Failure Analysis of Retrieved UHMWPE Tibial Insert in Total Knee Replacement

被引:0
|
作者
Liza, S. [1 ]
Haseeb, A. S. M. A. [1 ]
Abbas, A. A. [2 ]
Masjuki, H. H. [1 ]
机构
[1] Univ Malaya, Fac Engn, Dept Mech Engn, Kuala Lumpur, Malaysia
[2] Univ Malaya, Fac Med Building, Dept Orthopaed Surg, Kuala Lumpur, Malaysia
关键词
UHMWPE; retrieval study; total knee replacement; wear; oxidation; ACETABULAR LINERS; POLYETHYLENE INSERTS; GAMMA-STERILIZATION; ORTHOPEDIC UHMWPE; WEAR; ARTHROPLASTY; PROSTHESES; OXIDATION; COMPONENTS; PARTICLES;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study involves the failure analysis of an ultra high molecular polyethylene (UHMWPE) tibial insert from Apollo (R) Total Knee System, which was removed after 10 years of service from an 70 years old female patient. The tibial insert was investigated by using a stereoscope, scanning electron microscope (SEM), infinite focus microscope (IFM) and energy disperse spectroscopy (EDS) to characterize the morphology and composition of the bearing surface. Differential scanning calorimetry (DSC) and Fourier transform spectroscopy (FTIR) were employed to characterize the degradation and crystallinity of the component. Gel-permeation chromatography (GPC) was used to measure the polyethylene tibial insert molecular weight. Results showed that the failure of total knee replacement (TKR) was associated with high grade wear with oxidation degradation. Surface delamination, scratch marks, pitting, folding, and embedded third body particles were observed on the retrieved UHMWPE tibial surface. The damage features observed on the UHMWPE tibial insert suggested degradation is due to fatigue related wear and oxidative-induced. From the depth measurement of the pit, the depths are large as 35 mu m and 60 mu m. Overall results showed that the UHMWPE tibial insert which retrieved from the patient who is an active and not overweight have a decreases in material properties and high grade wear that can contribute to the failure of the tibial insert after 10 years used.
引用
收藏
页码:73 / +
页数:3
相关论文
共 50 条
  • [1] Failure analysis of retrieved UHMWPE tibial insert in total knee replacement
    Liza, S.
    Haseeb, A. S. M. A.
    Abbas, A. A.
    Masjuki, H. H.
    ENGINEERING FAILURE ANALYSIS, 2011, 18 (06) : 1415 - 1423
  • [2] Failure Analysis of Retrieved Polyethylene Insert in Total Knee Replacement
    Grecu, Dan
    Antoniac, Iulian
    Trante, Octavian
    Niculescu, Marius
    Lupescu, Olivera
    MATERIALE PLASTICE, 2016, 53 (04) : 776 - 780
  • [3] Mechanical Properties and Tribological Behaviour of Retrieved UHMWPE Tibial Insert in Total Knee Replacement after Implantation 30 Months
    Kang, Xueqin
    Ge, Shirong
    Dai, Xiaofeng
    POLYMERS & POLYMER COMPOSITES, 2014, 22 (06): : 561 - 567
  • [4] A New Method of In Vitro Wear Assessment of the UHMWPE Tibial Insert in Total Knee Replacement
    Affatato, Saverio
    Cristofolini, Luca
    Leardini, Walter
    Erani, Paolo
    Zavalloni, Mara
    Tigani, Domenico
    Viceconti, Marco
    ARTIFICIAL ORGANS, 2008, 32 (12) : 942 - 948
  • [5] Comparative Failure Analysis of Retrieved UHMWPE Tibial Components
    Camacho, N.
    Stafford, S. W.
    Trueba, L., Jr.
    MEDICAL DEVICE MATERIALS VI, 2013, : 63 - 68
  • [6] Tibial interface wear in retrieved total knee components and correlations with modular insert motion
    Rao, AR
    Engh, GA
    Collier, MB
    Lounici, S
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (10): : 1849 - 1855
  • [7] Surface analysis of early retrieved polyethylene tibial inserts for both knees in total knee replacement
    Tan, M. Y.
    Liza, S.
    Khadijah, S. M. P.
    Abbas, A. A.
    Merican, A. M.
    Ayob, K. A.
    Zulkifli, N. W. M.
    Masjuki, H. H.
    ENGINEERING FAILURE ANALYSIS, 2020, 109
  • [8] Contact analysis of UHMWPE tibial component based on geometrical measurement for retrieved knee prosthesis
    Cho, Changhee
    Murakami, Teruo
    Sawae, Yoshinori
    Sakai, Nobuo
    Miura, Hiromasa
    Kawano, Tsutomu
    Iwamoto, Yukihide
    Nippon Kikai Gakkai Ronbunshu, C Hen/Transactions of the Japan Society of Mechanical Engineers, Part C, 2002, 68 (04): : 1284 - 1290
  • [9] The effect of tibial insert design on rising from a chair;: motion analysis after total knee replacement
    Saari, T
    Tranberg, R
    Zügner, R
    Uvehammer, J
    Kärrholm, J
    CLINICAL BIOMECHANICS, 2004, 19 (09) : 951 - 956
  • [10] FAILURE ANALYSIS OF RETRIEVED CEMENTED TOTAL KNEE PROSTHESES
    HALVORSON, TL
    COOK, SD
    SANDBORN, PM
    BRUNET, ME
    SOUTHERN MEDICAL JOURNAL, 1988, 81 (09) : S54 - S54