EPID-based verification of the MLC performance for dynamic IMRT and VMAT

被引:47
|
作者
Rowshanfarzad, Pejman [1 ]
Sabet, Mahsheed [2 ]
Barnes, Michael P. [2 ]
O'Connor, Daryl J. [1 ]
Greer, Peter B. [1 ,2 ]
机构
[1] Univ Newcastle, Sch Math & Phys Sci, Newcastle, NSW 2308, Australia
[2] Calvary Mater Newcastle Hosp, Dept Radiat Oncol, Newcastle, NSW 2310, Australia
基金
英国医学研究理事会;
关键词
linac QA; MLC performance; VMAT; IMRT; EPID; LEAF POSITION ACCURACY; PORTAL IMAGING DEVICE; INTENSITY-MODULATED RADIOTHERAPY; MULTILEAF COLLIMATOR; QUALITY-ASSURANCE; CLINICAL IMPLEMENTATION; LOG FILES; DELIVERY; ARC; THERAPY;
D O I
10.1118/1.4752207
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In advanced radiotherapy treatments such as intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT), verification of the performance of the multileaf collimator (MLC) is an essential part of the linac QA program. The purpose of this study is to use the existing measurement methods for geometric QA of the MLCs and extend them to more comprehensive evaluation techniques, and to develop dedicated robust algorithms to quantitatively investigate the MLC performance in a fast, accurate, and efficient manner. Methods: The behavior of leaves was investigated in the step-and-shoot mode by the analysis of integrated electronic portal imaging device (EPID) images acquired during picket fence tests at fixed gantry angles and arc delivery. The MLC was also studied in dynamic mode by the analysis of cine EPID images of a sliding gap pattern delivered in a variety of conditions including different leaf speeds, deliveries at fixed gantry angles or in arc mode, and changing the direction of leaf motion. The accuracy of the method was tested by detection of the intentionally inserted errors in the delivery patterns. Results: The algorithm developed for the picket fence analysis was able to find each individual leaf position, gap width, and leaf bank skewness in addition to the deviations from expected leaf positions with respect to the beam central axis with sub-pixel accuracy. For the three tested linacs over a period of 5 months, the maximum change in the gap width was 0.5 mm, the maximum deviation from the expected leaf positions was 0.1 mm and the MLC skewness was up to 0.2 degrees. The algorithm developed for the sliding gap analysis could determine the velocity and acceleration/deceleration of each individual leaf as well as the gap width. There was a slight decrease in the accuracy of leaf performance with increasing leaf speeds. The analysis results were presented through several graphs. The accuracy of the method was assessed as 0.01 mm for both the gap size and peak position determination. Conclusions: This study provides fast, easy, and accurate test methods for routine QA of the MLC performance and helps in faster troubleshooting of MLC problems in both IMRT and VMAT treatments. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4752207]
引用
收藏
页码:6192 / 6207
页数:16
相关论文
共 50 条
  • [41] Establishing Institution-Specific Action Levels for EPID-Based QA for IMRT
    Gray, A.
    Saldan, A.
    Rosenzweig, D.
    Meiler, R.
    Clark, D.
    Abu-Aita, R.
    Schell, M.
    Cavanaugh, D.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [42] Is Weekly MLC QA Necessary? Two Year EPID-Based Weekly MLC QA Experience at the University of Maryland
    Kalavagunta, C.
    Yang, X.
    Xu, H.
    Zhang, B.
    Mossahebi, S.
    Sawant, A.
    Yi, B.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3753 - 3753
  • [43] A novel EPID-based MLC QA method with log files achieving submillimeter accuracy
    Liu, Chenlu
    Wang, Binbing
    Bai, Xue
    Cheng, Xiaolong
    Wang, Xiaotong
    Yang, Xiaohua
    Shan, Guoping
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (08):
  • [44] Sensitivity analysis of EPID-based 3D dose reconstruction for VMAT QA
    Alhazmi, A.
    Gianoli, C.
    Neppl, S.
    Martins, J.
    Veloza, S.
    Podesta, M.
    Verhaegen, F.
    Reiner, M.
    Belka, C.
    Parodi, K.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S960 - S961
  • [45] Modulation factors calculated with an EPID-derived MLC fluence model to streamline IMRT/VMAT second checks
    Steciw, Stephen
    Rathee, Satyapal
    Warkentin, Brad
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (06): : 62 - 81
  • [46] Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries
    Agnew, A.
    Agnew, C. E.
    Grattan, M. W. D.
    Hounsell, A. R.
    McGarry, C. K.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (09): : N49 - N63
  • [47] Clinical results of an EPID-based in-vivo dosimetry for prostate cancer treated by VMAT
    Falco, M. D.
    Giancaterino, S.
    De Nicola, A.
    Perrotti, F.
    Menna, S.
    Fidanzio, A.
    Piermattei, A.
    Genovesi, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S706 - S706
  • [48] An EPID-Based Method for Testing Absolute Leaf Position for MLC Without Backup Jaws
    Hancock, S.
    Whitaker, M.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 229 - 229
  • [49] Dose Verification in IMRT and VMAT
    Feygelman, Vladimir
    Nelms, Benjamin E.
    [J]. CONCEPTS AND TRENDS IN MEDICAL RADIATION DOSIMETRY, 2011, 1345 : 145 - +
  • [50] A Viable Approach to Patient Specific QA for Spine VMAT SRS Using EPID-Based Dosimetry
    Song, Y.
    Zhang, P.
    Li, J.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3742 - 3742