Bicruciate substituting total knee replacement: how effective are the added kinematic constraints in vivo?

被引:28
|
作者
van Duren, B. H. [1 ]
Pandit, H. [1 ,2 ]
Price, M. [3 ]
Tilley, S. [3 ]
Gill, H. S. [1 ]
Murray, D. W. [1 ,2 ]
Thomas, N. P. [3 ]
机构
[1] Univ Oxford, NDORMS, Oxford, England
[2] Nuffield Orthopaed Ctr, Oxford OX3 7LD, England
[3] N Hampshire Hosp, Basingstoke, Hants, England
关键词
In vivo; Kinematics; Posterior stabilised TKR; Patella tendon angle; PFA; Cam-post mechanism; Knee; Arthroplasty; Fluoroscopy; Bicruciate retaining; FLUOROSCOPIC ANALYSIS; VITRO MEASUREMENT; MOBILE-BEARING; ARTHROPLASTY; FLEXION; RECONSTRUCTION; MODEL;
D O I
10.1007/s00167-011-1796-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The Journey bicruciate substituting (BCS) TKR was designed to restore normal knee kinematics. It has two cam-post mechanisms which substitute for the ACL and PCL. The aim of this study was to undertake a comprehensive study of the Journey BCS kinematics in vivo to assess the function of the cam-post mechanisms and their effect on functional kinematics and compared to the kinematics of a group of normal knees. The kinematics of 10 Journey BCS were assessed fluoroscopically during step-up and lunge exercises, and were compared to those of 20 normal knees. The fluoroscopic images were used to determine relative implant orientation using a 2D to 3D reconstruction method. The determined relative tibio-femoral orientations allowed for cam-post engagement and tibio-femoral contact points to be determined. Functional kinematics were assessed using the patella tendon angle (PTA) and the patella flexion angle (PFA) relationship with the knee flexion angle (KFA). The average maximum flexion achieved by the Journey was 124.7A degrees. Both cam mechanisms engaged: The anterior cam during extension at 12.6A degrees and the posterior cam in flexion at 45.4A degrees. During flexion, the contacts points on the tibia moved posteriorly with no paradoxical anterior translation. The PTA/KFA relationships of the Journey implant group for both the step-up and lunge exercises were broadly similar in terms of trend to those established for the normal knee but the PTA between 10A degrees KFA and 140A degrees KFA were significantly (P < 0.05) lower than that for the normal knees. The PFA/KFA trend for both the implant and normal groups showed a linear relationship; however, the values of PFA were higher for the Journey compared to the normal. The Journey BCS showed no paradoxical anterior movement and sufficient posterior femoral roll back which corresponded with the engagement of the anterior and posterior cam-post mechanisms. Trends shown by the PTA/KFA and PFA/KFA kinematic profiles observed for the Journey group were more normal than those seen with other designs of TKR. However, despite being more close to normal than other implants, the Journey group showed a different kinematic profile to that of the normal knees, which is most likely due to the femur being too far posterior relative to the tibia. Case-control study, retrospective, comparative study, Level III.
引用
收藏
页码:1998 / 2006
页数:9
相关论文
共 50 条
  • [31] Comparison of wear in a total knee replacement under different kinematic conditions
    Petra I. Barnett
    John Fisher
    Daniel D. Auger
    Martin H. Stone
    Eileen Ingham
    Journal of Materials Science: Materials in Medicine, 2001, 12 : 1039 - 1042
  • [32] Intraoperative evaluation of total knee replacement: kinematic assessment with a navigation system
    Casino, Daniela
    Zaffagnini, Stefano
    Martelli, Sandra
    Lopomo, Nicola
    Bignozzi, Simone
    Iacono, Francesco
    Russo, Alessandro
    Marcacci, Maurilio
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2009, 17 (04) : 369 - 373
  • [33] In vivo knee kinematics during high flexion after a posterior-substituting total knee arthroplasty
    Moynihan, Angela L.
    Varadarajan, Kartik M.
    Hanson, George R.
    Park, Sang-Eun
    Nha, Kyung Wook
    Suggs, Jeremy F.
    Johnson, Todd
    Li, Guoan
    INTERNATIONAL ORTHOPAEDICS, 2010, 34 (04) : 497 - 503
  • [34] In vivo knee kinematics during high flexion after a posterior-substituting total knee arthroplasty
    Angela L. Moynihan
    Kartik M. Varadarajan
    George R. Hanson
    Sang-Eun Park
    Kyung Wook Nha
    Jeremy F. Suggs
    Todd Johnson
    Guoan Li
    International Orthopaedics, 2010, 34 : 497 - 503
  • [35] In vivo kinematic determination of total knee arthroplasty from squatting to standing
    Liu, Fei
    Ohdera, Toshihiro
    Miyamoto, Hirohide
    Wasielewski, Ray C.
    Komistek, Richard D.
    Mahfouz, Mohamed R.
    KNEE, 2009, 16 (02): : 116 - 120
  • [36] In vivo kinematical validated knee model for preclinical testing of total knee replacement
    Shu, Liming
    Yao, Jiang
    Yamamoto, Ko
    Sato, Takashi
    Sugita, Naohiko
    COMPUTERS IN BIOLOGY AND MEDICINE, 2021, 132
  • [37] Kinematic behaviour and soft tissue management in guided motion total knee replacement
    Camilla Halewood
    Michael Risebury
    Neil P. Thomas
    Andrew A. Amis
    Knee Surgery, Sports Traumatology, Arthroscopy, 2014, 22 : 3074 - 3082
  • [38] Kinetic and kinematic gait analysis before and after total knee replacement.
    Perka, C
    Möckel, G
    Boenick, U
    ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2000, 138 (03): : 191 - 196
  • [39] Kinematic navigation in total knee replacement - Experience from the first 50 cases
    Chang, Chih-Wei
    Yang, Chyun-Yu
    JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 2006, 105 (06) : 468 - 474
  • [40] Kinematic behaviour and soft tissue management in guided motion total knee replacement
    Halewood, Camilla
    Risebury, Michael
    Thomas, Neil P.
    Amis, Andrew A.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2014, 22 (12) : 3074 - 3082