Retinal Laser Lesion Visibility in Simultaneous Ultra-High Axial Resolution Optical Coherence Tomography

被引:10
|
作者
Steiner, Patrick [1 ,2 ]
Enzmann, Volker [3 ]
Meier, Christoph [2 ]
Povazay, Boris [2 ]
Kowal, Jens H. [1 ,3 ]
机构
[1] Univ Bern, ARTORG Ctr, CH-3010 Bern, Switzerland
[2] Berne Univ Appl Sci, HuCE OptoLab, CH-2502 Biel, Switzerland
[3] Inselspital Bern, Dept Ophthalmol, CH-3010 Bern, Switzerland
来源
IEEE PHOTONICS JOURNAL | 2014年 / 6卷 / 06期
关键词
Retinal laser therapy; photocoagulation; optical coherence tomography; histology;
D O I
10.1109/JPHOT.2014.2374594
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Ex vivo porcine retina laser lesions applied with varying laser power (20 mW-2 W, 10 ms pulse, 196 lesions) are manually evaluated by microscopic and optical coherence tomography (OCT) visibility, as well as in histological sections immediately after the deposition of the laser energy. An optical coherence tomography system with 1.78 mu m axial resolution specifically developed to image thin retinal layers simultaneously to laser therapy is presented, and visibility thresholds of the laser lesions in OCT data and fundus imaging are compared. Optical coherence tomography scans are compared with histological sections to estimate the resolving power for small optical changes in the retinal layers, and real-time time-lapse scans during laser application are shown and analyzed quantitatively. Ultrahigh-resolution OCT inspection features a lesion visibility threshold 40-50 mW (17% reduction) lower than for visual inspection. With the new measurement system, 42% of the lesions that were invisible using state-of-the-art ophthalmoscopic methods could be detected.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Ultra-High Resolution Fiber Based Polarization Sensitive Spectral Domain Optical Coherence Tomography
    Goetzinger, Erich
    Pircher, Michael
    Baumann, Bernhard
    Hitzenberger, Christoph K.
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICINE XIV, 2010, 7554
  • [32] Ultra-High Resolution Optical Coherence Tomography (OCT) Using a Halogen Lamp as the Light Source
    Masato Ohmi
    Masamitsu Haruna
    Optical Review, 2003, 10 : 478 - 481
  • [33] Photoreceptor Lengthening During Dark Adaptation Detected With Ultra-High Resolution Optical Coherence Tomography
    Lu, Chen D.
    Choi, Woo Jhon
    Wang, Zhao
    Kraus, Martin F.
    Hornegger, Joachim
    Duker, Jay S.
    Fujimoto, James G.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)
  • [34] A comparative study of noise in supercontinuum light sources for ultra-high resolution Optical Coherence Tomography
    Maria, M.
    Gonzalo, I. B.
    Bondu, M.
    Engelsholm, R. D.
    Feuchter, T.
    Moselund, P. M.
    Leick, L.
    Bang, O.
    Podoleanu, A.
    DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES X, 2017, 10056
  • [35] Glass-clad Crystal Fiber based Ultra-high Resolution Optical Coherence Tomography
    Wang, Y. T.
    Tsai, C. C.
    Hsu, K. Y.
    Lin, Y. S.
    Jheng, D. Y.
    Huang, K. Y.
    Huang, S. L.
    Yeh, P. S.
    2009 LASERS & ELECTRO-OPTICS & THE PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, VOLS 1 AND 2, 2009, : 1293 - 1293
  • [36] Measurement of the retention time of different ophthalmic formulations with ultra-high resolution Optical Coherence Tomography
    Amato, Roberta
    Gagliano, Caterina
    Malaguarnera, Giulia
    Reibaldi, Michele
    Longo, Antonio
    Papa, Vincenzo
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [37] Ultra-high resolution spectral domain optical coherence tomography using supercontinuum light source
    Yiheng Lim
    Toyohiko Yatagai
    Yukitoshi Otani
    Optical Review, 2016, 23 : 180 - 186
  • [38] Ultra-high resolution spectral domain optical coherence tomography using supercontinuum light source
    Lim, Yiheng
    Yatagai, Toyohiko
    Otani, Yukitoshi
    OPTICAL REVIEW, 2016, 23 (02) : 180 - 186
  • [39] En-face scanning optical coherence tomography with ultra-high resolution for material investigation
    Wiesauer, K
    Pircher, M
    Götzinger, E
    Bauer, S
    Engelke, R
    Ahrens, G
    Grützner, G
    Hitzenberger, CK
    Stifter, D
    OPTICS EXPRESS, 2005, 13 (03): : 1015 - 1024
  • [40] Ultra-low noise supercontinuum source for ultra-high resolution optical coherence tomography at 1300 nm
    Gonzalo, I. B.
    Maria, M.
    Engelsholm, R. D.
    Feuchter, T.
    Leick, L.
    Moselund, P. M.
    Podoleanu, A.
    Bang, O.
    DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES XI, 2018, 10486