Globally Optimal Superconducting Homogeneous Magnet Design for an Asymmetric 3.0 T Head MRI Scanner

被引:6
|
作者
Ni, Zhipeng [1 ]
Hu, Geli [1 ]
Wang, Qiuliang [1 ]
Yan, Luguang [1 ]
机构
[1] Chinese Acad Sci, Inst Elect Engn, Key Lab Appl Superconduct, Beijing 100190, Peoples R China
基金
中国国家自然科学基金;
关键词
Actively-shielded technique; asymmetric structure; head imaging; hybrid numerical method; MINIMUM STORED ENERGY;
D O I
10.1109/TASC.2013.2283653
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes a hybrid numerical method with a combination of linear programming (LP) and nonlinear programming to design an asymmetric magnetic resonance imaging (MRI) system for head imaging. The method was successfully employed to a 0.7 T open-style MRI. Recently, we have updated the LP mathematical model in the hybrid optimization strategy by adding maximum axial and radial magnetic field strength limitation to make the magnet safe. The whole magnet consists of eight coaxial coils asymmetry about the z-axis. The six innermost coils are the main coils with positive current direction and the two outermost coils are shielding coils with negative current direction. All coils contribute their efforts to produce a central magnetic field strength of 3.0 T over a 20 cm diameter of spherical volume, and the peak to peak homogeneity is 12 ppm. The temperature bore for the magnet is formed with two different diameters cylindrical bores which makes head and shoulder access to the imaging volume easier. A 5 Gauss footprint area is restricted larger than an elliptical region with the sizes of 3.5 m in radial and 4.0 m in axial direction. The results show that the methodology is very flexible and efficient for asymmetric MRI magnet design.
引用
收藏
页数:5
相关论文
共 50 条
  • [41] Design of a 15 T hybrid superconducting magnet
    Gao, P.
    Xue, S. Q.
    Long, F.
    Liu, F.
    Liu, H. J.
    Wu, Y.
    Li, J. G.
    Qin, J. G.
    Zhou, C.
    FUSION ENGINEERING AND DESIGN, 2021, 167
  • [42] Design of a 15 T hybrid superconducting magnet
    Gao, P.
    Xue, S.Q.
    Long, F.
    Liu, F.
    Liu, H.J.
    Wu, Y.
    Li, J.G.
    Qin, J.G.
    Zhou, C.
    Zhou, C. (chao.zhou@ipp.ac.cn), 1600, Elsevier Ltd (167):
  • [43] OPTIMAL-DESIGN OF A SUPERCONDUCTING MHD SADDLE MAGNET
    LIA, GMA
    MONTANARI, I
    RIBANI, PL
    IEEE TRANSACTIONS ON MAGNETICS, 1992, 28 (01) : 466 - 469
  • [44] Design, Fabrication and Measurement of a 3 T Cryogen-Free Animal MRI Superconducting Magnet
    Wang, Yaohui
    Liu, Hui
    Chen, Shunzhong
    Wang, Weimin
    Liu, Feng
    Wang, Qiuliang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2024, 34 (05) : 1 - 4
  • [45] A Novel Design Method of Passive Shimming for 0.7-T Biplanar Superconducting MRI Magnet
    Zhu, Xuchen
    Wang, Houshen
    Wang, Hui
    Li, Yi
    Fang, Youtong
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2016, 26 (07)
  • [46] Resistive homogeneous MRI magnet design by matrix subset selection
    Morgan, PN
    Conolly, SM
    Macovski, A
    MAGNETIC RESONANCE IN MEDICINE, 1999, 41 (06) : 1221 - 1229
  • [47] Multiphysics Stress Analysis of a 1.5 T Superconducting MRI Magnet
    Thekkethil, Sankar Ram
    Rastogi, Vikas
    Kar, Soumen
    JOURNAL OF SUPERCONDUCTIVITY AND NOVEL MAGNETISM, 2023, 36 (02) : 467 - 476
  • [48] 1.5 T Cryogen Free Superconducting Magnet for Dedicated MRI
    Rybakov, Alexander
    Bagdinov, Anton
    Demikhov, Evgeniy
    Kostrov, Evgeniy
    Lysenko, Valeriy
    Piskunov, Nikolay
    Tysyachnykh, Yuriy
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2016, 26 (04)
  • [49] Multiphysics Stress Analysis of a 1.5 T Superconducting MRI Magnet
    Sankar Ram Thekkethil
    Vikas Rastogi
    Soumen Kar
    Journal of Superconductivity and Novel Magnetism, 2023, 36 : 467 - 476
  • [50] Design optimization of superconducting MRI magnet with polygonal iron yoke
    Li, Qingfu
    Yang, Xin
    Liu, Yunsong
    Hsi-An Chiao Tung Ta Hsueh/Journal of Xi'an Jiaotong University, 1996, 30 (11): : 14 - 19