To get physical insight into the 3D transfer characteristics of interference microscopy at high numerical apertures we study reflecting rectangular grating structures. In general, the height obtained from phase information seems to be reduced, whereas height values resulting from coherence scanning sometimes seem to be systematically overestimated. Increasing the numerical aperture of an interference microscope broadens the spectra of the resulting interference signals, thus offering a broad variety of wavelength contributions to be analyzed. If phase analysis of a measured far-field interference wavefront is performed at very short wavelengths the periodical profiles obtained from coherence scanning and phase shifting analysis differ only by the measured amplitude. However, at longer wavelength there is a 180 degrees phase shift of the measured profiles obtained from phase analysis compared to coherence peak analysis. Increasing the evaluation wavelength improves the lateral resolution since the long wavelength contributions are related to electromagnetic waves of high angles of incidence. This behavior is to the best of our knowledge not documented in literature so far. It was first observed experimentally and could be confirmed by simulation results obtained from either Kirchhoff diffraction theory or an extended Richards-Wolf model developed in our group. Compared to original input profiles used for the simulation the profiles obtained from phase evaluation correspond quite well at longer wavelength, whereas the results obtained from coherence peak analysis are typically inverted with respect to height. (C) 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
引用
收藏
页码:7376 / 7389
页数:14
相关论文
共 50 条
  • [1] Beyond the lateral resolution limit by phase imaging
    Cotte, Y.
    Toy, M. Fatih
    Depeursinge, C.
    [J]. THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING XVIII, 2011, 7904
  • [2] Beyond the lateral resolution limit by phase imaging
    Cotte, Yann
    Toy, M. Fatih
    Depeursinge, Christian
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2011, 16 (10)
  • [3] Breaking diffraction limit of lateral resolution in optical coherence tomography
    Wang, Benquan
    Lu, Rongwen
    Zhang, Qiuxiang
    Yao, Xincheng
    [J]. QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2013, 3 (05) : 243 - 248
  • [4] Lateral spatial resolution improvement in laser scanning fluorescence microscopy using a subdiffraction limit optical spot
    Nishimura, Takahiro
    Shinkawa, Daiki
    Ogura, Yusuke
    Tamada, Yosuke
    Tanida, Jun
    [J]. BIOMEDICAL IMAGING AND SENSING CONFERENCE, 2018, 10711
  • [5] Optical coherence microscopy for high resolution biological imaging
    Vabre, L
    Dubois, A
    Beaurepaire, E
    Boccara, C
    [J]. PHOTON MIGRATION, DIFFUSE SPECTROSCOPY AND OPTICAL COHERENCE TOMOGRAPHY: IMAGING AND FUNCTIONAL ASSESSMENT, 2000, 4160 : 24 - 30
  • [6] 3-D Optical Interference Microscopy at the Lateral Resolution
    Lehmann, Peter
    Niehues, Jan
    Tereschenko, Stanislav
    [J]. INTERNATIONAL JOURNAL OF OPTOMECHATRONICS, 2014, 8 (04) : 231 - 241
  • [7] Improving lateral resolution of optical coherence tomography for imaging of skins
    Shen, Kai
    Lu, Hui
    Wang, Michael R.
    [J]. THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING XXIII, 2016, 9713
  • [8] High-resolution line-scanning optical coherence microscopy
    Chen, Yu
    Huang, Shu-Wei
    Aguirre, Aaron D.
    Fujimoto, James G.
    [J]. OPTICS LETTERS, 2007, 32 (14) : 1971 - 1973
  • [9] Transverse priority scanning and microscopy for high resolution optical coherence tomography
    Hsiung, P
    Aguirre, AD
    Ko, TH
    Bourquin, S
    Schneider, K
    Hartl, I
    Fujimoto, JG
    [J]. OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE TECHNIQUES, 2003, 5140 : 110 - 112
  • [10] The lateral resolution limit for imaging periodic conducting surfaces in capacitive microscopy
    Bruce, NC
    García-Valenzuela, A
    Kouznetsov, D
    [J]. JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2000, 33 (22) : 2890 - 2898