摘要
This study involves the failure analysis of an ultra high molecular polyethylene (UHMWPE) tibial insert from Apollo (R) Total Knee System, which was removed after 10 years of service from an 70 years old female patient. The tibial insert was investigated by using a stereoscope, scanning electron microscope (SEM), infinite focus microscope (IFM) and energy disperse spectroscopy (EDS) to characterize the morphology and composition of the bearing surface. Differential scanning calorimetry (DSC) and Fourier transform spectroscopy (FTIR) were employed to characterize the degradation and crystallinity of the component. Gel-permeation chromatography (GPC) was used to measure the polyethylene tibial insert molecular weight. Results showed that the failure of total knee replacement (TKR) was associated with high grade wear with oxidation degradation. Surface delamination, scratch marks, pitting, folding, and embedded third body particles were observed on the retrieved UHMWPE tibial surface. The damage features observed on the UHMWPE tibial insert suggested degradation is due to fatigue related wear and oxidative-induced. From the depth measurement of the pit, the depths are large as 35 mu m and 60 mu m. Overall results showed that the UHMWPE tibial insert which retrieved from the patient who is an active and not overweight have a decreases in material properties and high grade wear that can contribute to the failure of the tibial insert after 10 years used.
引用
收藏
页码:73 / +
页数:3
相关论文
共 50 条
  • [31] The Effect of Malrotation of Tibial Component of Total Knee Arthroplasty on Tibial Insert during High Flexion Using a Finite Element Analysis
    Osano, Kei
    Nagamine, Ryuji
    Todo, Mitsugu
    Kawasaki, Makoto
    SCIENTIFIC WORLD JOURNAL, 2014,
  • [32] COLD-EXTRUSION DEFORMATION OF UHMWPE IN TOTAL KNEE REPLACEMENT PROSTHESES
    CHANG, DCM
    GOH, JCH
    TEOH, SH
    BOSE, K
    BIOMATERIALS, 1995, 16 (03) : 219 - 223
  • [33] Bilateral condyle fracture of tibial insert in mobile bearing total knee arthroplasty
    Yoon, Jung-Ro
    Jeong, Hyeon-Il
    Oh, Kwang-Jun
    Yang, Jae-Hyuk
    KNEE, 2014, 21 (01): : 318 - 321
  • [34] Isolated Tibial Polyethylene Insert Exchange Outcomes After Total Knee Arthroplasty
    Willson, Seann E.
    Munro, Michelle L.
    Sandwell, Julie C.
    Ezzet, Kace A.
    Colwell, Clifford W., Jr.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (01) : 96 - 101
  • [35] Angled polyethylene insert exchange for sagittal tibial malalignment in total knee arthroplasty
    Sah, Alexander P.
    Scott, Richard D.
    Iorio, Richard
    JOURNAL OF ARTHROPLASTY, 2008, 23 (01): : 141 - 144
  • [36] Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition
    Khadijah, S. M. P.
    Liza, S.
    Shahemi, N.
    Merican, A. M.
    Abbas, A. A.
    Ayob, K. A.
    ENGINEERING FAILURE ANALYSIS, 2020, 118 (118)
  • [37] Change in UHMWPE properties of retrieved ceramic total knee prosthesis in clinical use for 23 years
    Oonishi, H
    Kim, SC
    Kyomoto, M
    Masuda, S
    Asano, T
    Clarke, IC
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2005, 74B (02) : 754 - 759
  • [38] A QUANTITATIVE TECHNIQUE FOR REPORTING SURFACE DEGRADATION PATTERNS OF UHMWPE COMPONENTS OF RETRIEVED TOTAL KNEE REPLACEMENTS
    CORNWALL, GB
    BRYANT, JT
    HANSSON, CM
    RUDAN, J
    KENNEDY, LA
    COOKE, TDV
    JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) : 9 - 18
  • [39] Total knee replacement with tibial tubercle osteotomy in rheumatoid patients with stiff knee
    Eid, Ahmed Salem
    Nassar, Wael Ahmed Mohamed
    Fayyad, Tamer Abdelmeguid Mohamed
    INTERNATIONAL ORTHOPAEDICS, 2016, 40 (11) : 2289 - 2293
  • [40] Total knee replacement with tibial tubercle osteotomy in rheumatoid patients with stiff knee
    Ahmed Salem Eid
    Wael Ahmed Mohamed Nassar
    Tamer Abdelmeguid Mohamed Fayyad
    International Orthopaedics, 2016, 40 : 2289 - 2293