A novel approach to background subtraction in contrast-enhanced dual-energy digital mammography with commercially available mammography devices: Polychromaticity correction

被引:8
|
作者
Contillo, Adriano [1 ]
Di Domenico, Giovanni [1 ]
Cardarelli, Paolo [1 ]
Gambaccini, Mauro [1 ]
Taibi, Angelo [1 ]
机构
[1] Univ Ferrara, Dipartimento Fis & Sci Terra, I-44122 Ferrara, Italy
关键词
digital mammography; contrast medium; dual-energy; breast phantom; beam hardening; correction; COMPUTED TOMOGRAPHIC MAMMOGRAPHY; MEDICAL IMAGING FACILITY; X-RAY-SPECTRA; SCREENING MAMMOGRAPHY; ACCURATE METHOD; BREAST; OPTIMIZATION; RADIOGRAPHY; PERFORMANCE; CALIBRATION;
D O I
10.1118/1.4933199
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Contrast-enhanced digital mammography is an image subtraction technique that is able to improve the detectability of lesions in dense breasts. One of the main sources of error, when the technique is performed by means of commercial mammography devices, is represented by the intrinsic polychromaticity of the x-ray beams. The aim of the work is to propose an iterative procedure, which only assumes the knowledge of a small set of universal quantities, to take into account the polychromaticity and correct the subtraction results accordingly. Methods: In order to verify the procedure, it has been applied to an analytical simulation of a target containing a contrast medium and to actual radiographs of a breast phantom containing cavities filled with a solution of the same medium. Results: The reconstructed densities of contrast medium were compared, showing very good agreement between the theoretical predictions and the experimental results already after the first iteration. Furthermore, the convergence of the iterative procedure was studied, showing that only a small number of iterations is necessary to reach limiting values. Conclusions: The proposed procedure represents an efficient solution to the polychromaticity issue, qualifying therefore as a viable alternative to inverse-map functions. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:6641 / 6650
页数:10
相关论文
共 50 条
  • [41] Application of BI-RADS Descriptors in Contrast-Enhanced Dual-Energy Mammography: Comparison with MRI
    Knogler, Thomas
    Homolka, Peter
    Hoernig, Mathias
    Leithner, Robert
    Langs, Georg
    Waitzbauer, Martin
    Pinker, Katja
    Leitner, Sabine
    Helbich, Thomas H.
    BREAST CARE, 2017, 12 (04) : 212 - 216
  • [42] Dual-energy contrast-enhanced digital mammography: Glandular dose estimation using a Monte Carlo code and voxel phantom
    Tzamicha, E.
    Yakoumakis, E.
    Tsalafoutas, I. A.
    Dimitriadis, A.
    Georgiou, E.
    Tsapaki, V.
    Chalazonitis, A.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (07): : 785 - 791
  • [43] Technique and clinical applications of dual-energy contrast-enhanced digital mammography (CEDM) in breast cancer evaluation: a pictorial essay
    Sudhir, Rashmi
    Koppula, Veeraiah
    Mandava, Anitha
    Kamala, Sannapareddy
    Potlapalli, Alekya
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2021, 27 (01) : 28 - 36
  • [44] A method for material decomposition in dual-energy contrast enhancement digital mammography
    Kahani, M.
    Kamali-asl, A.
    Ghadiri, H.
    Hashemi, S.
    MEASUREMENT, 2016, 88 : 87 - 95
  • [45] A dual-energy subtraction imaging technique for enhanced mircrocalcification imaging and tissue composition measurement in digital mammography
    Shaw, CC
    Liu, X
    Whitman, GJ
    RADIOLOGY, 1999, 213P : 368 - 368
  • [46] Contrast-Medium-Enhanced Digital Mammography: Comparison of Standard vs. High-Energy Spectra for Dual-Energy Temporal Subtraction
    Rosado-Mendez, I.
    Villasenor, Y.
    Benitez-Bribiesca, L.
    Brandan, M. E.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 458 - 461
  • [47] Dual-energy approach to contrast-enhanced mammography using the balanced filter method: Spectral optimization and preliminary phantom measurement
    Saito, Masatoshi
    MEDICAL PHYSICS, 2007, 34 (11) : 4236 - 4246
  • [48] Algorithmic scatter correction in dual-energy digital mammography for calcification imaging
    Chen, Xi
    Nishikawa, Robert M.
    Chan, Suk-tak
    Lau, Beverly A.
    Zhang, Lei
    Mou, Xuanqin
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [49] A scatter correction method for contrast-enhanced dual-energy digital breast tomosynthesis
    Lu, Yihuan
    Peng, Boyu
    Lau, Beverly A.
    Hu, Yue-Houng
    Scaduto, David A.
    Zhao, Wei
    Gindi, Gene
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (16): : 6323 - 6354
  • [50] Scatter-to-primary ratio and grid performance in contrast-enhanced dual-energy digital mammography using Monte Carlo simulations
    Cunha, D. M.
    Godeli, J.
    Poletti, M. E.
    RADIATION PHYSICS AND CHEMISTRY, 2025, 232