Performance evaluation of a particle-enhanced turbidimetric cystatin C assay on the Hitachi 917 analyzer

被引:12
|
作者
Al-Turkmani, M. Rabie
Law, Terence
Kellogg, Mark D. [1 ]
机构
[1] Childrens Hosp, Dept Lab Med, Boston, MA 02115 USA
关键词
Cystatin C; Turbidimetric; Immunoassay; Method evaluation;
D O I
10.1016/j.cca.2008.08.016
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Human cystatin C is a low molecular weight protein that has been proposed as a new endogenous marker of glomerular filtration rate. We investigated the performance of the Genzyme cystatin C assay on the Hitachi 917 analyzer. Methods: Imprecision. linearity. recovery, and interference studies were performed on the Hitachi 917 analyzer. For method comparison, split sample aliquots were assayed using the described method and 2 other commercially available cystatin C assays. Results: The assay was linear from 0.24 to 6.36 mg/l. Within-run coefficient of variation (CV) was 4.2 and 0.8% at cystatin C concentrations of 0.50 and 2.00 mg/l, respectively. Between-run CV was 4.3 and 2.7% at the same concentrations. The average analytical recovery was 99%. Bilirubin (<= 30 mg/dl), triglycerides (<= 1000 mg/dl), intralipid (L index <= 1000), and rheumatoid factor (<= 1000 IU/ml) did not interfere with the assay. A > 10% change in cystatin C level was observed when hemoglobin concentration was > 800 mg/dl. The assay compared well with the Dade Behring immunonephelometric assay and the Dako immunoturbidimetric assay. Conclusion: The Genzyme cystatin C immunoassay is an acceptable method for the determination of cystatin C on the Hitachi 917 analyzer. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:75 / 77
页数:3
相关论文
共 50 条
  • [21] PERFORMANCE EVALUATION OF PARTICLE-ENHANCED TURBIDIMETRIC IMMUNOASSAY FOR ANTINUCLEAR ANTIBODIES DETECTION IN COMPARISON WITH LINE IMMUNOASSAY
    Qin, Yan
    Luo, Jing
    Zhao, Xiang-Cong
    ANNALS OF THE RHEUMATIC DISEASES, 2019, 78 : 2132 - 2132
  • [22] Determination of serum lipoprotein lipase using a latex particle-enhanced turbidimetric immunoassay with an automated analyzer
    Machida, Tetsuo
    Miyashita, Kazuya
    Sone, Takuya
    Tanaka, Sayori
    Nakajima, Katsuyuki
    Saito, Masayuki
    Stanhope, Kimber
    Havel, Peter
    Sumino, Hiroyuki
    Murakami, Masami
    CLINICA CHIMICA ACTA, 2015, 442 : 130 - 135
  • [24] Performance evaluation of a turbidimetric cystatin C assay on different high-throughput platforms
    Hansson, Lars-Olof
    Grubb, Anders
    Liden, Anders
    Flodin, Mats
    Berggren, Annacarin
    Delanghe, Joris
    Stove, Veronique
    Luthe, Hilmar
    Rhode, Karl-Heinz
    Beck, Claus
    Domke, Ingrid
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2010, 70 (05): : 347 - 353
  • [25] Analytical validation of a human particle-enhanced nephelometric assay for cystatin C measurement in feline serum and urine
    Ghys, Liesbeth F. E.
    Meyer, Evelyne
    Paepe, Dominique
    Delanghe, Joris
    Daminet, Sylvie
    VETERINARY CLINICAL PATHOLOGY, 2014, 43 (02) : 226 - 234
  • [26] DEVELOPMENT AND VALIDATION OF A PARTICLE-ENHANCED TURBIDIMETRIC IMMUNOASSAY FOR C-REACTIVE PROTEIN
    PRICE, CP
    TRULL, AK
    BERRY, D
    GORMAN, EG
    JOURNAL OF IMMUNOLOGICAL METHODS, 1987, 99 (02) : 205 - 211
  • [27] Analytical performance of a particle-enhanced nephelometric immunoassay for serum cystatin C using rate analysis
    Stowe, H
    Lawrence, D
    Newman, DJ
    Lamb, EJ
    CLINICAL CHEMISTRY, 2001, 47 (08) : 1482 - 1485
  • [28] Evaluation of the new Cedia® sirolimus assay on Hitachi 917
    Liddell, P
    Shumaker, S
    Johnson, S
    THERAPEUTIC DRUG MONITORING, 2005, 27 (02) : 233 - 233
  • [29] Development and validation of a particle-enhanced turbidimetric inhibition assay for urine albumin on the Dade aca(R) analyzer (vol 43, pg 109, 1997)
    Thakkar, H
    Newman, DJ
    Holownia, F
    Davey, CL
    Wang, CC
    Lloyd, J
    CLINICAL CHEMISTRY, 1997, 43 (08) : 1471 - 1471
  • [30] Performance evaluation of the Sysmex Cystatin C assay on the Roche Modular P Analyzer
    He, Min
    Ke, Pei-feng
    Xu, Jian-hua
    Chen, Wei-ye
    Lin, Hai-biao
    Zhang, Cheng
    Wang, Yun-xiu
    Zhuang, Jun-hua
    Huang, Xian-zhang
    CLINICAL BIOCHEMISTRY, 2016, 49 (12) : 915 - 918