Microfluidic chip-based valveless flow injection analysis system with gravity-driven flows

被引:11
|
作者
Huang, Yang-Zhen [1 ]
Du, Wen-Bin [1 ]
Pan, Jian-Zhang [1 ]
Fang, Qun [1 ]
机构
[1] Zhejiang Univ, Dept Chem, Inst Microanalyt Syst, Hangzhou 310058, Zhejiang, Peoples R China
基金
中国国家自然科学基金;
关键词
D O I
10.1039/b807494a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work, a microfluidic chip-based valveless flow injection analysis (FIA) system with gravity-driven flows and liquid-core waveguide (LCW) spectrometric detection was developed. Automated sample injection in the 0.3-6.4 nL range under gated injection mode was achieved by controlling the vertical position of the waste reservoir fixed on a moving platform and the residence time of the reservoir in each position, without the requirement of microvalves or electrokinetic manipulation. An integrated LCW spectrometric detection system was built on the chip by coupling a 20 mm-long Teflon AF 2400 capillary with the microchannel to function as a LCW flow cell, using a green LED as light source and a photodiode as detector. The performance of the system was demonstrated in the determination of [NO2](2-) based on the Saltzman reaction. Linear absorbance response was obtained in the range of 0.1-20 mg L-1 (R-2 = 0.9910), and a good reproducibility of 0.34% RSD (n = 17) was achieved.
引用
收藏
页码:1237 / 1241
页数:5
相关论文
共 50 条
  • [1] High-throughput nanoliter sample introduction microfluidic chip-based flow injection analysis system with gravity-driven flows
    Du, WB
    Fang, Q
    He, QH
    Fang, ZL
    ANALYTICAL CHEMISTRY, 2005, 77 (05) : 1330 - 1337
  • [2] A flow analysis system based on microfluidic chip with continuous sample introduction and gravity-driven flows
    Huang, YZ
    Fang, Q
    Li, DN
    CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE, 2004, 25 (09): : 1628 - 1631
  • [3] Valveless gated injection for microfluidic chip-based liquid chromatography system with polymer monolithic column
    Wang, Xiu-Li
    Zhu, Ying
    Fang, Qun
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1246 : 123 - 128
  • [4] A microfluidic chip-based flow injection system with gas diffusion separation and photometric detection
    Jia Hong-Xin
    Wang Shi-Li
    Xu Zhang-Run
    Fang Zhao-lun
    CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE, 2006, 27 (09): : 1621 - 1625
  • [5] Microfluidic chip-based flow injection system with gas diffusion separation and photometric detection
    Jia, Hong-Xin
    Wang, Shi-Li
    Xu, Zhang-Run
    Fang, Zhao-Lun
    Gaodeng Xuexiao Huaxue Xuebao/Chemical Journal of Chinese Universities, 2006, 27 (09): : 1621 - 1625
  • [6] Microfluidic chip-based flow-injection analysis with high throughput sampling interface
    Fang, Q
    Du, WB
    He, QH
    Fang, ZL
    Micro Total Analysis Systems 2004, Vol 1, 2005, (296): : 195 - 197
  • [7] Microfluidic chip-based flow injection system for gas diffusion separation with an evaporation pump drive
    Jia, Hong-Xin
    Wang, Shi-Li
    Guan, Yan-Xia
    Fang, Zhao-Lun
    PROCEEDINGS OF THE FIRST SHENYANG INTERNATIONAL COLLOQUIUM ON MICROFLUIDICS, 2007, : 90 - 91
  • [8] Paper spray mass spectrometry-based method for analysis of droplets in a gravity-driven microfluidic chip
    Zhang, Yandong
    Li, Haifang
    Ma, Yuan
    Lin, Jin-Ming
    ANALYST, 2014, 139 (05) : 1023 - 1029
  • [9] MODELING DRUG DELIVERY IN GRAVITY-DRIVEN MICROFLUIDIC SYSTEM
    Maki, Antti-Juhana
    Kreutzer, Joose
    Kallio, Pasi
    PROCEEDINGS OF THE ASME 12TH INTERNATIONAL CONFERENCE ON NANOCHANNELS, MICROCHANNELS, AND MINICHANNELS, 2014, 2014,
  • [10] ANALYSIS OF GRAVITY-DRIVEN SLURRY FLOW
    Ashrafi, Nariman
    Zeydabadi, Haniyeh
    PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION - 2012, VOL 7, PTS A-D, 2013, : 1043 - 1049