Machine-specific quality assurance procedure for stereotactic treatments with dynamic couch rotations

被引:5
|
作者
Wilson, Byron [1 ,2 ]
Gete, Ermias [1 ]
机构
[1] BC Canc Agcy, Vancouver Ctr, Med Phys, Vancouver, BC, Canada
[2] Univ British Columbia, Phys & Astron, Vancouver, BC, Canada
关键词
quality assurance; SABR; SRS; VMAT; MODULATED ARC THERAPY; PARTIAL BREAST IRRADIATION; LINEAR-ACCELERATOR; SINGLE-ISOCENTER; BRAIN METASTASES; PLAN QUALITY; RADIOSURGERY; DELIVERY; MOTION; FEASIBILITY;
D O I
10.1002/mp.12589
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: We present a method in which the treatment couch's accuracy is measured using the electronic portal imaging device (EPID) and a phantom of our own construction. Using this phantom, we were able to quantify the treatment couch walkout, and the rotation angle accuracy for both static and dynamic couch treatments. These measurements were used to provide an accurate measure of the treatment couch isocenter as well as to verify the couch rotation angle recorded in the trajectory log. Methods: The phantom was constructed using a polystyrene slab in which five ball bearings of 4 mm diameter are placed on the same plane at varying radii (0, 2.8, 4.4, 5.6, and 6.7 cm). The couch was rotated through its full extent (-90, 90 degrees) while MV images were acquired continuously. The couch rotational accuracy was calculated using a least squares minimization which fit the locations of the BBs to their expected locations relative to reference setup conditions. Using this approach, rotation angle and isocenter walkout was calculated in three dimensions. These measurements were used to quantify the accuracy of the couch as well as to validate the Varian TrueBeam trajectory logs. Additionally, a method for an EPID-based couch star-shot measurement was developed and compared with the traditional film-based method. Results: The measured couch center of rotation consisted of a cloud of points clustered around the room isocenter within 0.7 mm distance. The trajectory log couch angle values agreed with those recorded in the DICOM header of the EPID images to the third significant digit and the couch rotation angles recorded in the trajectory log and DICOM header agreed with the calculated values to 0.08 degrees. Comparison of couch star-shot measurement developed in this study with film-based star-shot measurements gave an agreement to within 0.2 mm. Conclusion: We have developed a quality assurance method for the treatment couch which is simple, accurate, and enables the user to access a multitude of consistent data with a single measurement. Using this method, we have shown that the treatment couch is accurate for both static and dynamic stereotactic deliveries. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:6529 / 6537
页数:9
相关论文
共 50 条
  • [1] Machine Specific Quality Assurance Procedure for Stereotactic Treatments with Dynamic Couch Rotations
    Wilson, B.
    Gete, E.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2955 - 2955
  • [2] Quality assurance of imaging for stereotactic treatments
    Novotny, J
    Novotny, JJ
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S95 - S95
  • [3] SGRT quality assurance for stereotactic treatments
    Rodriguez Romero, R.
    Lopez Corella, A.
    Torres Lopez, M.
    Pinto Monedero, M.
    Sanchez Rubio, P.
    Martinez Ortega, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1471 - S1473
  • [4] Comprehensive quality assurance for stereotactic radiosurgery treatments
    Ramaseshan, R
    Heydarian, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14): : N199 - N205
  • [5] Dose quality assurance for stereotactic radiotherapy treatments
    Perks, J
    Rosenberg, I
    Warrington, J
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (10): : N209 - N215
  • [6] QUALITY ASSURANCE PROCEDURE FOR SRS AND SRT TREATMENTS
    Meijers, Arturs
    Plaude, Sandija
    Utehina, Olga
    Popovs, Sergejs
    [J]. MEDICAL PHYSICS IN THE BALTIC STATES, 2011, : 69 - 72
  • [7] An Unified Imaging and Robotic Couch Quality Assurance Procedure, and Automation Software
    Schreibmann, E.
    Elder, E.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3357 - 3357
  • [8] Simplified "on-couch" daily quality assurance procedure for CT simulators
    Liu, Ruijie Rachel
    Prado, Karl
    Gillin, Michael
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (03): : 49 - 55
  • [9] Development and clinical implementation of a quality assurance (QA) procedure for automated couch movement
    Li, W.
    Moseley, D.
    Nordinger, B.
    Foxcroft, S.
    Smale, J.
    Affray, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S13 - S14
  • [10] A CT daily quality assurance on-couch procedure and phantom in radiation oncology
    Liu, R.
    Prado, K.
    Wu, R.
    Balter, P.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2350 - 2350