MAMMOGRAPHY DOSIMETRY USING AN IN-HOUSE DEVELOPED POLYMETHYL METHACRYLATE PHANTOM

被引:5
|
作者
Sharma, Reena [1 ]
Sharma, Sunil Dutt [1 ]
Mayya, Y. S. [1 ]
Chourasiya, G. [1 ]
机构
[1] Bhabha Atom Res Ctr, Radiol Phys & Advisory Div, Bombay 400094, Maharashtra, India
关键词
BREAST;
D O I
10.1093/rpd/ncr476
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Phantom-based measurements in mammography are well-established for quality assurance (QA) and quality control (QC) procedures involving equipment performance and comparisons of X-ray machines. Polymethyl methacrylate (PMMA) is among the best suitable materials for simulation of the breast. For carrying out QA/QC exercises in India, a mammographic PMMA phantom with engraved slots for keeping thermoluminescence dosemeters (TLD) has been developed. The radiation transmission property of the developed phantom was compared with the commercially available phantoms for verifying its suitability for mammography dosimetry. The breast entrance exposure (BEE), mean glandular dose (MGD), percentage depth dose (PDD), percentage surface dose distribution (PSDD), calibration testing of automatic exposure control (AEC) and density control function of a mammography machine were measured using this phantom. MGD was derived from the measured BEE following two different methodologies and the results were compared. The PDD and PSDD measurements were carried out using LiF: Mg, Cu, P chips. The in-house phantom was found comparable with the commercially available phantoms. The difference in the MGD values derived using two different methods were found in the range of 17.5-32.6%. Measured depth ranges in the phantom lie between 0.32 and 0.40 cm for 75% depth dose, 0.73 and 0.92 cm for 50 % depth dose, and 1.54 and 1.78 cm for 25% depth dose. Higher PSDD value was observed towards chest wall edge side of the phantom, which is due to the orientation of cathode anode axis along the chest wall to the nipple direction. Results obtained for AEC configuration testing shows that the observed mean optical density (O.D) of the phantom image was 1.59 and O.D difference for every successive increase in thickness of the phantom was within +/- 0.15 O.D. Under density control function testing, at -2 and -1 density settings, the variation in film image O.D was within +/- 0.15 O.D of the normal density setting '0' and at +2 and +1 density setting, it was observed to be within +/- 0.30 O.D. This study indicates that the locally made PMMA TLD slot phantom can be used to measure various mammography QC parameters which are essentially required for better outcomes in mammography.
引用
收藏
页码:379 / 385
页数:7
相关论文
共 50 条
  • [21] Waste-to-Energy Systems Modelling Using In-House Developed Software
    Kropac, Jiri
    Pavlas, Martin
    Fusek, Michal
    Klimek, Petr
    Tous, Michal
    PRES 2011: 14TH INTERNATIONAL CONFERENCE ON PROCESS INTEGRATION, MODELLING AND OPTIMISATION FOR ENERGY SAVING AND POLLUTION REDUCTION, PTS 1 AND 2, 2011, 25 : 533 - +
  • [22] Nanostructured scintillator developed in-house for radon detection
    Abdalla, Ayman M.
    Almalki, Shaimaa
    Kawaguchi, Noriaki
    Yanagida, Takayuki
    RADIATION PHYSICS AND CHEMISTRY, 2022, 197
  • [23] Verification of a treatment planning system using an in-house designed head and neck phantom
    Akpochafor, M. O.
    Aweda, M. A.
    Adeneye, S. O.
    Iloputaife, C.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S139 - S140
  • [24] Image and dosimetry quality control for digital mammography by using the Normi Mam Digital phantom
    Calderon-Mata, Angelica
    Arce-Corrales, Luis
    Mejias-Gamboa, Ricardo
    UNICIENCIA, 2018, 32 (02) : 70 - 81
  • [25] An in-house developed resettable MOSFET dosimeter for radiotherapy
    Verellen, Dirk
    Van Vaerenbergh, Sven
    Tournel, Koen
    Heuninckx, Karina
    Joris, Laurent
    Duchateau, Michael
    Linthout, Nadine
    Gevaert, Thierry
    Reynders, Truus
    Van de Vondel, Iwein
    Coppens, Luc
    Depuydt, Tom
    De Ridder, Mark
    Storme, Guy
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (04): : N97 - N109
  • [26] AN IN-HOUSE DEVELOPED RESETTABLE MOSFET DOSIMETER FOR RADIOTHERAPY
    Van Vaerenbergh, S.
    Verellen, D.
    Tournel, K.
    Heuninckx, K.
    Duchateau, M.
    Linthout, N.
    Gevaert, T.
    Reynders, T.
    Van de Vondel, I.
    Coppens, L.
    Depuydt, T.
    Coppens, L.
    Storme, G.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S468 - S468
  • [27] IN-HOUSE DEVELOPED METHODOLOGIES & TOOLS FOR DECOMMISSIONING PROJECTS
    Detilleux, Michel
    Centner, Baudouin
    ICEM2007: PROCEEDINGS OF THE 11TH INTERNATIONAL CONFERENCE ON ENVIRONMENTAL REMEDIATION AND RADIOACTIVE WASTE MANAGEMENT, PTS A AND B, 2009, : 1109 - 1117
  • [28] A quality assurance program for an amorphous silicon electronic portal imaging device using in-house developed phantoms: a method development for dosimetry purposes
    Jomehzadeh, A.
    Shokrani, P.
    Mohammadi, M.
    Amouheidari, A.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2014, 12 (03): : 257 - 264
  • [29] An In-House Developed MOSFET Dosimeter with Reset Capabilities
    Verellen, D.
    Van Vaerenbergh, S.
    Tournel, K.
    Duchateau, M.
    Linthout, N.
    Gevaert, T.
    Reynders, T.
    Van De Vondel, I.
    Coppens, L.
    Depuydt, T.
    Storme, G.
    MEDICAL PHYSICS, 2010, 37 (06) : 3271 - +
  • [30] An In-House Developed Resettable MOSFET Dosimeter for Radiotherapy
    Van Vaerenbergh, S.
    Verellen, D.
    Van De Vondel, I.
    Heuninckx, K.
    Tournel, K.
    Reynders, T.
    Duchateau, M.
    Gevaert, T.
    Leysen, K.
    Depuydt, T.
    Poels, K.
    De Ridder, M.
    MEDICAL PHYSICS, 2011, 38 (06)