Real-Time Continuous Glucose Monitoring Significantly Reduces Severe Hypoglycemia in Hypoglycemia-Unaware Patients With Type 1 Diabetes

被引:105
|
作者
Choudhary, Pratik [1 ,3 ]
Ramasamy, Sharmin [2 ]
Green, Louisa [3 ]
Gallen, Geraldine [3 ]
Pender, Siobhan [2 ]
Brackenridge, Anna [2 ]
Amiel, Stephanie A. [1 ,3 ]
Pickup, John C. [1 ,2 ]
机构
[1] Kings Coll London, Div Diabet & Nutr Sci, London, England
[2] Guys & St Thomas NHS Fdn Trust, Dept Diabet, London, England
[3] Kings Coll Hosp London, Dept Diabet, London, England
关键词
NOCTURNAL HYPOGLYCEMIA; IMPAIRED AWARENESS; GLYCEMIC CONTROL; METAANALYSIS; FREQUENCY; RESPONSES; THERAPY; ADULTS;
D O I
10.2337/dc13-0939
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVETo evaluate the effect of continuous glucose monitoring (CGM) on the frequency of severe hypoglycemia (SH) in patients with established hypoglycemia unawareness.RESEARCH DESIGN AND METHODSWe conducted a retrospective audit of 35 patients with type 1 diabetes and problematic hypoglycemia unawareness, despite optimized medical therapy (continuous subcutaneous insulin infusion/multiple daily insulin injections), who used CGM for >1 year.RESULTSOver a 1-year follow-up period, the median rates of SH were reduced from 4.0 (interquartile range [IQR] 0.75-7.25) episodes/patient-year to 0.0 (0.0-1.25) episodes/patient-year (P < 0.001), and the mean (SD) rates were reduced from 8.1 +/- 13 to 0.6 +/- 1.2 episodes/year (P = 0.005). HbA(1c) was reduced from 8.1 +/- 1.2% to 7.6 +/- 1.0% over the year (P = 0.005). The mean Gold score, measured in 19 patients, did not change: 5.1 +/- 1.5 vs. 5.2 +/- 1.9 (P = 0.67).CONCLUSIONSIn a specialist experienced insulin pump center, in carefully selected patients, CGM reduced SH while improving HbA(1c) but failed to restore hypoglycemia awareness.
引用
下载
收藏
页码:4160 / 4162
页数:3
相关论文
共 50 条
  • [1] Real-time continuous glucose monitoring decreases the risk of severe hypoglycemia in people with type 1 diabetes and impaired awareness of hypoglycemia
    Lucidi, Paola
    Porcellati, Francesca
    Bolli, Geremia B.
    Fanelli, Carmine G.
    ANNALS OF TRANSLATIONAL MEDICINE, 2018, 6
  • [2] Effect of real-time continuous glucose monitoring on hypoglycemia in adult type 1 diabetes patients
    Jia, Weiping
    JOURNAL OF DIABETES INVESTIGATION, 2018, 9 (03): : 478 - 480
  • [3] Annual Cost to Achieve Cost Effectiveness of Continuous Glucose Monitoring in Hypoglycemia-Unaware Type 1 Diabetes Patients Using Intensive Insulin in the United Kingdom
    Bilir, S. Pinar
    Campbell, Caresse
    Hellmund, Richard
    Munakata, Julie
    DIABETES, 2019, 68
  • [4] Real-Time Hypoglycemia Detection from Continuous Glucose Monitoring Data of Subjects with Type 1 Diabetes
    Jensen, Morten Hasselstrom
    Christensen, Toke Folke
    Tarnow, Lise
    Seto, Edmund
    Johansen, Mette Dencker
    Hejlesen, Ole Kristian
    DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (07) : 538 - 543
  • [5] Risk factors of severe hypoglycemia and occurrence of silent hypoglycemia in patients with type 1 diabetes assessed by continuous glucose monitoring
    Hoi-Hansen, T
    Pedersen-Bjergaard, U
    Pramming, S
    Thorsteinsson, B
    DIABETES, 2005, 54 : A153 - A154
  • [6] Improving Epinephrine Responses in Hypoglycemia Unawareness With Real-Time Continuous Glucose Monitoring in Adolescents With Type 1 Diabetes
    Ly, Trang T.
    Hewitt, Jacqueline
    Davey, Raymond J.
    Lim, Ee Mun
    Davis, Elizabeth A.
    Jones, Timothy W.
    DIABETES CARE, 2011, 34 (01) : 50 - 52
  • [7] Nocturnal Continuous Glucose Monitoring: Accuracy and Reliability of Hypoglycemia Detection in Patients with Type 1 Diabetes at High Risk of Severe Hypoglycemia
    Bay, Christiane
    Kristensen, Peter Lommer
    Pedersen-Bjergaard, Ulrik
    Tarnow, Lise
    Thorsteinsson, Birger
    DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (05) : 371 - 377
  • [8] Real-Time Continuous Glucose Monitoring Can Predict Severe Hypoglycemia in People with Type 1 Diabetes: Combined Analysis of the HypoDE and DIAMOND Trials
    Hermanns, Norbert
    Ehrmann, Dominic
    Heinemann, Lutz
    Freckmann, Guido
    Waldenmaier, Delia
    Calhoun, Peter
    DIABETES TECHNOLOGY & THERAPEUTICS, 2022, 24 (09) : 603 - 610
  • [9] Effect of Continuous Glucose Monitoring on Hypoglycemia in Type 1 Diabetes
    Battelino, Tadej
    Phillip, Moshe
    Bratina, Natasa
    Nimri, Revital
    Oskarsson, Per
    Bolinder, Jan
    DIABETES CARE, 2011, 34 (04) : 795 - 800
  • [10] Associations Between the Time in Hypoglycemia and Hypoglycemia Awareness Status in Type 1 Diabetes Patients Using Continuous Glucose Monitoring Systems
    Lin, Yu Kuei
    Groat, Danielle
    Chan, Owen
    Hung, Man
    Sharma, Anu
    Varner, Michael W.
    Gouripeddi, Ramkiran
    Facelli, Julio C.
    Fisher, Simon J.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2020, 22 (11) : 787 - 793