z.
    CEREBROVASCULAR DISEASES, 2013, 35 : 257 - 257
  • [4] Advances in MR angiography with 7T MRI: From microvascular imaging to functional angiography
    Park, Chan-A
    Kang, Chang-Ki
    Kim, Young-Bo
    Cho, Zang-Hee
    NEUROIMAGE, 2018, 168 : 269 - 278
  • [5] Volumetric glutamate imaging (GluCEST) using 7T MRI can lateralize nonlesional temporal lobe epilepsy: A preliminary study
    Hadar, Peter N.
    Kini, Lohith G.
    Nanga, Ravi Prakash Reddy
    Shinohara, Russell T.
    Chen, Stephanie H.
    Shah, Preya
    Wisse, Laura E. M.
    Elliott, Mark A.
    Hariharan, Hari
    Reddy, Ravinder
    Detre, John A.
    Stein, Joel M.
    Das, Sandhitsu
    Davis, Kathryn A.
    BRAIN AND BEHAVIOR, 2021, 11 (08):
  • [6] Single-shot spiral diffusion-weighted imaging at 7T using expanded encoding with compressed sensing
    Varela-Mattatall, Gabriel
    Dubovan, Paul I. I.
    Santini, Tales
    Gilbert, Kyle M. M.
    Menon, Ravi S. S.
    Baron, Corey A. A.
    MAGNETIC RESONANCE IN MEDICINE, 2023, 90 (02) : 615 - 623
  • [7] 7T MRI Loop Antenna for Carotid Imaging
    Hinge, Lars
    Mortensen, Nicolai
    Zhurbenko, Vitaliy
    Boer, Vincent O.
    Wang, Wenjun
    2020 50TH EUROPEAN MICROWAVE CONFERENCE (EUMC), 2020,
  • [8] 7T MRI Loop Antenna for Carotid Imaging
    Hinge, Lars
    Mortensen, Nicolai
    Zhurbenko, Vitaliy
    Boer, Vincent O.
    Wang, Wenjun
    2020 50TH EUROPEAN MICROWAVE CONFERENCE (EUMC), 2020, : 1007 - 1010
  • [9] 7T MRI Loop Antenna for Carotid Imaging
    Hinge, Lars
    Mortensen, Nicolai
    Zhurbenko, Vitaliy
    Boer, Vincent O.
    Wang, Wenjun
    2020 50TH EUROPEAN MICROWAVE CONFERENCE (EUMC), 2020,
  • [10] Improved characterization of lenticulostriate arteries using compressed sensing time-of-flight at 7T
    Zhe Zhang
    Qingle Kong
    Yingkui Zhang
    Wanlin Zhu
    Ning Wei
    Yuyuan Xu
    Yue Suo
    Xia Meng
    Patrick Liebig
    Zihao Zhang
    Yongjun Wang