Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging.

被引:1
|
作者
Antonio, Debora [1 ,2 ,3 ]
Di Maria, Salvatore [2 ]
Matela, Nuno [3 ]
Vieira, Sandra [4 ]
Vaz, Pedro [2 ]
机构
[1] Univ Nova Lisboa, Fac Ciencias & Tecnol, Dept Fis, P-2829516 Quinta Da Torre, Caparica, Portugal
[2] Univ Lisbon, Ctr Ciencias & Tecnol Nucl, Inst Super Tecn, Estr Nacl 10,Km139,7, P-2695066 Bobadela, Portugal
[3] Univ Lisbon, Fac Ciencias, Inst Biofis & Engn Biomed, P-1749016 Lisbon, Portugal
[4] Ctr Clin Champalimaud, Radioncol, Ave Brasilia, P-1400038 Lisbon, Portugal
关键词
Cone beam computed tomography; Cancer detection; Monte Carlo simulation; Organ dose; Medical image reconstruction; Lung and breast imaging; X-RAY-SPECTRA; CANCER; TOMOSYNTHESIS;
D O I
10.1016/j.radphyschem.2020.108972
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Cancer is the second leading cause of death in the world, and therefore, there is an undeniable need to ensure early screening and detection systems worldwide. The aim of this project was to study the feasibility of a Cone Beam Computed Tomography (CBCT) scanner for simultaneous breast and lung lesion imaging. Additionally, the development of reconstruction algorithms and the study of their impact to the image quality was considered. Monte Carlo (MC) simulations were performed using the PENELOPE code system. A geometry model of a CBCT scanner was implemented for energies of 30 keV and 80 keV for hypothetical scanning protocols. Microcalcifications were inserted into the breast and lung of the computational phantom (ICRP Adult Female Reference), used in the simulations for dose assessment and projection acquisition. Dosimetric and imaging performances were evaluated through Computed Tomography Dose Index and Catphan phantoms, respectively. Reconstructed images were analyzed in terms of Contrast-to-Noise Ratio (CNR) and dose calculations were performed for two protocols, using a normalization factor of 2 mGy in the breast and another with 5 mGy in the lungs. MC geometry model and reconstruction algorithm were validated by means of on-field measurements and data acquisition in a clinical center. Results indicate that for both protocols, the absorbed dose in both organs is the same, allowing the optimization of protocol strategies regarding radiosensitive organs. The best implementation of the reconstruction algorithm was achieved with 80 keV, using linear interpolation and hanning filter. More specifically for a spherical lung lesion (r = 7 mm) a 30% CNR gain was found when the number of projections varied from 12 to 36 (corresponding to a dose increase of a factor of 3). This study suggests the applicability of a CBCT modulated beam scanner for imaging simultaneously breast and lung lesions while ensuring dose reduction without compromising image quality as a possibility.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Design of CT reconstruction kernel specifically for clinical lung imaging
    Cody, DD
    Hsieh, J
    Gladish, GW
    MEDICAL IMAGING 2005: PHYSIOLOGY, FUNCTION, AND STRUCTURE FROM MEDICAL IMAGES, PTS 1 AND 2, 2005, 5746 : 101 - 108
  • [42] Task-based optimization of image reconstruction in breast CT
    Sanchez, Adrian A.
    Sidky, Emil Y.
    Pan, Xiaochuan
    MEDICAL IMAGING 2014: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2014, 9037
  • [43] A reconstruction algorithm for helical CT imaging on PI-planes
    Liang, Hongzhu
    Zhang, Cishen
    Yan, Ming
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 1090 - +
  • [44] Reconstruction algorithm effects on CT imaging measures in emphysema patients
    Boedeker, KL
    McNitt-Gray, MF
    Truong, DA
    Rogers, SR
    Gjertson, DW
    Goldin, JG
    RADIOLOGY, 2002, 225 : 547 - 547
  • [45] An Iterative CT Reconstruction Algorithm for Fast Fluid Flow Imaging
    Van Eyndhoven, Geert
    Batenburg, K. Joost
    Kazantsev, Daniil
    Van Nieuwenhove, Vincent
    Lee, Peter D.
    Dobson, Katherine J.
    Sijbers, Jan
    IEEE TRANSACTIONS ON IMAGE PROCESSING, 2015, 24 (11) : 4446 - 4458
  • [46] Deployment of a dose-reduction algorithm in Molecular breast Imaging - impact of reader experience on its optimization
    Stanke, Ashlee
    O'Connor, Michael
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [47] Evaluation and Optimization of a New PET Reconstruction Algorithm, Bayesian Penalized Likelihood Reconstruction, for Lung Cancer Assessment According to Lesion Size
    Otani, Tomoaki
    Hosono, Makoto
    Kanagaki, Mitsunori
    Onishi, Yasuyuki
    Matsubara, Naoko
    Kawabata, Kazuna
    Kimura, Hiroyuki
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2019, 213 (02) : W50 - W56
  • [48] Utility of FDG-PET in the evaluation of recurrence breast cancer. A comparison with CT imaging.
    Benitez, A.
    Albala, M. D.
    Maza, R.
    Contreras, P. I.
    Pacheco, C.
    Latre, J. M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 : S417 - S417
  • [49] A compressed sensing based iterative reconstruction algorithm for CT dose reduction
    Hsieh, Chia-Jui
    Chiang, Huihua Kenny
    Chiu, Yung-Hsiang
    Xiao, Bo-Wen
    Sun, Cheng-Wei
    Yeh, Ming-Hua
    Yeh, Ming-Hua
    Chen, Jvh-cheng
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [50] Reducing the dose of CT of the paranasal sinuses: potential of an iterative reconstruction algorithm
    Schaafs, Lars-Arne
    Lenk, Julian
    Hamm, Bernd
    Niehues, Stefan Markus
    DENTOMAXILLOFACIAL RADIOLOGY, 2016, 45 (07)