Intercomparison of lateral scales of scanning electron microscopes and atomic force microscopes in research institutes in Northern Europe

被引:4
|
作者
Seppa, Jeremias [1 ]
Korpelainen, Virpi [1 ]
Bergstrand, Sten [2 ]
Karlsson, Helge [3 ]
Lillepea, Lauri [4 ]
Lassila, Antti [1 ]
机构
[1] Ctr Metrol & Accreditat MIKES, FI-02151 Espoo, Finland
[2] SP Tech Res Inst Sweden, SE-50115 Boras, Sweden
[3] Justervesenet, NO-2027 Kjeller, Norway
[4] AS Metrosert, EE-12618 Tallinn, Estonia
关键词
SEM; AFM; comparison; scale; calibration;
D O I
10.1088/0957-0233/25/4/044013
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
An intercomparison of lateral scales of scanning electron microscopes (SEM) and atomic force microscopes (AFM) in various research laboratories in Northern Europe was organized by the local national metrology institutes. In this paper are presented the results of the comparison, with also an example uncertainty budget for AFM grating pitch measurement. Grating samples (1D) were circulated among the participating laboratories. The participating laboratories were also asked about the calibration of their instruments. The accuracy of the uncertainty estimates seemed to vary largely between the laboratories, and for some laboratories the appropriateness of the calibration procedures could be considered. Several institutes (60% of all results in terms of E-n value) also had good comprehension of their measurement capability. The average difference from reference value was 6.7 and 10.0 nm for calibrated instruments and 20.6 and 39.9 nm for uncalibrated instruments for 300 nm and 700 nm gratings, respectively. The correlation of the results for both nominally 300 and 700 nm gratings shows that a simple scale factor calibration would have corrected a large part of the deviations from the reference values.
引用
收藏
页数:7
相关论文
共 40 条
  • [21] Research Activities of Nanodimensional Standards Using Atomic Force Microscopes, Transmission Electron Microscope, and Scanning Electron Microscope at the National Metrology Institute of Japan (vol 5, pg 83,2022)
    Misumi, Ichiko
    Kizu, Ryosuke
    Itoh, Hiroshi
    Kumagai, Kazuhiro
    Kobayashi, Keita
    Sigehuzi, Tomoo
    NANOMANUFACTURING AND METROLOGY, 2022, 5 (04) : 445 - 445
  • [22] CYLINDRICAL VESICLES AS DIAGNOSTIC CELLULAR INDICES IN CARDIORENAL SYNDROME TYPES III AND IV: BIOPSY OF PERIPHERAL BLOOD SMEARS WITH THE ATOMIC FORCE AND SCANNING ELECTRON MICROSCOPES
    Stamopoulos, Dimosthenis
    Mpakirtzi, Nerantzoula
    Kotsovassilis, Costantine
    Pirgakis, Vlassios
    Afentakis, Nikolaos
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2015, 30
  • [23] CRYSTALLIZATION OF ELECTROLYTES OF THE REPLACEMENT MEDIUM AT THE PERIPHERY OF RED BLOOD CELLS IN THERAPEUTIC PLASMA EXCHANGE: AN INVESTIGATION WITH SCANNING-ELECTRON AND ATOMIC-FORCE MICROSCOPES
    Stamopoulos, Dimosthenis
    Mpakirtzi, Nerantzoula
    Gogola, Basiliki
    Zeibekis, Manos
    Stivarou, Dora
    Panagiotou, Maria
    Grapsa, Eirini
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2012, 27 : 224 - 225
  • [24] The use of a special work station for in situ measurements of highly reactive electrochemical systems by atomic force and scanning tunneling microscopes
    Cohen, Y
    Aurbach, D
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (12): : 4668 - 4675
  • [25] Application of Tuning Fork Sensors for In-situ Studies of Dynamic Force Interactions Inside Scanning and Transmission Electron Microscopes
    Andzane, Jana
    Poplausks, Raimonds
    Prikulis, Juris
    Lohmus, Ruenno
    Vlassov, Sergei
    Kubatkin, Sergey
    Erts, Donats
    MATERIALS SCIENCE-MEDZIAGOTYRA, 2012, 18 (02): : 197 - 201
  • [26] AU(111) 23X-ROOT-3 SURFACE AS A TEST SURFACE FOR COMPARING THE ATOMIC FORCE AND SCANNING TUNNELING MICROSCOPES
    ODEN, PI
    TAO, NJ
    LINDSAY, SM
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1993, 11 (02): : 137 - 140
  • [27] Bent-fiber near-field scanning optical microscopy probes for use with commercial atomic-force microscopes
    Taylor, RS
    Leopold, KE
    Wendman, M
    Gurley, G
    Elings, V
    MICROMACHINING AND IMAGING, 1997, 3009 : 119 - 129
  • [28] OBSERVATION OF BINDING AND POLYMERIZATION OF FUR REPRESSOR ONTO OPERATOR-CONTAINING DNA WITH ELECTRON AND ATOMIC-FORCE MICROSCOPES
    LECAM, E
    FRECHON, D
    BARRAY, M
    FOURCADE, A
    DELAIN, E
    PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1994, 91 (25) : 11816 - 11820
  • [29] The effects of two-dimensional bifurcations and quantum beats in a system of combined atomic force and scanning tunneling microscopes with quantum dots
    V. Ch. Zhukovsky
    V. D. Krevchik
    M. B. Semenov
    P. V. Krevchik
    R. V. Zaytsev
    I. A. Egorov
    Moscow University Physics Bulletin, 2016, 71 : 545 - 555
  • [30] The Effects of Two-Dimensional Bifurcations and Quantum Beats in a System of Combined Atomic Force and Scanning Tunneling Microscopes with Quantum Dots
    Zhukovsky, V. Ch.
    Krevchik, V. D.
    Semenov, M. B.
    Krevchik, P. V.
    Zaytsev, R. V.
    Egorov, I. A.
    MOSCOW UNIVERSITY PHYSICS BULLETIN, 2016, 71 (06) : 545 - 555