Electromagnetic Navigation in Distal Locking of Long Diaphyseal Interlocking Intramedullary Nailing

被引:0
|
作者
Zhou, Zhentao [1 ]
Zhou, Xiaozhong [1 ]
Shan, Bingchen [1 ]
Zhou, Haibin [1 ]
Lu, Zhengfeng [1 ]
Dong, Qirong [1 ]
机构
[1] Soochow Univ, Affiliated Hosp 2, Dept Orthopaed, Suzhou 215004, Peoples R China
关键词
intramedullary nailing; Distal locking; Femoral/tibial fractures; Electromagnetic navigation; RADIATION-EXPOSURE; TARGETING SYSTEM; FRACTURES; INSERTION; DEVICE; SCREWS;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To describe the applications and effects of electromagnetic navigation (EN) technology in distal locking for the treatment of long diaphyseal fracture (LDF) with interlocking intramedullary nailing (IIN). Study Design: An interventional study. Place and Duration of Study: The Second Affiliated Hospital of Soochow University, China, from March 2013 to July 2014. Methodology: Patients who underwent IIN-LDF were selected. Twenty-four (50%) of whom were operated under EN guidance (group A) and the other 24 (50%) under conventional targeting guidance (group B). The distal locking time and X-ray irradiation time of the two groups were compared. Results: Each group included 16 (33.3%) cases of femoral fracture and 8 (16.7%) cases of tibial fracture. The success rate of distal locking in group A was higher than that in group B (95.8% vs. 83.3%, p=0.045). There were statistically significant differences in the distal locking time and X-ray irradiation time of femoral intramedullary nailing between the two groups (p=0.027 and p=0.001, respectively). There were no statistically significant differences in the distal locking time and X-ray irradiation time of tibial intramedullary nailing between the two groups (p=0.347 and p=0.056, respectively). Conclusion: EN-IN was advantageous as it enabled easy targeting, significantly reduced intraoperative fluoroscopy and operation time and small trauma and had other advantages when used for treating LDFs, especially femoral diaphyseal fractures.
引用
收藏
页码:975 / 979
页数:5
相关论文
共 50 条
  • [21] Retrograde interlocking nailing of distal femoral fractures with the supracondylar intramedullary nail
    Ostermann, PAW
    Hahn, MP
    Ekkernkamp, A
    David, A
    Muhr, G
    CHIRURG, 1996, 67 (11): : 1135 - 1140
  • [22] An arrangement of supports to improve access for distal interlocking in intramedullary tibial nailing
    Haughton, B. A.
    Patel, M. P.
    Livesey, J. P.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2013, 95 (07) : 532 - 532
  • [23] INTERLOCKING INTRAMEDULLARY NAILING FOR SUPRACONDYLAR AND INTERCONDYLAR FRACTURES OF THE DISTAL PART OF THE FEMUR
    LEUNG, KS
    SHEN, WY
    SO, WS
    MUI, LT
    GROSSE, A
    SHATIN, NT
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (03): : 332 - 340
  • [24] Efficacy of TiRobot assistance in distal locking for femoral intramedullary nailing
    Peng, Yuanxiang
    Cai, Feng
    Liu, Luhua
    Qu, Gaoyang
    Wu, Pengyun
    Xia, Sijie
    Shen, Feng
    MEDICINE, 2025, 104 (01)
  • [25] Intramedullary nailing for humeral shaft fractures: Is distal locking necessary?
    Egrise, Francois
    Clowez, Gilles
    Recanatesi, Nicolas
    Tabutin, Jacques
    Borge, Paul Emile
    Gastaud, Olivier
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2023, 109 (05)
  • [26] Is second distal locking screw necessary in interlocking nailing of femur fractures?
    Karaarslan, A.
    Karakasli, A.
    Aycan, H.
    Taylan, O.
    Uzun, B.
    Sesli, E.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2015, 66 (01) : 9 - 15
  • [27] Robotic distal locking of intramedullary nailing: Technical description and cadaveric testing
    Panzica, Martin
    Suero, Eduardo M.
    Westphal, Ralf
    Citak, Musa
    Liodakis, Emmanouil
    Hawi, Nael
    Petri, Max
    Krettek, Christian
    Stuebig, Timo
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2017, 13 (04):
  • [28] The Role of Biplanar Distal Locking in Intramedullary Nailing of Tibial Shaft Fractures
    Alemdaroglu, Kadir Bahadir
    Iltar, Serkan
    Ozturk, Alper
    Gultac, Emre
    Yucens, Mehmet
    Aydogan, Nevres Hurriyet
    ARCHIVES OF BONE AND JOINT SURGERY-ABJS, 2019, 7 (01): : 33 - 37
  • [29] Reduced Surgical Time and Higher Accuracy of Distal Locking with the Electromagnetic Targeting System in Humeral Shaft Intramedullary Nailing
    Fu Huichao
    Wu Xiaoming
    ORTHOPAEDIC SURGERY, 2020, 12 (05) : 1413 - 1420
  • [30] Intramedullary interlocking nailing of humeral fractures
    Ikpeme, JO
    Farrar, MJ
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1997, 28 (08): : 571 - 571