Theoretical study on temporal and spatial performance of magnetic solenoid used in dilation x-ray imager

被引:10
|
作者
Wang, Qiangqiang [1 ]
Cao, Zhurong [1 ]
Chen, Tao [1 ]
Deng, Keli [1 ]
Deng, Bo [1 ]
机构
[1] China Acad Engn Phys, Res Ctr Laser Fus, Mianyang 621900, Sichuan, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2020年 / 91卷 / 07期
基金
中国国家自然科学基金;
关键词
SECONDARY-ELECTRON EMISSION; DISTORTION; LENSES; SHAPE;
D O I
10.1063/1.5133395
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In the Dilation X-ray Imager (DIXI), which is characterized by an ultra-short gating time, a magnetic solenoid is used to keep the photoelectrons from defocusing during the drift process. The performance of the magnetic solenoid has an important influence on the performance of the DIXI. We present in this paper the efforts on studying the spatial and temporal performance of the magnetic solenoid used in the DIXI by tracking the photoelectrons with the particle-in-cell method. The initial parameters of the photoelectrons of the Au cathode were sampled with a Monte Carlo code. A novel magnetic solenoid with a shielding shell made of soft iron was proposed. We compared the performance of this solenoid with a normal solenoid. The simulation results of magnetic field distribution, spatial resolution, transit time spread, and temporal distortion were presented in detail to demonstrate that the performance of the DIXI can be significantly improved by using the magnetic solenoid with the proposed iron shell.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Marshall Grazing Incidence X-ray Spectrometer Slitjaw Imager Implementation and Performance
    Vigil, Genevieve D.
    Winebarger, Amy R.
    Kobayashi, Ken
    Cheimets, Peter N.
    Champey, Patrick R.
    Savage, Sabrina L.
    Golub, Leon
    Watkinson, Benjamin
    Beabout, Brent
    Weddendorf, Bruce
    Walsh, Robert W.
    SOLAR PHYSICS, 2021, 296 (06)
  • [32] Theoretical analysis of the spectra of X-ray resonant magnetic reflectivity
    Andreeva, M. A.
    Smekhova, A. G.
    APPLIED SURFACE SCIENCE, 2006, 252 (15) : 5619 - 5621
  • [33] Theoretical study of X-ray photoemission, X-ray absorption and resonant X-ray emission spectroscopy of Mn films on Ag
    Taguchi, M.
    Kueger, P.
    Parlebas, J. C.
    Kotani, A.
    PHYSICA SCRIPTA, 2005, T115 : 122 - 124
  • [34] Improving the spatial resolution of a soft X-ray Charge Coupled Device used for Resonant Inelastic X-ray Scattering
    Soman, M. R.
    Hall, D. J.
    Tutt, J. H.
    Murray, N. J.
    Holland, A. D.
    Schmitt, T.
    Raabe, J.
    Schmitt, B.
    JOURNAL OF INSTRUMENTATION, 2011, 6
  • [35] Role of the activator in the performance of scintillators used in X-ray imaging
    Kandarakis, I
    Cavouras, D
    APPLIED RADIATION AND ISOTOPES, 2001, 54 (05) : 821 - 831
  • [36] Performance Assessment of X-Ray QA Instruments Used in Mammography
    Chu, C. H.
    Chen, I. J.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [37] SPATIAL AND TEMPORAL X-RAY DIAGNOSES OF LASER-PRODUCED PLASMA
    GEX, JP
    FLEUROT, N
    SAUNEUF, R
    REVUE DE PHYSIQUE APPLIQUEE, 1977, 12 (07): : 1049 - 1053
  • [38] Investigation into the influence of x-ray scatter on the imaging performance of an x-ray flat panel imager based cone beam volume CT
    Tang, XY
    Ning, R
    Yu, RF
    Conover, D
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 851 - 860
  • [39] STUDY ON MAGNETIC STARS AND X-RAY STARS
    DAUVILLIER, A
    BULLETIN DE LA CLASSE DES SCIENCES ACADEMIE ROYALE DE BELGIQUE, 1974, 60 (09): : 1060 - 1069
  • [40] Performance of Magnetic Penetration Thermometers for X-ray Astronomy
    Nagler, P. C.
    Adams, J. S.
    Balvin, M. A.
    Bandler, S. R.
    Denis, K. L.
    Hsieh, W. -T.
    Kelly, D. P.
    Porst, J. -P.
    Sadleir, J. E.
    Seidel, G. M.
    Smith, S. J.
    Stevenson, T. R.
    JOURNAL OF LOW TEMPERATURE PHYSICS, 2012, 167 (3-4) : 455 - 460