Cortical bone drilling and thermal osteonecrosis

被引:282
|
作者
Augustin, Goran [1 ,2 ]
Zigman, Tomislav [2 ]
Davila, Slavko [2 ]
Udilljak, Toma [3 ]
Staroveski, Tomislav [3 ]
Brezak, Danko [3 ]
Babic, Slaven [4 ]
机构
[1] Univ Zagreb, Dept Surg, Univ Hosp, Ctr Zagreb, Zagreb 10000, Croatia
[2] Univ Zagreb, Sch Med, Zagreb 10000, Croatia
[3] Univ Zagreb, Fac Mech Engn & Naval Architecture, Zagreb 10000, Croatia
[4] Clin Hosp Ctr Sestre Milosrdnice, Zagreb 10000, Croatia
关键词
Bone; Drill; Drilling; Parameters; Temperature; Thermal osteonecrosis; FEMORAL-HEAD OSTEONECROSIS; IMPLANT SITE PREPARATION; TEMPERATURE-MEASUREMENT; PIN INSERTION; SPEED; GENERATION; OSSEOINTEGRATION; CRYOSURGERY; IRRIGATION; PARAMETERS;
D O I
10.1016/j.clinbiomech.2011.10.010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Bone drilling is a common step in operative fracture treatment and reconstructive surgery. During drilling elevated bone temperature is generated. Temperatures above 47 degrees C cause thermal osteonecrosis which contributes to screw loosening and subsequently implant failures and refractures. Methods: The current literature on bone drilling and thermal osteonecrosis is reviewed. The methodologies involved in the experimental and clinical studies are described and compared. Findings: Areas which require further investigation are highlighted and the potential use of more precise experimental setup and future technologies are addressed. Interpretation: Important drill and drilling parameters that could cause increase in bone temperature and hence thermal osteonecrosis are reviewed and discussed: drilling speed, drill feed rate, cooling, drill diameter, drill point angle, drill material and wearing, drilling depth, pre-drilling, drill geometry and bone cortical thickness. Experimental methods of temperature measurement during bone drilling are defined and thermal osteonecrosis is discussed with its pathophysiology, significance in bone surgery and methods for its minimization. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:313 / 325
页数:13
相关论文
共 50 条
  • [41] A Model for the Prediction of Thermal Response of Bone in Surgical Drilling
    Maani, Nazanin
    Farhang, Kambiz
    Hodaei, Mohammad
    JOURNAL OF THERMAL SCIENCE AND ENGINEERING APPLICATIONS, 2014, 6 (04)
  • [42] INFLUENCE OF BONE DRILLING PARAMETERS ON THE THERMAL STRESS DISTRIBUTION
    Fernandes, Maria G.
    Fonseca, Elza M.
    Jorge, Renato N.
    IRF2016: 5TH INTERNATIONAL CONFERENCE INTEGRITY-RELIABILITY-FAILURE, 2016, : 517 - 528
  • [43] Optimization of drilling parameters for thermal bone necrosis prevention
    Akhbar, Mohd Faizal Ali
    Yusoff, Ahmad Razlan
    TECHNOLOGY AND HEALTH CARE, 2018, 26 (04) : 621 - 635
  • [44] An experimental investigation on thermal exposure during bone drilling
    Lee, JuEun
    Ozdoganlar, O. Burak
    Rabin, Yoed
    MEDICAL ENGINEERING & PHYSICS, 2012, 34 (10) : 1510 - 1520
  • [45] Measurement of thermal conductivity of bovine cortical bone
    Davidson, SRH
    James, DF
    MEDICAL ENGINEERING & PHYSICS, 2000, 22 (10) : 741 - 747
  • [46] TREATMENT OF OSTEONECROSIS OF THE FEMORAL-HEAD BY DRILLING AND MUSCLE-PEDICLE BONE-GRAFTING
    BAKSI, DP
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (02): : 241 - 245
  • [47] The trapdoor procedure using autogenous cortical and cancellous bone grafts for osteonecrosis of the femoral head
    Mont, MA
    Einhorn, TA
    Sponseller, PD
    Hungerford, DS
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (01): : 56 - 62
  • [48] Drilling Forces and Temperature Rises in Axial Low-frequency Vibration-assisted Cortical Bone Drilling
    Bai X.
    Hou S.
    Li K.
    Qu Y.
    Zhongguo Jixie Gongcheng/China Mechanical Engineering, 2021, 32 (03): : 321 - 330
  • [49] Investigation of transient machining in the cortical bone drilling process by conventional and axial vibration-assisted drilling methods
    Bai, Xiaofan
    Qiao, Guochao
    Liu, Zhiqiang
    Zhu, Weidong
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2023, 237 (04) : 489 - 501
  • [50] Modeling and verification of cortical bone drilling forces based on tissue structure heterogeneity
    Song, Shaokang
    Zhao, Jun
    Sun, Xianshun
    Li, Anhai
    Lu, Yongliang
    Zhang, Shihua
    CIRP JOURNAL OF MANUFACTURING SCIENCE AND TECHNOLOGY, 2024, 54 : 63 - 74