High Early Failure Rate of the Columbus® Posterior Stabilized High-flexion Knee Prosthesis

被引:19
|
作者
Jung, Woon-Hwa [2 ]
Jeong, Jae-Heon [1 ]
Ha, Yong-Chan [1 ]
Lee, Young-Kyun [3 ]
Koo, Kyung-Hoi [3 ]
机构
[1] Chung Ang Univ, Dept Orthopaed Surg, Coll Med, Seoul 156755, South Korea
[2] Murup Hosp, Dept Orthopaed Surg, Chang Won, South Korea
[3] Seoul Natl Univ, Bundang Hosp, Dept Orthopaed Surg, Songnam, South Korea
关键词
RECURRENT DISLOCATION; CRUCIATE LIGAMENT; FEMORAL COMPONENT; ARTHROPLASTY; REPLACEMENT; DESIGN; MOTION; SYSTEM;
D O I
10.1007/s11999-011-2202-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Most studies report high survivorship rates for TKAs, however, we observed higher than anticipated rates of dislocation and femoral component loosening after implanting a Columbus (R) posterior stabilized prosthesis. Questions/purpose We therefore determined (1) the incidence of dislocation and aseptic loosening that occurred after implantation of posterior stabilized high-flexion prostheses in TKAs, (2) the causative factors of dislocation and aseptic femoral component loosening when comparing two designs of prostheses, and (3) the mechanisms of dislocation. Methods We retrospectively reviewed 319 patients who underwent 384 TKAs from May 2007 to July 2008. These patients had been assigned alternately to receive a Scorpio (R) posterior stabilized knee prosthesis (Group I, 158 patients, 187 knees) or a Columbus (R) posterior stabilized knee prosthesis (Group II, 161 patients, 197 knees). We followed the patients with clinical and radiographic evaluations for a minimum of 24 months (mean, 26 months; range, 2438 months). Ten retrieved prostheses were examined visually. Results Ten dislocations (5.1%; 10 of 197 knees) and seven aseptic loosenings of femoral components (3.6%; seven of 197 knees) occurred in Group II at a mean of 10.9 months postoperatively. However, no dislocation or loosening occurred in patients in Group I. Most dislocations were associated with varus flexion or flexion rotation movements during normal daily activities. The cam jump distance at 908 flexion for the Columbus (R) prosthesis was lower than for the Scorpio (R) prosthesis. Conclusions We observed a high rate of early failure during short-term followup after implantation of a Columbus (R) posterior stabilized prosthesis. It appears that early failures of the Columbus (R) design were related to a different cam-post design attributable to a low jump distance during knee flexion. We no longer recommend using this device.
引用
收藏
页码:1472 / 1481
页数:10
相关论文
共 50 条
  • [41] THE POSTERIOR STABILIZED KNEE PROSTHESIS
    SCUDERI, GR
    INSALL, JN
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 1989, 20 (01) : 71 - 78
  • [42] A high-flexion design total knee prosthesis: a ten to twelve-year follow-up study
    Kim, Man Soo
    Choi, Keun Young
    Hur, Jae Hyeong
    In, Yong
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2024, 19 (01):
  • [43] A review on flexion angle in high-flexion total knee arthroplasty for indonesian's need
    Darmanto, D.
    Novriansyah, R.
    Anggoro, P. W.
    Ismail, R.
    Jamari, J.
    Bayuseno, A. P.
    FRONTIERS IN MECHANICAL ENGINEERING-SWITZERLAND, 2022, 8
  • [44] Long-Term Outcomes of High-Flexion Design Total Knee Arthroplasty with a Short Posterior Flange
    Lee, Chang-Rack
    Park, Dae-Hyun
    Heo, Ki-Seong
    Jo, Se-Myoung
    Seo, Kyung-Jae
    Seo, Seung-Suk
    CLINICS IN ORTHOPEDIC SURGERY, 2024, 16 (02) : 251 - 258
  • [45] High-flexion posterior-substituting versus cruciate-retaining prosthesis in total knee arthroplasty: functional outcome, range of motion and complication comparison
    Zhenxiang Zhang
    Wei Zhu
    Wenjie Zhang
    Archives of Orthopaedic and Trauma Surgery, 2015, 135 : 119 - 124
  • [46] High-flexion posterior-substituting versus cruciate-retaining prosthesis in total knee arthroplasty: functional outcome, range of motion and complication comparison
    Zhang, Zhenxiang
    Zhu, Wei
    Zhang, Wenjie
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2015, 135 (01) : 119 - 124
  • [47] Midterm outcomes of high-flexion total knee arthroplasty on Japanese lifestyle
    Ohno, Hiroshi
    Murata, Minoru
    Ozu, Satoshi
    Matsuoka, Nariyasu
    Kawamura, Hiroshi
    Iida, Hirokazu
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2016, 50 (05) : 527 - 532
  • [48] Improved Clinical Outcomes After High-Flexion Total Knee Arthroplasty
    Seng, Chusheng
    Yeo, Seng Jin
    Wee, James L.
    Subanesh, S.
    Chong, Hwee Chi
    Lo, Ngai Nung
    JOURNAL OF ARTHROPLASTY, 2011, 26 (07): : 1025 - 1030
  • [49] Does a mobile-bearing, high-flexion design increase knee flexion after total knee replacement?
    Nutton, R. W.
    Wade, F. A.
    Coutts, F. J.
    Van der Linden, M. L.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (08): : 1051 - 1057
  • [50] High-flexion knee designs: More hype than hope? In the affirmative
    Ritter, Merrill A.
    JOURNAL OF ARTHROPLASTY, 2006, 21 (04): : 40 - 41